When the AUTO OFF feature is enabled, the stethoscope will enter standby mode after a period of time determined by the 
POWER setting if there is no contact detected on the diaphragm AND the buttons are not activated. When the AUTO OFF 
feature is disabled, the stethoscope will enter standby mode after a period of time determined by the POWER setting from 
the last button activation. In addition, when the AUTO OFF feature is disabled, there will be an audible alert 10 seconds prior 
to entering standby mode. After entering standby mode, the Model 3100 will be powered on when contact is detected on the 
diaphragm or by button activation. Otherwise, the Model 3100 will completely power off after the period of time determined 
by the POWER setting (see below).

With AUTO OFF feature ENABLED:

SETUP Options

Detail

POWER

HIGH

Remains on full power for 

30 seconds

 before entering standby

Remains in standby for

 5 hours 

before powering off

Backlight remains lit for

 5 seconds 

after pressing the power button

MEDIUM

Remains on full power for 

20 seconds

 before entering standby

Remains in standby for 

2 hours

 before powering off

Backlight remains lit for 

5 seconds 

after pressing the power button

LOW

Remains on full power for 

10 seconds 

before entering standby

Remains in standby for 

30 minutes

 before powering off

Backlight remains lit for 

3 seconds 

after pressing power button

With AUTO OFF feature DISABLED:

SETUP Options

Detail

POWER

HIGH

Remains on full power for 

7 minutes

 before entering standby

Remains in standby for 

5 hours 

before powering off

Backlight remains lit for

 5 seconds 

after pressing power button

MEDIUM

Remains on full power for

 5 minutes

 before entering standby

Remains in standby for 

2 hours

 before powering off

Backlight remains lit 

5 seconds

 after pressing power button

LOW

Remains on full power for

 3 minutes 

before entering standby

Remains in standby for 

30 minutes

 before powering off

Backlight remains lit 

3 seconds

 after pressing power button

SETUP

Id

exit

power

To obtain model and software version information: 

Select the “ID” option in the Setup menu. Use the (-) and (+) buttons 

to scroll through the list of reference numbers. Press (M) to return to the operating mode of the stethoscope.

 

To leave the Setup menu:

 Press the Filter button or press the (M) button to select the EXIT option. This will return the 

stethoscope to its normal operating state.

 

10.  Other Operating Considerations 

Operating range is -22º to 104ºF (-30º to 40ºC), 15 to 93% relative humidity.

Storage and transport range is -40º to 131ºF (-40º to 55ºC), 15 to 93% relative humidity.

To extend the life of your stethoscope, avoid extreme heat, cold, solvents and oils. Remove the battery whenever the 
stethoscope will not be used for several months. 

If you plan to use the stethoscope below 0ºF (-18ºC) you should use a lithium battery to insure proper function.

Failure to follow care and maintenance recommendations could result in damage to the internal components of the Littmann 
Electronic Stethoscope. Internal damage could cause malfunction of the product, ranging from a slight decrease in auditory 
response to complete failure of the product.

If you experience any problems with the electronic stethoscope, do not attempt to repair it yourself. Please notify our 3M 
Health Care Service Center for directions on shipping and receiving.

Содержание Littmann 3100

Страница 1: ......

Страница 2: ...this latest electronic stethoscope from the Littmann brand Hear it With Confidence SAFETY INFORMATION Please read understand and follow all safety information contained in these instructions prior to...

Страница 3: ...ope Read understand and follow all the safety information on the battery package Intended Use The 3M Littmann Electronic Stethoscope Model 3100 is intended for medical diagnostic purposes only It may...

Страница 4: ...ope Read understand and follow all the safety information on the battery package Intended Use The 3M Littmann Electronic Stethoscope Model 3100 is intended for medical diagnostic purposes only It may...

Страница 5: ...o during standby mode When stethoscope diaphragm or any button is pressed LCD display settings will reappear and stethoscope will be ready to use After two hours of non use the stethoscope will leave...

Страница 6: ...o during standby mode When stethoscope diaphragm or any button is pressed LCD display settings will reappear and stethoscope will be ready to use After two hours of non use the stethoscope will leave...

Страница 7: ...hs As the alkaline battery life depletes the icon will change as follows The battery icon begins to blink when only a few hours of battery capacity remains CAUTION When the battery is completely deple...

Страница 8: ...hs As the alkaline battery life depletes the icon will change as follows The battery icon begins to blink when only a few hours of battery capacity remains CAUTION When the battery is completely deple...

Страница 9: ...OW setting uses less battery power To increase battery life ENABLE the Auto Off feature and select the LOW power setting The power settings provide different degrees of power consumption In general th...

Страница 10: ...OW setting uses less battery power To increase battery life ENABLE the Auto Off feature and select the LOW power setting The power settings provide different degrees of power consumption In general th...

Страница 11: ...ering off Backlight remains lit for 5 seconds after pressing power button MEDIUM Remains on full power for 5 minutes before entering standby Remains in standby for 2 hours before powering off Backligh...

Страница 12: ...ering off Backlight remains lit for 5 seconds after pressing power button MEDIUM Remains on full power for 5 minutes before entering standby Remains in standby for 2 hours before powering off Backligh...

Страница 13: ...Cleaning Other Parts of the Stethoscope Eartips eartubes plastic tubing and chestpiece can be wiped clean with alcohol Eartips may be removed for a more thorough cleaning NOTICE Do not immerse the ste...

Страница 14: ...Cleaning Other Parts of the Stethoscope Eartips eartubes plastic tubing and chestpiece can be wiped clean with alcohol Eartips may be removed for a more thorough cleaning NOTICE Do not immerse the ste...

Страница 15: ...e 3M Littmann Electronic Stethoscope Model 3100 is intended for use in the electromagnetic environment specified below The customer or the user of Model 3100 should assure that it is used in such an e...

Страница 16: ...e 3M Littmann Electronic Stethoscope Model 3100 is intended for use in the electromagnetic environment specified below The customer or the user of Model 3100 should assure that it is used in such an e...

Страница 17: ...accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If the measured field strength in the location in which the Model 31...

Страница 18: ...accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If the measured field strength in the location in which the Model 31...

Страница 19: ......

Отзывы: