GX-A/GF-A Maintenance manual
Page 43
3. Internal mass value correction
(
Cal 5et
)
(Only GX-A)
The following explains about how the internal mass value is adjusted precisely.
Perform the linearity adjustment before internal mass value correction. In the case of
performing the internal mass value correction only, the internal temperature of the balance
must be approached to the room (the air) temperature enough.
(1)
Preliminary load
Apply a preliminary load by placing a load with the full scale value and remove it. Perform
this pre-load three times.
(2) External mass calibration (
Cal out
)
Step 1. Display the data adjustment mode
d-d5p
after getting into check mode shown
on page 28,29.
Step 2. Press the SAMPLE key three times.
Calout
is displayed.
Step 3. Press the PRINT key. The motor starts and adjusts the internal mass position
correctly.
Step 4.
Cal 0
is displayed. Check the standard mass for calibration in table below.
Model (GX-A series)
Standard masses
203
200g
303
300g
403
400g
603
600g
1003
1000g
1603
2002
2kg
3002
3kg
4002
4kg
6002
5kg
10002
6001
10001
To change the calibration mass value, press the SAMPLE key and change the value
as necessary using the following keys.
SAMPLE key: changes the blinking digit position.
RE-ZERO key : changes the value of the blinking digit.
PRINT key:
saves the data.
Step 5. With nothing placed on the pan, press the PRINT key.
<
Cal 0
is displayed.
Step 6. After it stabilized,
200
is displayed.
(Example of the GX-203A)