35
Operation
1 Press
the
ON:OFF
key to turn off the display.
2 While pressing and holding the
key and the
SAMPLE
key, press the
ON:OFF
key. The balance displays
p5
.
3 Press
the
key. Then the balance displays the function switches.
4 Set the switches using the following keys.
SAMPLE
key
To select the switch to change the parameter.
RE-ZERO
key
To change the parameter of the switch selected.
0
: To inhibit changes.
1
: To permit changes
key
To store the new parameter and return to the weighing mode.
CAL
key
To cancel the operation (
Clr
is displayed.) To return to the weighing
mode, press the
CAL
key once again.
8-2 Initializing the Balance
This function returns the following parameters to factory settings.
Calibration
data
Function
table
The sample unit mass value (counting mode), 100%
reference mass value (percent mode)
External calibration weight
Function switch settings
Statistical
data
Note
Be sure to calibrate the balance after initialization.
Operation
1 Press the
ON:OFF
key to turn off the display.
2 While pressing and holding the
key and the
SAMPLE
key, press the
ON:OFF
key.
The balance displays
p5
.
3 Press
the
SAMPLE
key to display
Clr
.
4 Press
the
key.
To cancel this operation, press the
CAL
key.
5 Press
the
RE-ZERO
key.
6 Press
the
key to initialize the balance.
The balance will automatically return to the weighing mode.
With these keys
held down
Press the key