5.2 “Stand-by” Mode
If left idle for a prolonged time the sealer will automatically switch into stand-by
mode (factory set at 60 minutes) and the temperature of the heating plate will
be reduced to 60°C.
When entering into stand-by mode the display shows the decrease of the
temperature in real time and the amber LED is switched off.
The temperature of the heating element will be maintained at 60°.
The display shows the actual temperature of the heating element and not the set
sealing temperature.
By pressing any key the unit will automatically begin to warm up the heating
plate to the pre-selected sealing temperature.
Refer to section 4.2 to change the time after which the unit switches into stand-
by mode or to switch off the stand-by function.
5.3 “Power-down” Mode
If the unit remains idle in stand-by mode for a prolonged time the system will go
into power down mode and the heating element will be switched off (factory set
at 10 hours). The amber LED is switched off.
The display shows the temperature decrease until it reaches 45°C, after this “LO”
will be displayed.
By pressing any key the unit will automatically begin to warm up the heating
element to the pre-selected sealing temperature.
Refer to section 4.2 to change the time after which the unit switches into power-
down mode.
5.4 Drawer Locking
It is possible to lock the drawer for transport or when the unit is not used to
prevent damage or dust contaminating the internal parts of the unit.
To lock the drawer proceed as follows:
•
While pressing the
key close the drawer
•
The plate lifting platform moves up and the drawer is locked
•
Switch OFF the unit
To unlock the drawer switch ON the unit.
14