AD-4405A Weighing Indicator
Page
17
Press and hold the
ON/OFF
key and press the
ESC
key
No parameters are changed,
CanCel
is displayed and the calibration
mode is finished.
Step18 Press
the
ON/OFF
key to turn the display off.
5.3. Weighing Range Function
The weighing range function can select "single range" or "dual range". Specify each
weighing interval (division) for the multi-interval instrument. Each weighing interval is
displayed according to a net value or gross value.
Caution
When single range is used, performing this function is not required.
Example 1
The gross display.
Specified
parameters:
First range
Range = 50.00kg, division 0.02kg
Second range
Range = 100.00kg (maximum capacity), division 0.1kg
Display
0kg
to
50kg
: The
first range, division 0.02kg.
50kg to 100kg :
The second range, division 0.1kg.
Division
=
0.02kg
Division
=
0.1kg
0kg 50kg 100kg Gross
Weighing
value
Example 2
The net display using a 40kg tare value.
Specified parameters:
The same parameters as example 1.
Display
-40kg to 50kg :
The first range, division 0.02kg.
50kg
to
60kg
: The
second range, division 0.1kg.
Division = 0.02kg
Division=0.1kg
-40kg 0kg 50kg 60kg Net
Tare value
90kg 100kg Gross
40kg Weighing value
5.3. Weighing Range Function