8
3.0 Specifications
Description
Typical Value
Dimensions
Base unit: (152 x 152 x 152) mm
(6 x 6 x 6) inches
Weight
1.8 kg (4 lbs.)
High Voltage Cascade
> 1100 V positive or negative
Power Supplies
Built-in NiMH-rechargeable batteries, 1400 mAh -
Recharge time approximately 14 hours (unit off) when batteries are fully
discharged
AC adaptor: secondary side, 12VDC/500 mA
Operating Time
(rechargeable batteries):
4 hours (approx.) with full charge.
Accuracy
± 10 % of range end value (digitized)
Operating Functions
CPM (positive/negative/automatic), Voltmeter, and Fieldmeter
Interfaces
Analog output ± 2V (± 1 V, in 500V range for voltmeter).
Input to output ratio in Voltmeter Mode: 25VDC = 12.5 to 1, 100VDC = 50 to 1,
500VDC = 500 to 1, 1,000VDC = 500 to 1, 5,000VDC = 2500 to 1.
Serial PC-COM RS232
3M
™
Remote Field
Sensor Probe, 711RS
34 mm diameter - 190 gms (6.7 oz.)
Displays
Two, 11-segment positive & negative LED-bar charge indicators 16-digit
alphanumeric dual row LCD
Settings: CPM - Operating
Function
Starting voltage: 600V - 1200V in 1V-steps
Stop voltage: 1V - 500V in 1V-steps
(in decimal mode)
Static Decay Time
0.1 seconds - 99.9 seconds
Offset-voltage Time
0 seconds (Indefinite time, CPM in floating mode)
1 - 10 seconds (1 second steps)
10 - 60 seconds (10 second steps)
Voltmeter Operating
Function
Ranges: 25V, 100V, 500V, 1.0 kV, 5.0 kV and auto range
Fieldmeter Operating
Function
Ranges: Manual 1.25 kV/m, 5 kV/m, 25 kV/m, 50 kV/m, 250 kV/m, and
automatic
Plate Electrode
VA-steel (152 x 152) mm/(6 x 6) inches, removable, capacitance (20 ± 2) pF
Cup Electrode
Gold-plated electrode with 4 mm-banana socket, for voltage measurements
Selection of Operating
Function
Pre-setting is “FIELDMETER”, additional automatic settings by applying the
plate or cup electrode