Page
56
AD-4405A Weighing Indicator
15.3.1. Setting the Interval Time
Step 1 Press the
SETPOINT(3)
key while pressing the
key to display
intval
for 2 seconds.
Step 2 Select which interval time to set, using the following keys.
MODE
key........The key to select serial output
int 5i
or printer output
int pr
.
ESC
key...........The key to return to the weight display.
ENTER
key ......The key to proceed to the time setting (Step 3).
Step 3 Set the interval time using the following keys. The key operation is the same as the
time setting.
0
-
9
key ..........The key to set the value at the blinking digit. Once set, the
blinking digit shifts to the right.
+/-
key..............The key to shift the blinking digit to the right.
CLEAR
key.......The key to reset the setting and display the interval time stored in
memory.
ESC
key...........The key to return to Step 2 without changing the interval time
setting.
ENTER
key ......The key to store the interval time setting and return to Step 2.
15.3.2. Interval Data Output / Print Output
Set the data output mode or print mode to “Manual mode” respectively to interval
data output or print out.
Pressing
the
START(7)
key while pressing the
key will start the interval
timer and output or print out the first data in the specified format.
The optional printer prints the interval time setting before printing the data.
Pressing the
STOP (8)
key while pressing the
key will stop the interval
timer and output or print out the data.
The optional printer prints “Intv. Stop” before printing the data.
The optional printer replaces overload data with spaces except for the decimal
point.
15. Common Items for Printer and Data Output