3.
Tap [Power off].
4.
A confirmation message the luminometer will shut down appears.
5.
Tap [OK]. The display screen goes black to confirm the power off process has completed.
6.
Or, tap [Cancel] to cancel the power off process if you wish to continue using the
luminometer.
Sleep Mode
The Clean-Trace LM1 luminometer conserves battery power if the touch screen or the cursor
control buttons are inactive. After 5 minutes of inactivity, the luminometer goes into Sleep
mode and the screen darkens.
After 2 further hours of inactivity, the luminometer automatically shuts down and must be
powered on again before use.
Waking the Clean-Trace LM1 luminometer from Sleep Mode
1. Press and hold the up-arrow button on the cursor control pad for 3 seconds.
2. The screen displays the information that appeared before the Sleep mode began so you can
continue working.
If the luminometer does not wake from the Sleep mode, the luminometer might be turned off.
Press and hold the on/off button for 3 seconds to turn it on.
If the battery is at a low charge before the luminometer goes into Sleep mode, this may result
in the battery becoming completely discharged. This will then result in the luminometer failing
to switch on due to insufficient charge.
In this case, the luminometer will need to be put on charge from the mains charger (not the PC
USB). After 30 minutes of charging you should then be able to switch on the machine
however it will likely require further charge to reach adequate charge for use. To maintain
battery performance, the luminometer should be switched on within 3 hours after placing the
unit on charge.
For further support, please visit
www.3M.com/foodsafety/cleantracesupport
Web: www.3M.com/foodsafety
Email:
3M and Clean-Trace are trademarks of 3M. © 3M 2018