19
Liqui-Cel
®
Membrane Contactors
Cleaning Guidelines
B. Final Drying
The final drying step involves flowing a clean, dry, filtered (0.2 micron) gas into the top shellside
port. Using a warm gas will reduce drying time. Nitrogen is preferred in the final drying step as hot
air can shorten the membrane life. See Table 9 for estimated flow rate and drying time guidelines.
Keep the lower shellside port closed.
Table 8: Final Drying Conditions
* Maximum gas pressure =
10 psig (0.7 kg/cm
2
)
C. Drying Schematics
Illustration 5: Bulk Water Removal/Initial Drying Step
F
P
= Flow Indicator
= Pressure Gauge
= Valve
F
P
Shellside
Shellside
Lumenside
Lumenside
Keep liquid outlet port open gas can
exit through the port
Liqui-Cel
®
Membrane Contactor Size
Gas Flow Rate*
Estimated Drying Time
2 x 6
0.5 scfm (0.84 m
3
/hr)
1 hr @ 60 ºC (140 ºF)
2.5 x 8
1 scfm (1.7 m
3
/hr)
1 hr @ 60 ºC (140 ºF)
4 x 13
5 scfm
4 hr @ 60 ºC (140 ºF)
4 x 28
10 scfm (17 m
3
/hr)
4 hr @ 60 ºC (140 ºF)
6 x 28
25 scfm ( 40 m
3
/hr)
8 hr @ 60 ºC (140 ºF)
8 x 20
20 scfm
8 hr @ 60 ºC (140 ºF)
8 x 40
30 scfm
16 hr @ 60 ºC (140 ºF)
8 x 80
60 scfm
24 hr @ 60 ºC (140 ºF)
10 x 28
40 scfm (120 m
3
/hr)
16 hr @ 60 ºC (140 ºF)
14 x 28
50 scfm ( 130 m
3
/hr)
24 hr @ 60 ºC (140 ºF)
14 x 40
100 scfm
24 hr @ 60 ºC (140 ºF)