26
7. Response Adjustment / Self Check Function
This function stabilizes the weight value, reducing the influence on weighing that is caused by drafts and/or
vibration at the place where the balance is installed. This function adjusts by automatically analyzing the
environment or by hand-operation. The function has three stages as follows : Changing the weighing speed
changes the display refresh rate.
Display
Function
setting
Response characteristic
FAST
Cond 0
Fast response,
Sensitive value
Slow response, Stable value
MID.
Cond 1
SLOW
Cond 2
7-1
Response Adjustment
Response adjustment can be changed by the following method.
1. Press and hold the MODE key until RESPONSE is displayed, and
then release the key.
2. Press the MODE key to select a weighing speed.
Either FAST , MID or SLOW can be selected.
3. After a few seconds of inactivity the balance displays
end
. Then, it
returns to the weighing mode and displays the updated response
indicator.
The response indicator remains displayed for a while.
Note
The value set for response adjustment can be changed by
"Condition
Cond
" of "Environment display" in the Function
Table.
For the setting method, refer to "10.Function Table".
Response
indicator
Press and Hold
Release and
press again
Each pressing switches
the indicators
After a while