Declaration – electromagnetic immunity - continued

The 3M Littmann® Electronic Stethoscope, Model 3100, is intended for use in the electromagnetic environment specifi ed 
below. The customer or the user of Model 3100 should assure that it is used in such an environment.

Immunity test

IEC 60601 test level

Compliance level

Electromagnetic environment –guidance

Portable and mobile RF communications equipment 
should be used no closer to any part of the Model 
3100, including cables, than the recommended 
separation distance calculated from the equation 
applicable to the frequency of the transmitter.
Recommended separation distance:

Conducted RF IEC 
61000-4-6

3 Vrms
150 kHz to 80 MHz

Not applicable

d = 1,2 

9

 P

Radiated RF IEC 
61000-4-3

3 V/m 
80 MHz to 2.5 GHz

3 V/m

80 MHz to 2.5 GHz

d = 1,2 

9

 P 80 MHz to 800 MHz

d = 2,3 

9

 P 800 MHz to 2,5 GHz

where P is the maximum output power rating of the 
transmitter in watts (W) according to the transmitter 
manufacturer and d is the recommended separation 
distance in meters (m).
Field strengths from fi xed RF transmitters, as determined 
by an electromagnetic site survey, 

a

 should be less than 

the compliance level in each frequency range, 

b

Interference may occur in the vicinity of equipment 
marked with the following symbol:

NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and 
refl ection from structures, objects and people.

a

 Field strengths from fi xed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile 

radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. 
To assess the electromagnetic environment due to fi xed RF transmitters, an electromagnetic site survey should be 
considered. If the measured fi eld strength in the location in which the Model 3100 is used exceeds the applicable RF 
compliance level above, the Model 3100 should be observed to verify normal operation. If abnormal performance is 
observed, additional measures may be necessary, such as re-orienting or relocating the Model 3100. 

b

 Over the frequency range 150 kHz to 80 MHz, fi eld strengths should be less than 3 V/m.

Recommended Separation Distances Between Portable and Mobile RF Communications Equipment and the 3M™ 
Littmann® Electronic Stethoscope Model 3100

The Model 3100 is intended for use in an electromagnetic environment in which radiated RF disturbances are 
controlled. The customer or the user of the Model 3100 can help prevent electromagnetic interference by maintaining a 
minimum distance between portable and mobile RF communications equipment (transmitters) and the Model 3100 as 
recommended below, according to the maximum output of the communications equipment.

Rated maximum output 

power of transmitter, P [W]

Separation distance according to frequency of transmitters, d [m]

150 kHz to 80 MHz

80 Mz to 800 MHz

d = 1,2 

9

 

P

800 MHz to 2,5 GHz

d = 2,3 

9

 

P

0.01

0.12

0.12

0.23

0.1

0.38

0.38

0.73

1

1.2

1.2

2.3

10

3.8

3.8

7.3

100

12

12

23

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters 
(m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output 
power rating of the transmitter in watts (W) according to the transmitter manufacturer.

NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.

NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and 
refl ection from structures, objects and people.

Содержание Littmann 3100

Страница 1: ......

Страница 2: ...this latest electronic stethoscope from the Littmann brand Hear it With Confidence SAFETY INFORMATION Please read understand and follow all safety information contained in these instructions prior to...

Страница 3: ...ope Read understand and follow all the safety information on the battery package Intended Use The 3M Littmann Electronic Stethoscope Model 3100 is intended for medical diagnostic purposes only It may...

Страница 4: ...ope Read understand and follow all the safety information on the battery package Intended Use The 3M Littmann Electronic Stethoscope Model 3100 is intended for medical diagnostic purposes only It may...

Страница 5: ...o during standby mode When stethoscope diaphragm or any button is pressed LCD display settings will reappear and stethoscope will be ready to use After two hours of non use the stethoscope will leave...

Страница 6: ...o during standby mode When stethoscope diaphragm or any button is pressed LCD display settings will reappear and stethoscope will be ready to use After two hours of non use the stethoscope will leave...

Страница 7: ...hs As the alkaline battery life depletes the icon will change as follows The battery icon begins to blink when only a few hours of battery capacity remains CAUTION When the battery is completely deple...

Страница 8: ...hs As the alkaline battery life depletes the icon will change as follows The battery icon begins to blink when only a few hours of battery capacity remains CAUTION When the battery is completely deple...

Страница 9: ...OW setting uses less battery power To increase battery life ENABLE the Auto Off feature and select the LOW power setting The power settings provide different degrees of power consumption In general th...

Страница 10: ...OW setting uses less battery power To increase battery life ENABLE the Auto Off feature and select the LOW power setting The power settings provide different degrees of power consumption In general th...

Страница 11: ...ering off Backlight remains lit for 5 seconds after pressing power button MEDIUM Remains on full power for 5 minutes before entering standby Remains in standby for 2 hours before powering off Backligh...

Страница 12: ...ering off Backlight remains lit for 5 seconds after pressing power button MEDIUM Remains on full power for 5 minutes before entering standby Remains in standby for 2 hours before powering off Backligh...

Страница 13: ...Cleaning Other Parts of the Stethoscope Eartips eartubes plastic tubing and chestpiece can be wiped clean with alcohol Eartips may be removed for a more thorough cleaning NOTICE Do not immerse the ste...

Страница 14: ...Cleaning Other Parts of the Stethoscope Eartips eartubes plastic tubing and chestpiece can be wiped clean with alcohol Eartips may be removed for a more thorough cleaning NOTICE Do not immerse the ste...

Страница 15: ...e 3M Littmann Electronic Stethoscope Model 3100 is intended for use in the electromagnetic environment specified below The customer or the user of Model 3100 should assure that it is used in such an e...

Страница 16: ...e 3M Littmann Electronic Stethoscope Model 3100 is intended for use in the electromagnetic environment specified below The customer or the user of Model 3100 should assure that it is used in such an e...

Страница 17: ...accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If the measured field strength in the location in which the Model 31...

Страница 18: ...accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If the measured field strength in the location in which the Model 31...

Страница 19: ......

Отзывы: