-15-
●
Selecting a way to set the density of a liquid
Select the liquid density input method from the function table below. The function table
is available only when the density mode is selected.
Operation
(1) Press and hold the SAMPLE key to display
ba5fnc
.
(2) Press the SAMPLE key several times to display
d5 fnc
.
(3) Press the PRINT key to display
ldin
.
(4) Press the RE-ZERO key to select
ldin 0
or
ldin 1 .
(5) Press the PRINT key to store. The balance displays
end
d
and then displays
Unit
.
(6) Press the CAL key to enter the density mode.
Class Item
Parameter
Description
0
Water temperature
d5 fnc
Density function
ld in
Liquid density input
1
Liquid density
0
= Factory setting
Note :
“
d5 fnc
”
appears after
“
5if
”.
Setting the density of a liquid
1. Press the MODE key as necessary to select the density mode. When the processing
indicator blinks with the unit “g” displayed, it indicates that the density mode is
selected.
2. In the density mode, press and hold the MODE key to enter the mode to set the liquid
density.
Note
In the normal weighing mode, the same procedure will activate the automatic
response adjustment. This function is not available in the density mode.