User Manual
FireCR Dental
TM-801-EN
37
5.2.
M
anufacturer’s Declaration - Electromagnetic Immunity
The FireCR Dental system is intended for use in the electromagnetic environment specified
below. The customer or the user of FireCR Dental system should assure that it is used in such
an environment
Immunity test
IEC 60601
Test level
Compliance level
Electromagnetic
Environment -guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
6 kV Contact
8 kV Air
6 kV Contact
8 kV Air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the relative
humidity should be at least
30%.
Electrical fast
Transient / burst
IEC 61000-4-4
2kV for power supply lines
1kV for input/output lines
2kV for power supply lines
1kV for input/output lines
Main power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
1 kV differential mode
2 kV common mode
1 kV differential mode
2 kV common mode
Main power quality should be
that of a typical commercial or
hospital environment.
Power frequency
(50/60Hz)
Magnetic field
IEC 61000-4-8
3.0 A/m
3.0 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
Voltage dips, short
Interruptions and
Voltage variations
on power supply
input lines
IEC 61000-4-11
<5%
U
т (>95% dip in
U
т)
for 0.5cycle
40%
U
т (60% dip in
U
т )
for 5 cycle
70%
U
т (30% dip in
U
т)
for 25 cycle
<5%
U
т (<95% dip in
U
т )
for 5 s
<5%
U
т (>95% dip in
U
т)
for 0.5cycle
40%
U
т (60% dip in
U
т )
for 5 cycle
70%
U
т (30% dip in
U
т)
for 25 cycle
<5%
U
т (<95% dip in
U
т )
for 5 s
Main power quality should be
that of a typical commercial or
hospital environment. If the
user of the BSVD-1000 system
requires continued operation
during power main
interruptions, it is
recommended that the
FireCR
Dental
system be powered
from an uninterruptible power
supply or a battery.