9. Function Switch And Initialization
9.1.
Permit Or Inhibit
The balance stores parameters that must not be changed unintentionally.
Example:
Calibration
data for accurate weighing, Data for adapting to the operating environment,
Control data for the RS-232C interface.
The balance is equipped with five switches for the purpose of protecting parameters. Each switch
can select either "permit" or "inhibit". "Inhibit" protects parameters against unintentional operations.
1 Press
the
ON:OFF
key to turn off the display.
2 While pressing and holding the
key and the
RANGE
key,
press the
ON:OFF
key to display
p5
.
3 Press
the
key. Then the balance displays the function switches.
4 Specify the switches using the following keys.
RANGE
key........... The key to select blinking digit.
RE-ZERO
key........ The key to change the parameter for the selected switch.
0
To inhibit changes. (Can not be used.)
1
To
permit
changes. (Can be used.)
key ............ The key to store the new parameter and return to the weighing mode.
CAL
key................ The key to cancel the operation (
Clr
is displayed ). To return to the
weighing mode, press the
CAL
key once again.
Function table
0
To inhibit changes to the function table.
1
To
permit
changes to the function table.
Calibration using the internal mass (One-touch calibration)
0
To inhibit calibration using the internal mass.
1
To permit calibration using the internal mass.
Calibration using the external weight
0
To inhibit calibration using the external weight.
1
To permit calibration using the external weight.
Automatic self calibration
(for variation of ambient temperature)
0
To inhibit automatic self calibration.
1
To permit automatic self calibration.
Internal mass correction
0
To inhibit correction.
1
To
permit
correction.