6

Scanning Step 3: Basic Scanning Procedure

1. 

Place the open end of the tip of the heron over the area of interest, and keep it 

stationary.

2. 

Click the scan button 

(A) 

once in QuantorClinic, or 

press the main button 

on the handpiece

.

3. 

Keep the handpiece 

stationary over the 

object of interest until 

you hear a clicking sound 

(if there is no clicking, 

please adjust the sound 

settings on your computer)

. You should see a 3D surface in the model view 

window, with a green rectangle which indicates the last scanned position 

(B)

.

4. 

Continue scanning by moving the scanner slowly over and around the area 

of interest. The Heron IOS can scan at a range of 0 to 12mm, however, better 

scans are obtained at a closer range

and

 it is okay if the tip touches the 

teeth. 

During scanning, you should continue to hear a rapid clicking sound and 

observe a green rectangle indicating the last scanned area. 

5.  If the clicking sound 

stops

 and a red rectangle 

is visible on the model, 

this indicates that data 

from the handpiece is no 

longer being integrated 

into the model. 

In order 

to continue scanning, 

place the handpiece 

over any area that 

you have previously 

scanned.

 When new data is being added to the model, the red rectangle 

will change to green, the clicking sound will resume, and you can continue 

scanning. 

6.  The scan can be paused by pressing the scan button on the handpiece, or 

the scan button (A) in QuantorClinic.

 It is recommended to inspect the 3D 

model, as it may be necessary to re-scan certain areas in order to fill in gaps, or 

improve the scan quality.

A

B

Содержание Heron IOS

Страница 1: ...g the Tip 2 Patient Environment 2 Tip Cleaning Sterilization 3 Scanning Step 1 Creating a Case 5 Scanning Step 2 Scanning Preparation 5 Scanning Step 3 Basic Scanning Procedure 6 Scanning Step 4 Final...

Страница 2: ...ng station on a flat stable surface and place the Heron IOS hand piece securely in its docking station Step 2 Connect the AC DC power adapter cable to the docking station the connector socket is locat...

Страница 3: ...ine extending 1 8 meter 6 feet around patient and system Depress bottom button 1 and rotate tip at 90 increments 2 Tip should be locked into position prior to scanning 1 2 1 2 Attach the tip to the ha...

Страница 4: ...following procedure Step 1 Clean the tip with soap water ensuring that the mirror is clean and free of smudges stains or any milky haze Do not use an abrasive surface to clean the mirror Step 2 After...

Страница 5: ...Insert and seal the tip into a sterilization pouch Step 2 Sterilize the wrapped tip in a steam autoclave at a temperature of 132 C 270 F and 210kPa 30 46 psi for 6 minutes 132 C 270 F 210kPa 30 46psi...

Страница 6: ...Click the yellow SCAN icon in top right corner of screen 7 A B E Quantor3D will auto launch taking you to the scan window Scanning Step 2 Scanning Preparation Prepare the patient for the scan by ensur...

Страница 7: ...ns are obtained at a closer range and it is okay if the tip touches the teeth During scanning you should continue to hear a rapid clicking sound and observe a green rectangle indicating the last scann...

Страница 8: ...Rescan any area that is missing critical data for the success of the final restoration Save and proceed to the next step in the scan strategy Scanning Step 5 Exporting Files with Dentalshare STL After...

Страница 9: ...can scanning is a skill that requires practice Prior to use on a patient practice should be conducted using a dental typodont model Lower Upper Jaw Scans 1 Start the scan path on the working arch jaw...

Страница 10: ...g the buccal surface of the posterior teeth and the facial surface of the anterior teeth to complete your scan strategy 5 You can turn the scanner off at any point during the scan make sure you restar...

Страница 11: ...will ring and a green check mark will appear on the upper jaw image in the upper right corner of the application 2 Once the upper jaw is aligned move the handpiece down to the lower jaw Again move th...

Страница 12: ...d do not throw it away The box is optimal for any necessary transportation or shipping of the Heron IOS 3D Imaging and Simulation Corp Americas 22560 Glenn Drive Suite 116 Sterling VA 20164 USA Phone...

Отзывы: