Bite Alignment Procedure:

The bite alignment procedure will be necessary to align the upper and lower jaws. 

This process will only be allowed if lower and upper jaw scans have been previously 

conducted. Once the Bite button is pressed, it may take a few seconds to load the 

previous scans.

1. 

Start scanning from the back of the upper buccal area. Move forward and 

backwards along the molar and pre-molar area, making sure to scan some 

of the gingiva area and to tilt the handpiece upward slightly. When an initial 

alignment is found the jaw will be transformed into alignment with the bite 

scan. When a final alignment is found the application will “ring” and a green 

check mark will appear on the upper jaw image in the upper right corner of the 

application.

2. 

Once the upper jaw is aligned, move the handpiece down to the lower jaw 

Again, move the handpiece forward and backwards along the buccal of the 

molar and pre-molar areas, making sure to scan some of the gingiva area, and 

tilt the handpiece downward slightly.  The update process for the lower jaw 

is like that of the upper jaw, except when a final alignment is found a green 

check mark will appear on the lower jaw image in the upper right corner of the 

application.

3. 

When both jaws have been aligned, pause the scan and review the alignment. 

If occlusal between the upper and lower scans is observed, you can choose 

to restart the bite alignment process or export the scans to fine tune the 

alignment in a third-party application.

Содержание Heron IOS

Страница 1: ...g the Tip 2 Patient Environment 2 Tip Cleaning Sterilization 3 Scanning Step 1 Creating a Case 5 Scanning Step 2 Scanning Preparation 5 Scanning Step 3 Basic Scanning Procedure 6 Scanning Step 4 Final...

Страница 2: ...ng station on a flat stable surface and place the Heron IOS hand piece securely in its docking station Step 2 Connect the AC DC power adapter cable to the docking station the connector socket is locat...

Страница 3: ...ine extending 1 8 meter 6 feet around patient and system Depress bottom button 1 and rotate tip at 90 increments 2 Tip should be locked into position prior to scanning 1 2 1 2 Attach the tip to the ha...

Страница 4: ...following procedure Step 1 Clean the tip with soap water ensuring that the mirror is clean and free of smudges stains or any milky haze Do not use an abrasive surface to clean the mirror Step 2 After...

Страница 5: ...Insert and seal the tip into a sterilization pouch Step 2 Sterilize the wrapped tip in a steam autoclave at a temperature of 132 C 270 F and 210kPa 30 46 psi for 6 minutes 132 C 270 F 210kPa 30 46psi...

Страница 6: ...Click the yellow SCAN icon in top right corner of screen 7 A B E Quantor3D will auto launch taking you to the scan window Scanning Step 2 Scanning Preparation Prepare the patient for the scan by ensur...

Страница 7: ...ns are obtained at a closer range and it is okay if the tip touches the teeth During scanning you should continue to hear a rapid clicking sound and observe a green rectangle indicating the last scann...

Страница 8: ...Rescan any area that is missing critical data for the success of the final restoration Save and proceed to the next step in the scan strategy Scanning Step 5 Exporting Files with Dentalshare STL After...

Страница 9: ...can scanning is a skill that requires practice Prior to use on a patient practice should be conducted using a dental typodont model Lower Upper Jaw Scans 1 Start the scan path on the working arch jaw...

Страница 10: ...g the buccal surface of the posterior teeth and the facial surface of the anterior teeth to complete your scan strategy 5 You can turn the scanner off at any point during the scan make sure you restar...

Страница 11: ...will ring and a green check mark will appear on the upper jaw image in the upper right corner of the application 2 Once the upper jaw is aligned move the handpiece down to the lower jaw Again move th...

Страница 12: ...d do not throw it away The box is optimal for any necessary transportation or shipping of the Heron IOS 3D Imaging and Simulation Corp Americas 22560 Glenn Drive Suite 116 Sterling VA 20164 USA Phone...

Отзывы: