8.7.
Correcting The Internal Mass Value
The balance can correct the internal mass
value within the range shown below. This
function corrects the internal mass value to
conform to an external weight. The corrected
mass value is maintained in non-volatile
memory even if the AC adapter is removed.
The internal mass value is corrected as follows:
Model
The internal mass
The normal range
BM-20
,
BM-22
20.00000 g
±
0.15 mg
BM-252
,
BM-200
,
BM-300
,
BM-500
200.0000 g
±
1.5 mg
Example
: 200.0000 g is corrected to +0.6 mg (200.0006 g). When using a 100 g external weight by
+0.6 mg for the correction, and the weight changed to 200 g, the correction value is +1.2 mg.
1 Perform one-touch calibration. Weigh an external weight and get the correction value.
2 Press
the
ON:OFF
key to turn off the display.
3 While pressing and holding the
key and the
RANGE
key,
press the
ON:OFF
key.
p5
is displayed.
4 Press
the
key to display the function switches.
Set the function table switch and internal mass correction
switch to "
1
" as shown above using the following keys.
RANGE
key.......The key to select blinking figure.
RE-ZERO
key....The key to change the value of the blinking figure.
key.........The key to store it and return to weighing mode.
CAL
key............The key to cancel current operation.
5 Press and hold the
RANGE
key to enter the function
table and release the key when
ba5fnc
is displayed.
6 Press the
RANGE
key several times until
C5
in
is
displayed, then release the key.
7 Press
the
key.
Correct the internal mass value using the following keys.
RE-ZERO
(
+
)key ....The key to increase the value.
MODE
(
-
)key ........The key to decrease the value.
key...........The key to store the new value and display
the next menu item of the function table.
CAL
key.............The key to cancel this correction and display
the next menu item of the function table.
8 Press
the
CAL
key to return the weighing mode.
9 Press
the
CAL
key to calibrate the balance using the internal mass.
10 Check that the correction has been performed properly with the
external weight. If the value is incorrect, repeat the correction.
Internal mass correction
Press and hold Press
Press and hold
Weighing mode
Weighing mode
Function switch
Function table
Corrected value
Standby
The internal mass correction switch
The function table switch
Perform one-touch calibration
200g
200g
Correct the internal mass
by +0.6mg at 200g.
Calibrate with this
internal mass.
The same mass
200.0000
g
200.0006
g
Corrected external weight