7. CALIBRATION
The AD-4530 measures voltage signals from sensors and displays the
values. Calibration is performed so that it performs correctly.
The decimal point (CF01), minimum scale value (CF02) and rated
capacity (CF03) are set using the function mode.
The zero point input voltage (CF04), the span input voltage (CF05)
and the display value for the span input voltage (CF06) are set using
calibration mode.
Calibration setting by function mode is also available. (Digital
calibration)
* During calibration, maintain a stable environment or a calibration
error may occur.
* You can confirm stabilization by confirming that the HOLD LED is on.
* The decimal point blinks to indicate that a number is being displayed,
not a weighing value.
7.1. Calibration Modes
With the weighing mode, press and to enter calibration
mode.
Enter zero point calibration mode.
Return to the weighing mode.
7.1.1.
Zero point calibration mode
With nothing on the load cell, wait for
it to stabilize, and then press .
Perform zero point calibration, and
proceed to span calibration mode.
Cancel zero point calibration, and
proceed to span calibration mode.
Hold to display the mV/V of the zero
point.
7.1.2.
Span calibration mode
Add the actual load for span calibration to the load
cell, and then input the value to be displayed when
the load is loaded.
After stabilization, press .
Select the digit to be changed.
Increase the value of the digit to be changed.
Change the polarity.
Perform span calibration, and
proceed to storing mode.
Cancel span calibration, and
proceed to storing mode.
* After the span calibration, the AD-4530 displays the mV/V value of the span
calibration for 3 seconds, and then proceeds to the storing mode.
7.1.3. Storing calibration mode
Store the calibration zero, span and displayed
value acquired.
When calibration was not performed, data is not
stored.
Store the data acquired, and return to
weighing mode.
Do not store the data acquired, and
return to weighing mode.
7.2. Calibration Error
Display Cause
Remedy
C?e2
Voltage at zero point
calibration was too far in the
positive direction.
C?e3
Voltage at zero point
calibration was too far in the
negative direction.
Confirm the rating
and connection of the
load cell.
Display Cause
Remedy
C?e4
The mass value exceeds the
weighing capacity.
Use a proper mass.
C?e5
The mass value is less than
the minimum scale value.
Use a proper mass.
C?e6
The load cell sensitivity is
insufficient.
Confirm the load cell
connection. Use a proper
mass.
C?e7
Voltage at span calibration is
less than the zero point.
Confirm the load cell
connection.
C?e8
When adding a load of the
weighing capacity, the load cell
output voltage is too high.
Use a load cell with a
greater weighing capacity
or set a smaller weighing
capacity value.
8. FUNCTION MODE
Use the function mode to set various functions and data. The set values are
saved in non-volatile memory and are maintained even if the power is
switched off.
8.1. Structure of Functions
The first 2 digits of the Function No. are the function group. The last 2 digits of the
Function No. are the function item.
CF Calibration
function
F0 Basic
function
F1 Comparator
function
Use this function to set the comparator operation.
F2
Analog output function
Use this function to set the output value of the analog voltage
output and analog current output.
F3, F4 Serial communication function
Use this function to set the RS-232C and RS-485.
* Set the zero point input voltage (CF04), the span input voltage (CF05) and
the display value for the span input voltage (CF06) in calibration mode.
* Set the upper limit value (F101) and lower limit value (F102) in
comparator mode.
When setting a function, the decimal point blinks to indicate that a
number is being displayed, not a weighing value.
8.2.Key operation
In weighing mode, press and to enter the function selection
mode.
8.2.1. Function selection mode
Select the function group. (First 2 digits)
Select the function item. (Last 2 digits)
Enter the setting changing mode.
Save the setting in non-volatile memory, and then return to
weighing mode.
8.2.2. Setting changing mode (Two methods)
P
Parameter selection method (All digits blinking)
Change the parameter.
Enter the setting and return to function selection mode.
Cancel the setting and return to function selection mode.
D
Digital input method (Change the blinking digit only)
Move the digit to be changed to the right.
Change the value of the blinking digit.
Change the polarity.
Enter the setting and return to function selection mode.
Cancel the setting and return to function selection mode.
8.3. Function Items
8.3.1. Calibration (C function)
Function No.
SETTING range
Function
Description
Default value
Setting type
Cf01
0 to 3
Decimal point
position
Decimal point position of the weighing
value
0
:
0000
1
:
000.0
2
:
00.00
3
:
0.000
0
P
Cf02
1 to 50
Minimum
scale value
Minimum division (d) of the weighing value
1
:1
10
:10
2
:2
20
:20
5
:5
50
:50
1
P
Cf03
1 to 9999
Maximum
weighing
capacity
Weighing is possible up to the value of
this setting plus 8 d (8 minimum
divisions)
Decimal point position depends on
Cf01
.
7000
D
Cf04
-7.000
to 7.000
Input voltage
of zero point
Input voltage from the load cell at zero
point
(Unit: mV/V)
0.000
D
Cf05
0.001
to 9.999
Input voltage
of span
Input voltage from the load cell at span
(measurement point - zero point)
(Unit: mV/V)
7.000
D
Cf06
-9999
to 9999
Display value
for input
voltage of span
Display value for span (measurement
point - zero point)
Decimal point position depends on
Cf01
.
7000
D
Cf07
0 to 100
Zero
adjustment
range
Range to enable zero adjustment by the
ZERO key
Expressed in percent of the weighing
capacity with the calibration zero point
as the center
100
D
Cf08
0.0 to 5.0
Zero tracking
time
Performed in combination with zero
tracking width. (Unit: second)
When
0.0
, zero tracking is not performed.
0.0
D
Cf09
0.0 to 9.9
Zero tracking
width
Performed in combination with zero
tracking time. (Unit: 0.1 d)
When
0.0
, zero tracking is not performed.
0.0
D
Cf10
0 to 2
Power on zero
Digital zero when turning power on
0
: Digital zero off
1
: Perform digital zero again
2
: Use state when the power was turned
off
0
P
Cf11
1 to 2
Zero operation
1
: On with , Off with
+
2
: On or off with
+
1
P
8.3.2.
Basic Functions
Function No.
Setting range
Function
Description
Default value
setting type
f001
0000
to 1111
Disable key
Each digit of the setting corresponds to a key
switch. Only available in the weighing mode.
Key assignment
0
: Enabled
0 0 0 0
1
: Disabled
+
0000
(Binary)
D
f002
0 to 11
Digital filter
Cutoff frequency
0
: Off
6
:0.5 Hz
1
:2.5 Hz
7
:0.35 Hz
2
:2.0 Hz
8
:0.25 Hz
3
:1.5 Hz
9
:0.20 Hz
4
:1.0 Hz
10
:0.15 Hz
5
:0.7 Hz
11
:0.10 Hz
4
P
f003
0 to 4
Hold mode
0
: Off
1
: Sample hold
2
: Peak hold
3
: Bottom hold
4
: Bipolar peak hold
1
P
f004
0000
to 1111
Latch
function
Function corresponds to an external input
latch.
Setting and latch
assignment
0
: Off
0 0 0 0 1
: On
Display value latch
Comparator latch
Analog output latch
Serial output latch
0000
(Binary)
D
* If you want to confirm the weighing value when setting the digital filter (F002),
press .
When the weighing value is displayed, the OK LED blinks and the display can
be set to zero by pressing .
Press to return the setting display.
8.3.3. Comparator
Function No.
setting range
Function
Description
Default value
setting type
f101
-9999 to
9999
Upper limit
value
Upper limit value of comparator.
Decimal point position depends on
Cf01
.
0000
D
f102
-9999 to
9999
Lower limit
value
Lower limit value of comparator.
Decimal point position depends on
Cf01
.
0000
D
f103
0 to 2
Comparator
mode
0
:Off
1
:On when the measured value is not
near zero
2
:Always on
2
P
f104
-9999 to
9999
Near zero Set the near zero range for Comparator
mode.
0000
D
f105
1 to 3
Hysteresis
mode
Hysteresis direction
1
:Upward 2-level judgment
2
:Upper / lower limit judgment
3
:Downward 2-level judgment
2
P
f106
0.0 to 5.0
Hysteresis
time
Set the hysteresis time by units of 0.1 seconds.
When
0.0
, the hysteresis is not used.
0.0
D
f107
00 to 99
Hysteresis
width
Set the hysteresis width to a two-digit value.
When
00
, the hysteresis is not used.
99
D
8.3.4. Analog
Function No.
setting range
Function
Description
Default value
setting type
f201
-9999 to 9999
0 V
output
Measurement value at DAV 0V output.
Decimal point position depends on
Cf01
.
0000
D
f202
-9999 to 9999
10 V
output
Measurement value at DAV 10V output.
Decimal point position depends on
Cf01
.
1000
D
f203
-9999 to 9999
4 mA
output
Measurement value at DAI 4 mA output.
Decimal point position depends on
Cf01
.
0000
D
F204
-9999 to 9999
20 mA
output
Measurement value at DAI 20 mA output.
Decimal point position depends on
Cf01
.
1000
D
8.3.5. Serial Communication
Function No.
Setting range
Function Description
Default value
Setting type
f301
2.4 to 38.4
Baud rate
2.4
: 2400 bps
4.8
: 4800 bps
9.6
: 9600 bps
19.2
: 19200 bps
38.4
: 38400 bps
2.4
P
f302
7 to 8
Data bit
length
7
:7 bits
8
:8 bits
7
P
f303
0 to 2
Parity
0
:None
1
: Odd
2
: Even
2
P
f304
1 to 2
Stop bit
1
:1 bit
2
:2 bits
1
P
f305
1 to 2
Terminator
1
:CRLF
2
:CR
1
P
f306
1 to 2
Communicat
ion mode
1
: Stream
2
: Manual print
2
P
f307
00 to 99
Model No.
ID that is added to the serial output
When setting to
0.0
, the ID is not
added.
00
D
f401
f402
f403
f404
f405
00 to 7F
Unit character1
Unit character2
Unit character3
Unit character4
Unit character5
Unit character added to serial
output.
Set with the hexadecimal ASCII
code.
All characters after
00
are ignored.
00
00
00
00
00
(Hexadecimal)
D
8.3.6. Error
Display Cause
Remedy
ad?e
The data cannot be
acquired from A/D
converter.
Repair is required.
eepe
Correct data cannot
be read from
EEPROM.
Perform initialization.
If the initialization does not clear
the error, repair is required.
Cale
Calibration data
error.
Perform calibration.
dt?e
A setting value is out
of range.
Check setting values and correct
if necessary.
Storing mode
Weighing mode
Span calibration