8.  Clip the microphone, with windscreen attached to one’s shoulder and 

     secure the NoisePro to the person’s waist.  
9.  Press         key to begin data collection.    The run icon        will appear in 

     the lower right corner of the display.

End Study

10.

  

Press          key to stop study.  The pause icon “

II

” will appear in the 

      lower right corner of the display.

Post-Calibration

11. 

 

From the START menu, press the CAL softkey.  The CAL screen 

      appears with CALIBRATE highlighted. 
12. Turn the calibrator on, connect microphone to calibration adapter, 

      connect calibration adapter to calibrator.  

13.  Press        key and POST-CALIBRATION screen appears.  Press         

       key to save (store).  

• 

 

NOTE: 

 in a Post-Calibration, the SPL value cannot be adjusted.

View Data

14.  From the “START” screen, highlight “VIEW CURRENT STUDY” and press the     

             key.   Press                keys to scroll through LEVEL, AVG, and DOSE 

       measurements.  Press            keys to scroll to other dosimeter measurements.  

• 

 NOTE: 

 the displayed measurement screens are dependent on the model     

      type and number of defined setups/virtual meters on your instrument.  

(Examples:  

       NoisePro DLX dosimeters 1, 2, 3, & 4; NoisePro & NoisePro DL dosimeters 1 & 2).

Содержание NoisePro Series

Страница 1: ...3MOccupational Health and Environmental Safety Division 3M NoisePro Personal Noise Dosimeter NoisePro Series Quick Start Guide ...

Страница 2: ...oisePro erases all previously stored data from memory Quick Start Steps Pre Calibration Reset 4 From the START menu press the CAL softkey The CAL screen appears with CALIBRATE highlighted 5 Turn the calibrator on connect microphone to calibration adapter connect calibration adapter to calibrator 6 Press and the PRE CALIBRATION screen appears If required press the keys to adjust the displayed value...

Страница 3: ...2 Turn the calibrator on connect microphone to calibration adapter connect calibration adapter to calibrator 13 Press key and POST CALIBRATION screen appears Press key to save store NOTE in a Post Calibration the SPL value cannot be adjusted View Data 14 From the START screen highlight VIEW CURRENT STUDY and press the key Press keys to scroll through LEVEL AVG and DOSE measurements Press keys to s...

Страница 4: ...NoisePro Overview Softkey menus Softkeys used for additional menu options Up Down arrow keys Left Right arrow keys Escape On Off Enter key Start Stop studies ...

Страница 5: ......

Страница 6: ...e highest A weighted sound that occurred applying the Slow response time Peak level the highest instantaneous sound level that the microphone detects Unlike the Max Level the peak is detected independently of the slow or fast response time the unit is set for Response Time Fast Slow Impulse how quickly the circuitry responds to changing noise levels These are ANSI IEC defined response times Most o...

Страница 7: ...eighting Slow response 5 dB ER 90 dB CL 80 dB Threshold OSHA PEL Permissible Exposure Level A Weighting Slow response 5 dB ER 90 dB CL 90 dB Threshold MSHA HC Hearing Conservation A Weighting Slow response 5 dB ER 90 dB CL 80 dB Threshold MSHA PEL Permissible Exposure Level A Weighting Slow response 5 dB ER 90 dB CL 90 dB Threshold ACGIH A Weighting Slow response 3 dB ER 85 dB CL 80 dB Threshold 2...

Страница 8: ... 9001 Registered Company ISO 17025 Accredited Calibration Lab 1060 Corporate Center Drive Oconomowoc WI 53066 Customer Service 262 567 9157 Toll Free 800 245 0779 www 3m com detection www questtechnologies com Please recycle Printed in USA 2011 3M All rights reserved 098 574 Rev D 1 12 ...

Отзывы: