EN
12
Display
Error code
Description
EC, E00
Communications error
A protocol error occurred in communications.
Confirm the format, baud rate and parity.
EC, E01
Undefined command error
An undefined command was received.
Confirm the command.
EC, E02
Not ready
A received command cannot be processed.
Example: The balance received a "Q" command, but not in the
weighing mode.
The balance received a "Q" command while
processing a RE-ZERO command.
Adjust the delay time to transmit a command.
EC, E03
Timeout error
If the timeout parameter is set to
, the balance did
not receive the next character of a command within the time limit
of one second. Confirm the communication.
EC, E04
Excess characters error
The balance received excessive characters in a command.
Confirm the command.
EC, E06
Format error
A command includes incorrect data.
Example: The data is numerically incorrect.
Confirm the command.
EC, E07
Parameter setting error
The received data exceeds the range that the balance can
accept. Confirm the parameter range of the command.
4-5 Other Display
When this indicator (
) blinks, automatic self calibration is waiting. If the
balance is not used for several minutes with this indicator blinking, the
balance automatically performs calibration using the internal mass. The
blinking period depends on the operating environment.
Advice
The balance can be used when this indicator is blinking, but we recommend
that you perform calibration before weighing.
Содержание GX-AE Series
Страница 60: ...CT 60 GX AE GX A GF A 0 0001g GX A GF A 0 001g GX A GF A GX AE GX AE GX AE GX A GF A 0 01g 0 1g...
Страница 61: ...CT 61 2 1 20 2 45 60 RH 30 7 Calibration...
Страница 65: ...CT 65 4 4 7 1 During Use CAL CAL CAL...
Страница 67: ...CT 67 Q RE ZERO Q 1 4 5...
Страница 69: ...CT 69 GX 6001A GF 6001A GX 10001A GF 10001A 6200g 10200g 6208 4g 10208 4g 0 1g 0 1g 0 1g 0 5g 5000g...
Страница 70: ...MEMO...
Страница 71: ......