3M™ PROTÉGÉ SG
PORTABLE, REUSABLE, SINGLE-GAS MONITOR
MAINTENANCE
087-0065, Rev D/February 2019
6-11
Figure 6-10: Installing New Filter
6.3. Error Codes
provides a list of error codes that display on the LCD when applicable.
Table 6-3: Error Codes
Error
Code
Cause
Solution
E01
Configuration memory
error
Reset the monitor configuration
E02
Datalog memory error
Erase the datalog memory
E03
Program memory error
Reprogram monitor firmware
E04
Datalog memory is full
Use the PGD software to read the datalog and save the
datalog file. Then delete the datalog through the PGD soft-
ware, refer to PGD software user guide.
E05
Battery depleted
Chargeable battery: charge the battery
un-rechargeable battery: replace the battery
E06
Missing sensor/wrong sen-
sor installation
Reinstall the sensor or contact the service center
E07
Sensor can not communi-
cate with Mipex sensor
Re-install sensor module and restart the monitor.
E08
Temperature sensor error
or exceed the range
Take the monitor to a normal temperature area.
E09
Temperature of Mipex sen-
sor has a dramatic change
Take the monitor to a normal temperature area.
E10
G-sensor error
This is not applied to the Protégé SG.
E11
Charge error
This is not applied to the Protégé SG.
E12
ADC full
Contact the service
Sensor Cover
(Inside)
New
Filter