8
5.
Function Table of Balance
The function table stores parameters that specify how to operate the balance, and it is used to
monitor or update those parameters. These parameters are stored in memory even if the AC
adaptor is removed and are retained until overwritten. Balance operations are specified by this
table when using the GXM
-
04
/
GXL
-
04 and the GXM
-
06
/
GXL
-
06. The table menu consists
of two layers as shown in the diagram below. The first layer is the “
Class
” and the second layer
is the “
Item
”. Each item stores a "
Parameter
". The last displayed parameter is in effect. After
the PRINT key is pressed, these updated parameters are applied to operations of the
balance.
Example of settings and menu structure (Example: the GXM
-
04)
Example : Set to "
Center value using sample weight input
" of "
Input method
" in "
Comparator
".
Caution
The balance may not function properly, depending on the settings and the operating
environment. Be sure to set parameters correctly.
5.1.
Operation Keys and Display of Function Table
The
indicates that the parameter displayed is in effect.
Enters the function table when pressed and held in the weighing mode.
Selects the class or item in the function table.
Changes the parameter.
Moves to an item in the class when a class is displayed.
Stores the new parameter and displays the next class when an item is displayed.
Cancels new parameter and displays the next class when an item is displayed.
Exits the function table and returns to the weighing mode when a class is displayed.
1
Press
with the GX
-
M
/
GF
-
M series and
with the GX
-
L
/
GF
-
L series.
Start from weighing mode
Class
Finish to weighing mode
Press and hold
Press
Item
Input method
Press
Parameter
Center value using numerical input
Parameter
Center value using sample weight
1