ATLAS THE ALS SIMULATOR P75
17 ATLAS THE ALS SIMULATOR P75
| FAQ
FAQ
My Atlas is unable to connect to REALITi?
Cause
Solution
Incorrect Wi-Fi Network selected
• Connect to the correct Wi-Fi on both iPads
NOTE:
Default SSID Wi-Fi “iSimulate-5.0”
Default password is “iSimulate2012”
• If operating more than one Atlas, ensure that each simulator has its own unique SSID name. If
operating more than one Atlas, ensure each simulator has its own router/network to connect to.
NOTE:
After 15 minutes of not being able to connect, Atlas will temporarily revert to the default SSID and
Password to allow you to rename the connection Wi-Fi and Password.
NOTE
:
Follow the procedure from the Quick Start Guide
(How to connect Atlas with REALITi360?)
REALITi 360 App is not updated
Go to the App Store on both your iPads and update the REALITi 360 app to the latest version
available
Atlas not turned on
Turn Atlas off and on again
Atlas battery low
Charge Atlas
Atlas too far from Wi-Fi
Move your network and REALITi system closer to Atlas
My connection with REALITi and Atlas is lost?
Multiple network connections may interfere with transmissions between REALITi and Atlas.
Forget other networks on both your iPads beside the network required to connect with Atlas. Default network is “iSimulate-5.0”
How to change the Wi-Fi setting on Atlas?
Refer to the ‘Additional Information (How to change the Wi-Fi setting on Atlas?).’
NOTE
: The steps are also available on page 15 of the Atlas manual.
Atlas airways have become contaminated from mouth-to-mouth resuscitation
DO NOT perform mouth-to-mouth on Atlas. If damage is reported:
• Clean the outside of Atlas with manikin wipes.
• Clean the inside of the oral cavity with manikin wipes.
• Change the lung bag.
No chest rise visible on Atlas
• If chest-rise is unilateral, ETT may be inserted into the right main stem bronchus. Correct the insertion of the ETT.
• Lung bag is leaking or tubing to lung is twisted, kinked, or disconnected for which the lung bag will need to be replaced.
• Seal the airway.
• Check air tubing for leakage.
• Check that all connections are intact.
• Replacing tubing if necessary.
Atlas carotid pulse can not be felt
The following may be the cause:
• The blood pressure is set below 50mmHg Systolic.
• The connection is lost between the REALITi 360 system and Atlas.
• Non-perfusing cardiac rhythm
• PEA rhythm