background image

Turn On

   

1.  Turn on the NoisePro by pressing and releasing          key.  

     The display will initialize and sequence to the “START” screen. 
2.  If “LOBAT” is in display, put new batteries in the NoisePro.

3.  Press and hold RESET softkey.  The display counts down from 5 then 

     shows “Sessions DELETED” for NoisePro DLX or “Studies DELETED” 

     for the NoisePro DL and NoisePro models.

• 

 NOTE: 

 Resetting the NoisePro erases all previously stored data from memory.

Quick Start Steps

Pre-Calibration

Reset

   4.  From the START menu, press the CAL softkey.  The CAL screen 

        appears with CALIBRATE highlighted. 
   5.  Turn the calibrator on, connect microphone to calibration adapter, 

        connect calibration adapter to calibrator.  

   6.   Press       and the PRE-CALIBRATION screen appears.  If required, 

         press the             keys to adjust the displayed value to match 

         the  calibrator output.  Press       key to  save (store) the pre-calibration.  
   7.  Press and release  

 key to return to START screen.

Pre-Calibration 

screen

and matching 

output

Содержание NoisePro Series

Страница 1: ...3MOccupational Health and Environmental Safety Division 3M NoisePro Personal Noise Dosimeter NoisePro Series Quick Start Guide ...

Страница 2: ...oisePro erases all previously stored data from memory Quick Start Steps Pre Calibration Reset 4 From the START menu press the CAL softkey The CAL screen appears with CALIBRATE highlighted 5 Turn the calibrator on connect microphone to calibration adapter connect calibration adapter to calibrator 6 Press and the PRE CALIBRATION screen appears If required press the keys to adjust the displayed value...

Страница 3: ...2 Turn the calibrator on connect microphone to calibration adapter connect calibration adapter to calibrator 13 Press key and POST CALIBRATION screen appears Press key to save store NOTE in a Post Calibration the SPL value cannot be adjusted View Data 14 From the START screen highlight VIEW CURRENT STUDY and press the key Press keys to scroll through LEVEL AVG and DOSE measurements Press keys to s...

Страница 4: ...NoisePro Overview Softkey menus Softkeys used for additional menu options Up Down arrow keys Left Right arrow keys Escape On Off Enter key Start Stop studies ...

Страница 5: ......

Страница 6: ...e highest A weighted sound that occurred applying the Slow response time Peak level the highest instantaneous sound level that the microphone detects Unlike the Max Level the peak is detected independently of the slow or fast response time the unit is set for Response Time Fast Slow Impulse how quickly the circuitry responds to changing noise levels These are ANSI IEC defined response times Most o...

Страница 7: ...eighting Slow response 5 dB ER 90 dB CL 80 dB Threshold OSHA PEL Permissible Exposure Level A Weighting Slow response 5 dB ER 90 dB CL 90 dB Threshold MSHA HC Hearing Conservation A Weighting Slow response 5 dB ER 90 dB CL 80 dB Threshold MSHA PEL Permissible Exposure Level A Weighting Slow response 5 dB ER 90 dB CL 90 dB Threshold ACGIH A Weighting Slow response 3 dB ER 85 dB CL 80 dB Threshold 2...

Страница 8: ... 9001 Registered Company ISO 17025 Accredited Calibration Lab 1060 Corporate Center Drive Oconomowoc WI 53066 Customer Service 262 567 9157 Toll Free 800 245 0779 www 3m com detection www questtechnologies com Please recycle Printed in USA 2011 3M All rights reserved 098 574 Rev D 1 12 ...

Отзывы: