background image

 22 

5-4-2 Using The Statistical Calculation Mode 

Entering data for statistical calculation 

 

Use the following keys to operate the statistical calculation mode. 

 

MODE

 

 key............

 

When the data is entered, moves between the displaying items (weighing mode, 
statistical results and data operation) each time the key is pressed. 

 

 

When no data has been entered, selects the unit. 

 

SAMPLE

 

 key........

 

Turns the minimum display on or off, in the weighing mode. 

 

RE-ZERO

 

 key ......

 

Sets the display to zero in the weighing mode. 

 

PRINT

 

 key............

 

Outputs the data number and the weight data and includes the weight data to 
statistical calculation in the weighing mode. (Output is not in the data format 
specified in the function table because of the data number added.) 

 

 

Outputs the statistical results while the statistical results are displayed. (Output is 
not in the data format specified in the function table.) 

 

CAL

 

 key................

 

Returns to the weighing mode.   

 

 
1 Press 

the 

 

RE-ZERO

 

  key to set the display to zero. 

 

2  Place the sample on the weighing pan and wait for the 

stabilization indicator to turn on. 

 

3 Press 

the 

 

PRINT

 

  key to add the data displayed to statistical 

calculation. The number of data on the upper left of the display 
increases by 1. 

 
4  Repeat steps 1 to 3 for each weighing. 

 
 
 

 
 

Содержание FX-1200iWP

Страница 1: ...FZ 120iWP FZ 200iWP FZ 300iWP FZ 1200iWP FZ 2000iWP FZ 3000iWP FX 120iWP FX 200iWP FX 300iWP FX 1200iWP FX 2000iWP FX 3000iWP 1WMPD4001591A ...

Страница 2: ... or translated into any language in any form by any means without the written permission of A D Company Ltd The contents of this manual and the specifications of the instrument covered by this manual are subject to change for improvement without notice Windows Word and Excel are the registered trademarks of the Microsoft Corporation ...

Страница 3: ... Counting Mode PCS 17 5 3 Percent Mode 19 5 4 Statistical Calculation Mode 20 5 5 Statistical Calculation Mode Example of Use 25 Adapting to the Environment 6 RESPONSE ADJUSTMENT 27 7 CALIBRATION 28 7 1 Calibration Mode 28 7 2 Calibration Using the Internal mass One Touch Calibration only for FZ iWP series 29 7 3 Calibration Using an External Weight 30 7 4 Calibration Test Using an External Weight...

Страница 4: ... UNDERHOOK 53 12 PROGRAMMABLE UNIT 54 RS 232C Serial Interface 13 RS 232C SERIAL INTERFACE 55 14 CONNECTION TO PERIPHERAL EQUIPMENT 56 14 1 Connection to the AD 8121B Printer 56 14 2 Connection to a Computer 57 14 3 Using Windows Communication Tools WinCT 58 15 COMMANDS 59 15 1 Command List 59 15 2 Acknowledge Code and Error Codes 60 15 3 Settings Related to the RS 232C 61 Maintenance 16 MAINTENAN...

Страница 5: ...pose balance can be installed almost anywhere Casing construction strong protection against dust and water Stabilization time of one second When FAST is selected for the response rate a stabilization time of one second to read a displayed value after a sample is placed on the pan has been achieved Multiple weighing units with most of the common units used around the world Standard RS 232C serial i...

Страница 6: ... interference and under these circumstances the user would be required to take at his own expense whatever measures are necessary to eliminate the interference FCC Federal Communications Commission in the U S A Compliance with EMC Directives This device features radio interference suppression and safety regulation in compliance with the following Council Directives Council directive 89 336 EEC EN6...

Страница 7: ...5 ...

Страница 8: ...6 ...

Страница 9: ...cable The breeze break is an accessory for FZ i WP series FX 120i WP FX 200i WP FX 300i WP The breeze break Assemble the breeze break as shown below Follow the numbered sequence 4 1 2 3 Waterproof diaphragm Pan boss Pan support Weighing pan Bubble spirit level Display Keys Leveling foot AC adapter AC adapter ID label Note Please confirm that the AC adapter type is correct for your local voltage an...

Страница 10: ... 2000i WP FX 3000i WP Assemble the pan support and weighing pan on the balance as shown in the illustration on page 7 3 Adjust the leveling feet to level the balance Confirm it using the bubble spirit level 4 Confirm that the AC adapter type is correct for the local voltage and power receptacle type 5 Connect the AC adapter to the balance Warm up the balance for at least 30 minutes with nothing on...

Страница 11: ...ty are not excessive The best operating temperature is about 20 C 68 F at about 50 relative humidity Install the balance where it is not exposed to direct sunlight and it is not affected by heaters or air conditioners Install the balance where it is free of dust Install the balance away from equipment which produces magnetic fields Install the balance in a stable place avoiding vibration and shock...

Страница 12: ... ambient humidity above 45 RH Use a metal shield case for a charged sample Wipe charged plastic samples with a moistened cloth This balance uses a strong magnet as part of the balance assembly so please use caution when weighing magnetic materials such as iron If there is a problem use the underhook on the bottom of the balance to suspend the material away from the influence of the magnet Eliminat...

Страница 13: ... spray 3 3 After Use Do not allow the balance to be immersed in water Even though the balance complies with IP code the balance will not withstand being completely immersed in water Avoid mechanical shock to the balance Do not disassemble the balance Contact the local A D dealer if the balance needs service or repair Do not use organic solvents to clean the balance Clean the balance with a lint fr...

Страница 14: ...on table mode Refer to 9 FUNCTION TABLE Switches the weighing units stored in the function table Refer to 4 WEIGHING UNITS Enters the response adjustment mode Cancels the operation when performing settings Enters the calibration mode Outputs the weight data to a printer or personal computer using the RS 232C serial interface depending on the function table settings Confirms the operation when perf...

Страница 15: ...mode for weighing press the MODE key For details about the units and modes see the table below Name unit mode Display Function table Storing mode Conversion factor 1 g Gram g g 1 g Counting mode pcs pcs Percent mode Ounce Avoir oz oz 28 349523125 g Pound lb lb 453 59237 g Pound Ounce Loz LO 1 Lb 16 oz 1 oz 28 349523125 g Troy Ounce ozt ozt 31 1034768 g Metric Carat ct ct 0 2 g Momme 3 75 g Pennywe...

Страница 16: ...el Taiwan 3 25330 5 86670 8 53330 0 00005 Tael China 3 90400 7 04000 10 24000 0 00005 Tola India 10 4597 18 8618 27 4353 0 0001 Messghal 26 0265 46 9335 68 2665 0 0005 Capacity Unit FX 1200i WP FX 2000i WP FX 3000i WP Minimum display Gram 1220 00 2200 00 3200 00 0 01 Ounce Avoir 43 0340 77 6025 112 8765 0 0005 Pound 2 68965 4 85015 7 05480 0 00005 Pound Ounce 2L 11 03oz 4L 13 60oz 7L 0 88oz 1L 0 0...

Страница 17: ...Specify a unit or mode in the order to be displayed using the following keys SAMPLE key To sequentially display the units RE ZERO key To specify a unit or mode The stabilization indicator appears when the displayed unit or mode is specified 5 Press the PRINT key to store the units or modes The balance displays end and then displays the next menu of the function table 6 Press the CAL key to exit th...

Страница 18: ...e While the stabilization indicator is on pressing the PRINT key will output the weight value using the RS 232C serial interface Note Peripheral equipment that is sold separately such as a printer or a personal computer is required 4 Remove the sample and container from the pan Notes To use other units press the MODE key and select an appropriate unit Press the SAMPLE key to turn on or off the min...

Страница 19: ... A greater number of samples will yield more accurate counting result 4 Place a container on the weighing pan if necessary Press the RE ZERO key to cancel the weight tare The number specified in step 3 appears e g 25 0 is displayed if 25 is selected in step 3 5 Place the number of samples specified on the pan In this example 25 pieces 6 Wait for the stabilization indicator to come on Press the PRI...

Страница 20: ... unit mass automatically by increasing the number of samples and averaging the unit mass variable to minimize the weighing error as the counting process proceeds 8 If a few more samples are added the processing indicator turns on To prevent an error add three or more The processing indicator does not turn on if overloaded Try to add the same number of samples as displayed 9 The balance re calculat...

Страница 21: ...e mass on the pan or in the container 5 Press the PRINT key to store the reference mass The balance displays 100 00 The decimal point position depends on the reference value The reference mass stored even if the AC adapter is removed is maintained in non volatile memory Note If the balance judges that the mass of the sample is too light to be used as a reference it displays lo Do not use the sampl...

Страница 22: ...er the input Turning the balance off will delete the statistical data The standard deviation and coefficient of variation are obtained by the equation below Standard deviation where Xi is the i th weight data N is number of data Standard deviation Coefficient of variation CV x 100 Average 5 4 1 Getting Started Switching to the Statistical Function Mode Changing The Function Table 1 Press and hold ...

Страница 23: ...fficient of variation 7 Press the PRINT key to store the setting Note The statistical calculation mode will be enabled using the ON OFF key or when the power is turned on 8 Press the CAL key to return to the weighing mode Selecting the unit 9 Press the MODE key to select the unit to be used for the statistical calculation mode Notes Selecting the unit using the MODE key is not available after the ...

Страница 24: ...ta number and the weight data and includes the weight data to statistical calculation in the weighing mode Output is not in the data format specified in the function table because of the data number added Outputs the statistical results while the statistical results are displayed Output is not in the data format specified in the function table CAL key Returns to the weighing mode 1 Press the RE ZE...

Страница 25: ...n and coefficient of variation are displayed as When the average is 0 the coefficient of variation is displayed as Statistical items are indicated on the upper left of the display using the following symbols Symbol Statistical item SUM Sum MAX Maximum MIN Minimum R Range Maximum minimum AVE Average SD Standard deviation CV Coefficient of variation 6 While the results are displayed press the PRINT ...

Страница 26: ...ete the latest data and exclude it from statistical calculation The number of data decreases by 1 when the balance returns to the weighing mode Clearing the statistical data All the statistical data will be deleted and the number of data will be 0 zero 1 In the weighing mode press the MODE key and then press the SAMPLE key several times to display Clear 2 Press the PRINT key to display Clr no 3 Pr...

Страница 27: ...en release the key 2 Select the application function Press the SAMPLE key several times to display ap fnc Then press the PRINT key to display apf 0 3 Change the application function parameter to 2 Press the RE ZERO key to display apf 2 Press the PRINT key to confirm the change After end ba5fnc is displayed Enabling Zero after output 4 Select Zero after output Press the SAMPLE key several times to ...

Страница 28: ...s 0 00 g Storing the weight value of formula 1 The weight value data is output when the peripheral output equipment is connected 4 Weigh formula 2 and press the PRINT key The balance displays 0 00 g Storing the weight value of formula 2 The weight value data is output when the peripheral output equipment is connected 5 When there are some more formulae to be added repeat step 4 6 After mixing is c...

Страница 29: ...NSE is displayed And then release the key 2 Press the MODE key to select a rate of the response adjustment Either FAST MID or SLOW can be selected 3 After a few seconds of inactivity or when the PRINT key is pressed the balance displays eend Then it returns to the weighing mode and displays the updated response indicator The response indicator remains displayed for a while Note Setting the respons...

Страница 30: ...ration or drafts to affect the balance during calibration To output the data for GLP using the RS 232C serial interface set GLP output info of Data output dout For details refer to 9 FUNCTION TABLE Calibration test is available only when GLP output info of Data output dout is set to 1 or 2 Caution on using an external weight The accuracy of an external weight will influence the accuracy of weighin...

Страница 31: ...on If the GLP output info parameter of the function table is set to 1 or 2 the balance displays glp and outputs Calibration Report using the RS 232C interface or stores the data in memory For details on the calibration report format refer to 10 2 GLP Report 5 The balance will automatically return to the weighing mode after calibration About the internal mass The mass of the internal mass may chang...

Страница 32: ...libration weight value stored in the balance proceed to step 5 4 Specify the calibration weight value as follows SAMPLE key To switch the display condition to All of the segments blinking calibration weight selection mode or The last two digits blinking value adjustment mode RE ZERO key To increase the value MODE key To decrease the value To select the calibration weight or adjust the value PRINT ...

Страница 33: ...eight Do not allow vibration or drafts to affect the balance 7 The balance displays end Remove the weight from the pan 8 When the GLP output is set the balance displays glp and outputs Calibration Report 9 The balance will automatically return to the weighing mode 10 Place the calibration weight on the pan and confirm that the value displayed is within 2 digits of the specified value If it is not ...

Страница 34: ...you want to change the target weight a list of usable weights is shown on page 28 press the SAMPLE key and proceed to step 4 If you use the target weight value stored in the balance proceed to step 5 4 Specify the target weight value as follows SAMPLE key To switch the display condition to All of the segments blinking target weight selection mode or The last two digits blinking value adjustment mo...

Страница 35: ...ce the displayed target weight on the pan and press the PRINT key The balance measures the target weight and displays the measured value Do not allow vibration or drafts to affect the balance 7 The balance displays end Remove the weight from the pan 8 When the GLP output is set the balance displays glp1 and outputs Calibration Test Report 9 The balance will automatically return to the weighing mod...

Страница 36: ...ches FX iWP series Function table 0 To inhibit changes to the function table 1 To permit changes to the function table Calibration using the internal mass One touch calibration 0 To inhibit calibration using the internal mass 1 To permit calibration using the internal mass Calibration using an external weight 0 To inhibit calibration using an external weight 1 To permit calibration using an extern...

Страница 37: ...he CAL key once again 8 2 Initializing the Balance This function returns the following parameters to factory settings Calibration data Function table The sample unit mass value counting mode 100 reference mass value percent mode External calibration weight Function switch settings Statistical data Note Be sure to calibrate the balance after initialization Operation 1 Press the ON OFF key to turn o...

Страница 38: ...mbol indicates that the parameter displayed is in effect When pressed and held in the weighing mode enters the function table mode Selects the class or item in the function table mode Changes the parameter When a class is displayed moves to an item in the class When an item is displayed stores the new parameter and displays the next class When an item is displayed cancels the new parameter and dis...

Страница 39: ...mma Decimal point format 0 OFF p on Auto display ON 1 ON Turns on the weighing mode display when the AC adapter is connected 0 OFF poff Auto display OFF 1 ON 10 minutes Turns off the display after 10 minutes of inactivity 0 Displays rng Display at start 1 Does not display Select whether or not to display the minimum weighing value at weighing start 0 Does not sound ba5fnc Environment Display beep ...

Страница 40: ...ference 1 Minus only Displayed value Reference ap p Auto print polarity 2 Both Regardless of displayed value 0 10 digits 1 100 digits ap b Auto print difference 2 1000 digits Difference between reference value and displayed value 0 Every measurement 1 2 seconds 2 5 seconds 3 10 seconds 4 30 seconds 5 1 minute 6 2 minute 7 5 minute int Interval time 8 10 minute Interval time for the interval output...

Страница 41: ...mable unit Multi unit Sets an arbitrary coefficient Available only when programmable unit mode is selected Unit Unit Refer to 4 WEIGHING UNITS id ID number Refer to 10 ID NUMBER AND GLP REPORT 0 Normal weighing mode 1 Capacity indicator apf Application function 2 Statistical calculation mode 0 Number of data sum 1 Number of data sum maximum minimum average range maximum minimum ap fnc Application ...

Страница 42: ...nt a weight value from drifting due to vibration or drafts Note With Hold function Hold set to ON 1 this item is used to set the stabilization range Hold function Hold Animal weighing mode This function is used to weigh a moving object such as an animal When the weight data is over the weighing range from zero and the display fluctuation is within the stabilization range for a fixed period of aver...

Страница 43: ...trong zero tracking Display refresh rate 5pd Period to refresh the display This parameter influences Stream mode along with Baud rate and Data output pause Note This item is selected automatically when the response rate is changed Decimal point pnt The decimal point format can be selected Auto display ON p on When the AC adapter is connected the display is automatically turned on in the weighing m...

Страница 44: ... added with ar d set to 1 to adjust zero after the data is output Mode B Required setting dout prt 2 Auto print mode B reference laststablevalue dout ap p Auto print polarity dout ap b Auto print difference Example For outputting the weight value while a sample is added Stream mode The balance outputs the weight data continuously regardless of the display condition The display does not blink in th...

Страница 45: ...mat If an AD 8121B is used set the printer to MODE 1 or 2 This format consists of fifteen characters excluding the terminator A header of two characters indicates the balance condition The polarity sign is placed before the data with the leading zeros If the data is zero the plus sign is used The unit consisting of three characters follows the data DP Dump print format 5if type 1 This format is us...

Страница 46: ... format 5if type 3 A header of two characters indicates the balance condition The polarity sign is used only for negative data The weight data uses spaces in place of the leading zeros The character length of this format changes dependent upon the unit NU numerical format 5if type 4 This format outputs only numerical data This format consists of nine characters excluding the terminator The polarit...

Страница 47: ...45 9 7 Data Format Examples ...

Страница 48: ...apacity indicator apf 1 In the weighing mode the indicator displays the weight data relative to the weighing capacity in percentage Zero 0 maximum capacity 100 Statistical calculation mode apf 2 The mode statistically calculates the weight data and displays or outputs the results For details refer to 5 4 Statistical Calculation Mode ...

Страница 49: ...d mode Entering the upper limit value 6 With Cp Hi displayed press the PRINT key The current setting of the upper limit value is displayed with all the digits blinking When the current setting is not to be changed press the PRINT or CAL key to proceed to step 7 When the current setting is to be changed press the RE ZERO key Change the setting using the following keys SAMPLE key To select the digit...

Страница 50: ...e digits blinking 5 Set the Year Month and Date using the following keys SAMPLE key To select the digits to change the value The selected digits blink RE ZERO key To increase the value by one MODE key To decrease the value by one PRINT key To store the new setting display end and go to step 6 CAL key To cancel the new setting and go to step 6 Confirming the time 6 The current time is displayed wit...

Страница 51: ...tes Do not enter invalid values such as a non existing date when setting the time and date When the clock backup battery has been depleted the balance displays rtc pf Under this condition press any key and set the time and date The dead battery only affects the clock and calendar function Even so the function works normally as long as the AC adapter is connected to the balance ...

Страница 52: ...lock and End block for the weight data 10 1 Setting the ID Number 1 Press and hold the SAMPLE key until ba5fnc of the function table is displayed 2 Press the SAMPLE key several times to display id 3 Press the PRINT key Set the ID number using the following keys SAMPLE key To select the digit to change the value RE ZERO key To set the character of the digit selected Refer to the display character s...

Страница 53: ...D ABCDEFG DATE 2006 02 06 TIME 12 23 34 CALIBRATED EXT CAL WEIGHT 2000 00 g SIGNATURE TERM Space ASCII 20h Terminator or Carriage return ASCII 0Dh Line feed ASCII 0Ah CR CR LF CR LF AD 8121 format General data format Manufacturer Model Serial number ID number Date Time Calibration test type Zero point value Target weight value Target weight Signature A D TERM MODEL FX 3000i TERM S N 01234567 TERM ...

Страница 54: ...en the setting is info 2 Weight data Time Signature End block AD 8121 format Manufacturer Model Serial number ID number Date Time Title block General data format A D TERM MODEL FX 3000i TERM S N 01234567 TERM ID ABCDEFG TERM DATE TERM TERM START TERM TIME TERM TERM WT 1234 56 g TERM WT 1234 61 g TERM WT 1234 62 g TERM WT 1234 63 g TERM WT 1234 51 g TERM WT 1234 53 g TERM WT 1234 71 g TERM WT 1234 ...

Страница 55: ... as shown below Caution Do not apply excessive force to the underhook When the underhook is not in use attach the plastic cap to prevent dust from getting into the balance The cap was sealed to the bottom of the balance using a special adhesives to be dust tight and protected against water jets The balance does not comply with IP65 once the cap has been removed Underhook Cap Bottom of the balance ...

Страница 56: ...t the coefficient Confirming the coefficient 4 The current coefficient is displayed with the first digit blinking When it is not to be changed press the CAL key and proceed to step 6 When it is to be changed press the RE ZERO key and proceed to step 5 Setting the coefficient 5 Set the coefficient using the following keys SAMPLE key To select a digit to change the value The selected digit blinks RE...

Страница 57: ...ata bits 7 bits None Data bits 8 bits Stop bit 1 bit Code ASCII D Sub 9 pin assignments Pin No Signal name Direction Description 1 No connection 2 TXD Output Transmit data 3 RXD Input Receive data 4 No connection 5 SG Signal ground 6 DSR Output Data set ready 7 RTS Input Request to send 8 CTS Output Clear to send 9 No connection Signal names of the balance side are the same as the DTE side with TX...

Страница 58: ... bp5 Baud rate 2 2 2 2 btpr Data bit parity bit 0 0 0 0 Crlf Terminator 0 0 0 0 5if Serial interface type Data format 0 0 0 1 1 Set appropriate parameters for ap p Auto print polarity and ap b Auto print difference when auto print mode A or B prt 1 or 2 is selected Set the AD 8121B DIP switch No 3 to ON when unstable data is printed with prt 4 2 Set 1 when multiple lines are printed in the FX iWP ...

Страница 59: ...mit the balance data to a personal computer The FX iWP series balance is a DCE Use a straight through cable If purchasing the RS 232C cable on the market check the interface connections and type Use the waterproof RS 232C cable AX KO2737 500 that is sold separately as an option when the RS 232C serial interface is used with IP 65 Refer to page 67 D Sub 9 pin Female connector Inch screw thread ...

Страница 60: ...rosoft Excel Can be used with most application software Using the WinCT software the balance can do the following 1 Analyzing the weight data and the statistics with RsKey The weight data can be input directly into an Excel worksheet Then Excel can analyze the data to obtain sum average standard deviation maximum and minimum value and display them in a graph 2 Controlling the balance using command...

Страница 61: ...hen stabilized Note The Q and SI commands the S and E SCP commands behave the same Commands to control the balance CAL Same as the CAL key OFF Turns the display off ON Turns the display on P Same as the ON OFF key PRT Same as the PRINT key R Same as the RE ZERO key SMP Same as the SAMPLE key T Tare key Z Same as the RE ZERO key E SCT Same as the RE ZERO key U Same as the MODE key ID Requests the I...

Страница 62: ...command to control the balance and can not process it the balance transmits an error code EC Exx When the balance receives a command to control the balance and can process it the balance transmits the acknowledge code Among commands to control the balance the following transmit the acknowledge code both when the balance receives the command and when the balance has accomplished the command If the ...

Страница 63: ...noise or a parity error has occurred due to transmission error the balance transmits an error code In this case send the command again 15 3 Settings Related to the RS 232C Concerning the RS 232C the balance has two functions Data output dout and Serial interface 5if Set each function as necessary ...

Страница 64: ... D dealer if the balance needs service or repair Use the original packing material for transportation While cleaning the balance and keeping it waterproof cover the RS 232C interface connector using the terminal cover or connect the waterproof RS 232C cable AX KO2737 500 and cover the AC adapter jack Also insure that the underhook cap is in place If water accumulates on the waterproof diaphragm a ...

Страница 65: ...lance level Refer to 3 1 Before Use Is the operating environment free from vibration and drafts For the FZ 120i WP FZ 200i WP FZ 300i WP FX 120i WP FX 200i WP FX 300i WP has the breeze break been installed Is there a strong electrical or magnetic noise source such as a motor near the balance Weighing method Is the weighing pan installed correctly Is the waterproof diaphragm deformed Is the waterpr...

Страница 66: ...y to return to the weighing mode EC E21 Calibration weight error The calibration weight is too light Confirm that the weighing pan is properly installed Confirm the calibration weight value Press the CAL key to return to the weighing mode Overload error A sample beyond the balance weighing capacity has been placed on the pan Remove the sample from the pan Weighing pan Error The weight value is too...

Страница 67: ... one second Confirm the communication EC E04 Excess characters error The balance received excessive characters in a command Confirm the command EC E06 Format error A command includes incorrect data e g The data is numerically incorrect Confirm the command EC E07 Parameter setting error The received data exceeds the range that the balance can accept Confirm the parameter range of the command Other ...

Страница 68: ... as Microsoft Excel Word and memo pad Driver installation is not necessary FXi 08 Ethernet interface Used to connect the balance to a LAN The WinCT Plus data communication software is provided as an accessory and can perform the following Acquire data from multiple balances connected to a LAN Control these balances with commands Acquire data transmitted from balances Example When pressing the PRIN...

Страница 69: ... more height AX USB 9P EX USB converter Adds a COM port to a PC Enables bi directional communication between the PC and the balance when a USB driver is installed Can use serial communication software such as WinCT on a PC without COM ports AX KO2737 500 Waterproof RS 232C cable Length 5 m D Sub 9 pin D Sub 9 pin Only the 9 pin of the balance side is of waterproof type Note Be sure to insert the w...

Страница 70: ... FX i WP series balance using the RS 232C serial interface to display the weight data away from the balance AD 8922 Remote controller Connected to the FX i WP series balance using the RS 232C serial interface to display the weight data and to remotely control the balance Analog output and comparator output available as an option to be installed AD 8121B Printer Compact dot matrix printer Statistic...

Страница 71: ... pieces Minimum 100 reference mass 0 100 g 1 00 g Percent mode Minimum 100 display 0 01 0 1 1 Depends on the reference mass stored External calibration weight 100 g 50 g 200 g 100 g 300 g 200 g 100 g 1000 g 500 g 2000 g 1000 g 3000 g 2000 g 1000 g Weighing pan diameter 130 mm 150 mm Net weight Approx 3 2 kg External dimensions 193 W x 262 5 D x 84 5 H mm AC adapter Confirm that the adapter type is...

Страница 72: ... rate 5 times second 10 times second or 20 times second Minimum unit mass 0 001 g 0 01 g Counting mode Number of samples 5 10 25 50 or 100 pieces Minimum 100 reference mass 0 100 g 1 00 g Percent mode Minimum 100 display 0 01 0 1 1 Depends on the reference mass stored External calibration weight 100 g 50 g 200 g 100 g 300 g 200 g 100 g 1000 g 500 g 2000 g 1000 g 3000 g 2000 g 1000 g Weighing pan d...

Страница 73: ...weighing pan up to the cap of the breeze break Inside dimension 5 Height of the opening when a clear plate is removed 6 Underhook position FZ 1200i WP FZ 2000i WP FZ 3000i WP FX 1200i WP FX 2000i WP FX 3000i WP 6 Underhook position Unit mm 184 171 5 2 85 5 85 4 82 83 5 5 176 115 1 φ130 3 162 6 3 184 168 5 171 5 2 262 5 193 115 1 71 5 61 5 194 5 192 6 193 71 5 61 5 168 5 85 5 162 φ150 262 5 194 5 6...

Страница 74: ...72 FXi WP 11 Large breeze break Unit mm 199 5 112 Internal size above the pan Opening 294 6 156 When fully opened Opening 270 105 31 198 107 When fully opened When fully opened When fully opened ...

Страница 75: ...erational function Re zero To set the display to zero GLP Good Laboratory Practice Repeatability Variation in measured values obtained when the same weight is placed and removed repetitively Usually expressed as a standard deviation e g Standard deviation 1 digit This means that measured values fall within 1 digit in the frequency of about 68 Stabilization time Time required after a sample being p...

Страница 76: ...larity 38 AVE Average 23 Average 20 23 39 AX KO2737 500 3 9 57 62 67 AX USB 9P 67 B ba5fnc Environment display 15 37 Baud rate 39 55 56 beep Beep 37 bep HI buzzer 37 bep LO buzzer 37 bep OK buzzer 37 bp5 Baud rate 39 56 Breeze break 7 67 btpr Data bit parity bit 39 56 Built in battery unit 66 C Cal e Calibration weight error 64 Cal e Calibration weight error 64 CAL key 12 32 Calibration 3 12 28 30...

Страница 77: ...8 Inhibit 34 Initializing the Balance 35 int Interval time 38 Interval output mode 38 43 Interval output mode active indicator 12 Interval output mode standby indicator 12 Interval time 38 IP65 3 9 53 66 69 70 73 K Key mode 38 42 Key mode B 38 42 Key mode C 38 42 KF format 39 44 L Leveling foot 7 Linearity 69 70 lo 64 Lower limit value 47 M Main unit cover 7 MAX Maximum 23 Maximum 20 23 39 Maximum...

Страница 78: ...ndicator 12 15 Stabilization time 69 70 Standard deviation 20 23 39 Standby indicator 12 5taf Statistical function mode output items 21 39 Statistical calculation mode 3 20 21 39 46 Statistical function mode output items 39 5t b Stability band width 37 40 Stop bit 55 Stream mode 38 42 SUM Sum 20 23 39 T Tare 16 17 19 26 Terminal cover 7 9 62 Terminator 39 Timeout 39 Title block 52 Transmission for...

Страница 79: ......

Страница 80: ...9 0 4102 459231 A D MERCURY PTY LTD 32 Dew Street Thebarton South Australia 5031 AUSTRALIA Telephone 61 8 8301 8100 Fax 61 8 8352 7409 A D KOREA Limited 한국에이 엔 디 주 대한민국 서울시 영등포구 여의도동 36 2 맨하탄 빌딩 8층 8th Floor Manhattan Bldg 36 2 Yoido dong Youngdeungpo ku Seoul KOREA 전화 82 2 780 4101 팩스 82 2 782 4280 A D RUS CO LTD Компания ЭЙ энд ДИ РУС 121357 Российская Федерация г Москва ул Верейская 112 Квартал...

Отзывы: