ATLAS THE ALS SIMULATOR P75

17  ATLAS THE ALS SIMULATOR P75 

| FAQ

FAQ

My Atlas is unable to connect to REALITi?

Cause

Solution

Incorrect Wi-Fi Network selected

• Connect to the correct Wi-Fi on both iPads

NOTE: 

Default SSID Wi-Fi “iSimulate-5.0”

            Default password is “iSimulate2012”

 

• If operating more than one Atlas, ensure that each simulator has its own unique SSID name. If 

operating more than one Atlas, ensure each simulator has its own router/network to connect to.

NOTE: 

After 15 minutes of not being able to connect, Atlas will temporarily revert to the default SSID and      

            Password to allow you to rename the connection Wi-Fi and Password.

  

NOTE

Follow the procedure from the Quick Start Guide  

            (How to connect Atlas with REALITi360?)

REALITi 360 App is not updated

Go to the App Store on both your iPads and update the REALITi 360 app to the latest version 

available

Atlas not turned on

Turn Atlas off and on again

Atlas battery low

Charge Atlas

Atlas too far from Wi-Fi

Move your network and REALITi system closer to Atlas

My connection with REALITi and Atlas is lost? 

Multiple network connections may interfere with transmissions between REALITi and Atlas.  

Forget other networks on both your iPads beside the network required to connect with Atlas. Default network is “iSimulate-5.0” 

How to change the Wi-Fi setting on Atlas?

Refer to the ‘Additional Information (How to change the Wi-Fi setting on Atlas?).’ 

NOTE

: The steps are also available on page 15 of the Atlas manual. 

Atlas airways have become contaminated from mouth-to-mouth resuscitation 

DO NOT perform mouth-to-mouth on Atlas. If damage is reported:

 

•  Clean the outside of Atlas with manikin wipes. 

•  Clean the inside of the oral cavity with manikin wipes.

•  Change the lung bag. 

 

No chest rise visible on Atlas

•  If chest-rise is unilateral, ETT may be inserted into the right main stem bronchus. Correct the insertion of the ETT. 

•  Lung bag is leaking or tubing to lung is twisted, kinked, or disconnected for which the lung bag will need to be replaced.

•  Seal the airway. 

•  Check air tubing for leakage.

•  Check that all connections are intact.

•  Replacing tubing if necessary.

Atlas carotid pulse can not be felt

The following may be the cause: 

•  The blood pressure is set below 50mmHg Systolic. 

•  The connection is lost between the REALITi 360 system and Atlas. 

•  Non-perfusing cardiac rhythm 

•  PEA rhythm

Содержание ATLAS THE ALS SIMULATOR P75

Страница 1: ...ATLAS THE ALS SIMULATOR P75 USER MANUAL 1023796 1024113...

Страница 2: ...Safety Instruction 5 Installation 5 Airway 6 Cardiopulmonary resuscitation CPR 8 Circulation 9 Network Router Settings 11 How to connect Atlas with REALITi360 11 How to change the Wi Fi setting on At...

Страница 3: ...s for advanced life support in simulation based training and further training DELIVERY CONTENT 1 2 3 4 5 6 7 8 9 11 10 13 1 Consumable bag 2 Atlas charger 3 Tape 4 Lubricant 5 Adapter 6 IV cannulation...

Страница 4: ...the monitor interface selected CPR dashboard screen for viewing all relevant parameters and values of cardiopulmonary resuscitation according to the AHA ERC Guidelines Atlas allows observation and rec...

Страница 5: ...stored and charged in dry rooms Operating temperature 10 C up to 35 C Storage temperature 10 C up to 35 C Air humidity 15 80 Only use the charger provided by the manufacturer Charger connection specif...

Страница 6: ...mulated dry aspiration removing inserted foreign bodies and inserting Guedel tubes OPA or Wendl tubes NPA When single assistant or dual assistant bag mask ventilation BVM is performed correctly on Atl...

Страница 7: ...eal intubation Supra glottic airway alternatives like iGel Laryngeal Masks LMA and Laryngeal Tubes LT can be used in accordance with manufacturers instructions Moreover a gastric probe can be inserted...

Страница 8: ...rrying out manual thoracic compressions and continuing resuscitation measures The CPR dashboard screen and Generic Defibrillator gives you feedback on the hand position With the use with of REALITi360...

Страница 9: ...top of the Atlas s left arm We recommend the use of drilling systems for this procedure When performing administration of medication via the IV pad and or IO access site use a drain bag which should...

Страница 10: ...and place the new training pad in the slot At the end of the training session we recommend drying removing the cannulas and taking the drainage system apart When the day s sessions are done flush the...

Страница 11: ...Control iPad to the Wi Fi network Both iPads must be on the same Wi Fi network iSimulate 5 0 If the mini router PIX Link is not included in the delivery content set up the Wi Fi net work with the REA...

Страница 12: ...hat appears select the corresponding REALITi monitor to your REALITi ecosystem If the monitor does not appear in the list check that the Control iPad and the Monitor iPad Pro are both con nected to th...

Страница 13: ...om the main menu on the Monitor iPad 10 On the Monitor iPad Pro the monitor selected from step 9 should appear Start the defibrillator 11 Swipe in from the right screen edge of the monitor iPad show t...

Страница 14: ...14 ATLAS THE ALS SIMULATOR P75 Connect Atlas with REALITi360 14 Compress the manikin You should be able to see the CPR compressions in the ECG waveform and the bar graph 13 The connected manikin will...

Страница 15: ...the delivery content of Atlas 2 Connect the REALITi Monitor iPad Pro to the Wi Fi network iSimulate 5 0 If the mini router PIX Link is not included in the delivery content set up the Wi Fi network wi...

Страница 16: ...e router of your choice Press the arrow button to save Please wait for 30 seconds Turn off the manikin and then power on again Now connect both iPads to the Wi Fi network of your choice The mini route...

Страница 17: ...t Multiple network connections may interfere with transmissions between REALITi and Atlas Forget other networks on both your iPads beside the network required to connect with Atlas Default network is...

Страница 18: ...meral IO replacement light skin dark skin 1024050 1024347 IV cannulation pad forearm light skin dark skin 1024047 1024345 IV cannulation pad hand light skin dark skin 1024048 1024344 Lung bag 1024056...

Отзывы: