GB / 34-8718-8817-7

1

3M™ Bair Hugger™ Warming Unit Model 775

Service Manual

Please forward to Biomedical Engineering Department

Introduction

Description of the 3M™ Bair Hugger™ Temperature

Management System

The 3M™ Bair Hugger™ temperature management system consists of a

Model 775 warming unit (with optional rolling stand, and other accessories)

along with a 3M warming blanket, Bair Hugger warming gown, or the 3M™

Ranger™ 241 blood/fluid warming set.
The Bair Hugger warming unit is attached to the warming blanket or

warming gown by means of a flexible hose. Warm air is generated in the

warming unit and flows through the warming unit hose and into the blanket

or gown. Depending on the model, the 3M blanket or gown is placed

either around, over, or underneath the patient. Small perforations on the

blanket or gown allow the warm air to be dispersed over the patient. For

blood/fluid warming applications, the Model 241 blood/fluid warming

set is inserted in the warming unit hose. When the warming unit is turned

ON and a temperature setting is selected, warm air flows over the Model

241 tubing and warmed blood/fluid exits from the distal end of the tubing.

For additional information on 3M blankets, gowns, the 241 warming set, or

other accessories visit us online at bairhugger.com.
This manual includes operating and service instructions and warming unit

specifications for the Model 775 warming unit. The warming unit can be

used in all clinical settings where the patient may feel too warm or too

cold. For information about using 3M blankets, gowns, or the 241 blood/

fluid warming set with the warming unit, refer to the respective Instructions

for Use.

Indications

The Bair Hugger temperature management system is intended to prevent

and treat hypothermia. In addition, the temperature management system

can be used to provide patient thermal comfort when conditions exist

that may cause patients to feel too warm or too cold. The temperature

management system can be used with adult and pediatric patients.

The Bair Hugger temperature management system should only be

used by trained medical professionals.

Federal law (USA) restricts this device to sale by or on the order of a

licensed healthcare professional.

Definition of Symbols

The following symbols may appear on the product’s labeling or

exterior packaging.

An equipotentiality plug (grounded) conductor

other than a protective earth conductor or a neutral

conductor, providing a direct connection between

electrical equipment and the potential equalization

busbar of the electrical installation. Please consult IEC

60601-1 for requirements.

Fuse

Type BF equipment (patient applied)

Date of manufacture

Manufacturer

Voltage, alternating current (AC)

This system is subject to European WEEE

Directive 2002/96/EC. This product contains

electrical and electronic components and must not be

disposed of using standard refuse collection. Please

consult local directives for disposal of electrical and

electronic equipment.

Protective earth ground

No free hosing

High fan setting

Low fan setting

Non-explosion proof

Consult instructions for use

Follow instructions for use

Recycle to avoid environmental contamination. This

product contains recyclable parts. For information on

recycling, please contact your nearest 3M Service

Center for advice.

Attention; see instructions for use

Keep dry

Temperature limit

PDF Scaled at 75.0%

Содержание Bair Hugger 775

Страница 1: ... license in Canada 2016 3M All rights reserved Issue Date 2016 08 34 8718 8817 7 3M Deutschland GmbH Health Care Business Carl Schurz Str 1 41453 Neuss Germany Model 775 Warming Unit Service Manual Normothermia System 1 3M Bair Hugger Warming Unit Model 775 2 Appareil de réchauffement Bair Hugger 3M modèle 775 3 3M Bair Hugger Wärmeeinheit Modell 775 PDF Scaled at 75 0 ...

Страница 2: ...3MTM Bair Hugger Patient Adjustable Warming Unit Requester Frank Boeshart Debra Staack Reference Die Doc Size 8 5 x 11 Item Spec 34 8718 8817 7 Supersedes Cat Product Model 775 Structure Service Manual PA 34871888177 indb GA InDesign CC 2014 BOOK PROCESS BLACK PMS 280 4394547 3 08 04 16 New Build kmh 08 26 16 Changes per RA proof report TN 09 15 16 Final release DJ PDF Scaled at 75 0 ...

Страница 3: ...ernative Modes 6 Accessing the Alternative Modes 7 Exiting an Alternative Mode 7 Required Preventative Maintenance 7 Calibrating the Operating Temperature 7 Testing the Over Temperature Detection System 8 Viewing the Fault Code Log 8 Clearing the Fault Code Log 8 Fault Code Table 9 Viewing the Hour Meter 9 Viewing the Operating Timer 9 Replacing the Air Filter 9 Replacing the Hose 9 Detaching Warm...

Страница 4: ... with the warming unit refer to the respective Instructions for Use Indications The Bair Hugger temperature management system is intended to prevent and treat hypothermia In addition the temperature management system can be used to provide patient thermal comfort when conditions exist that may cause patients to feel too warm or too cold The temperature management system can be used with adult and ...

Страница 5: ... with the warming unit refer to the respective Instructions for Use Indications The Bair Hugger temperature management system is intended to prevent and treat hypothermia In addition the temperature management system can be used to provide patient thermal comfort when conditions exist that may cause patients to feel too warm or too cold The temperature management system can be used with adult and ...

Страница 6: ...ing blanket gown The Bair Hugger warming unit is intended to be used with the hose properly attached to its corresponding warming blanket gown and in accordance with good practices for operating room sterile technique CAUTION To reduce the risk of patient or caregiver injury If mounted on an IV pole the distance from the bottom of the warming unit to the floor must be less than 44 112 cm and the I...

Страница 7: ...ing blanket gown The Bair Hugger warming unit is intended to be used with the hose properly attached to its corresponding warming blanket gown and in accordance with good practices for operating room sterile technique CAUTION To reduce the risk of patient or caregiver injury If mounted on an IV pole the distance from the bottom of the warming unit to the floor must be less than 44 112 cm and the I...

Страница 8: ...ation mode allows the technician to view important system information including the firmware system identification serial port baud rate 5V power supply status and the control board serial number The Over temperature Detection Test mode allows the user to test the operation of the independent over temperature shutdown system Accessing the Alternative Modes To access the alternative modes you must ...

Страница 9: ...ation mode allows the technician to view important system information including the firmware system identification serial port baud rate 5V power supply status and the control board serial number The Over temperature Detection Test mode allows the user to test the operation of the independent over temperature shutdown system Accessing the Alternative Modes To access the alternative modes you must ...

Страница 10: ...urs and minutes 3 Press either the 32 C or 38 C button to view the next fault code in the log The warming unit stores a maximum of 5 fault codes in non volatile memory 4 Press the Alternative Mode button to return the warming unit to its previous operating mode Clearing the Fault Code Log Put the warming unit in Fault Code mode by pressing the Alternative Mode button and the 32 C button simultaneo...

Страница 11: ...urs and minutes 3 Press either the 32 C or 38 C button to view the next fault code in the log The warming unit stores a maximum of 5 fault codes in non volatile memory 4 Press the Alternative Mode button to return the warming unit to its previous operating mode Clearing the Fault Code Log Put the warming unit in Fault Code mode by pressing the Alternative Mode button and the 32 C button simultaneo...

Страница 12: ...ordance with hospital practices for cleaning OR equipment After every use wipe the warming unit the outside of the warming unit hose and any other surfaces that may have been touched Use a damp soft cloth and a hospital approved mild detergent germicidal disposable wipes disinfecting towelettes or antimicrobial spray 3 Let air dry or dry with a separate clean soft cloth Notice Do not use cleaning ...

Страница 13: ...ordance with hospital practices for cleaning OR equipment After every use wipe the warming unit the outside of the warming unit hose and any other surfaces that may have been touched Use a damp soft cloth and a hospital approved mild detergent germicidal disposable wipes disinfecting towelettes or antimicrobial spray 3 Let air dry or dry with a separate clean soft cloth Notice Do not use cleaning ...

Страница 14: ...spenser Shipping label Marking pen METHOD 1 Call 3M technical support see page 12 for contact information to get a Return Authorization RA number and a service carton 2 Remove the top foam piece If the plastic liner is available use it to enclose the warming unit 3 Lower the warming unit carefully into the carton 4 Replace the top foam piece Observe proper orientation Hose will extend out in the c...

Страница 15: ...spenser Shipping label Marking pen METHOD 1 Call 3M technical support see page 12 for contact information to get a Return Authorization RA number and a service carton 2 Remove the top foam piece If the plastic liner is available use it to enclose the warming unit 3 Lower the warming unit carefully into the carton 4 Replace the top foam piece Observe proper orientation Hose will extend out in the c...

Страница 16: ... 5 GHz d 2 3 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manuf...

Страница 17: ... 5 GHz d 2 3 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manuf...

Страница 18: ...ct the filter 2 Return filter or replace as necessary Note Replace every 12 months or every 500 hours PN 90047 CALIBRATE THE OPERATING TEMPERATURE 1 Attach the Model 22110 temperature test unit to the end of the warming unit hose 2 Run in Ambient mode and high fan setting for 10 minutes 3 Enter the calibration mode a Press the Alternate Mode button and Ambient button together for approximately 3 s...

Страница 19: ...ct the filter 2 Return filter or replace as necessary Note Replace every 12 months or every 500 hours PN 90047 CALIBRATE THE OPERATING TEMPERATURE 1 Attach the Model 22110 temperature test unit to the end of the warming unit hose 2 Run in Ambient mode and high fan setting for 10 minutes 3 Enter the calibration mode a Press the Alternate Mode button and Ambient button together for approximately 3 s...

Страница 20: ...irculation d air 24 Mode Veille 24 Pannes 24 Surchauffe 25 Autres modes 25 Accès aux autres modes 25 Sortie des autres modes 25 Entretien préventif nécessaire 25 Étalonnage de la température de fonctionnement 25 Test du système de détection de surchauffe 26 Affichage du journal des codes de panne 27 Effacement du journal des codes de panne 27 Tableau des codes de panne 27 Affichage du compteur d h...

Отзывы: