background image

4

 

5.1  Current measurement 

 

The  nominal  voltage  of  the  voltage 
source  may  not  exceed  600  V.  Do  not 
make measurements in the low-voltage 
mains region. 

The multimeter is to be connected in series with 
the  load  on  the  side  where  the  voltage  with  re-
spect to ground is lowest. 

 

Before making any current measurements, set 
the mode switch to 

 or 

 as appropri-

ate. 

 

Connect  the  terminal  at  the  lower  potential 
to the earth socket. 

 

Currents  of  less  than  3  A  can  be  measured 
between  the  earth  socket  and  measuring 
socket "A". 

 

Currents  of  less  than  3  A  can  be  measured 
between  the  earth  socket  and  measuring 
socket  "A".  Currents  higher  than  3  A  should 
be  measured  between  the  earth  socket  and 
measuring socket "10A". 

 

Set  the  switch  to 

A

  mode  and  select  the 

desired measuring range. lf the current to be 
measured  is  unknown  beforehand,  set  the 
measurement  range  dial  to  the  highest 
range and work down. 

The fact that measurements can be made with-
out  disconnecting  the  instrument  and  using 
overload  protection  without  fuses  means  that 
converters  may  also  be  connected  into  the  cir-
cuit. 

 

V

A

10A

 

Fig. 1 Current measurement 

 

5.2  Voltage measurement 

 

The  nominal  voltage  of  the  voltage 
source  may  not  exceed  600  V.  Do  not 
make measurements in the low-voltage 
mains region. 

 

Before making any voltage measurements, set 
the mode switch to 

 or 

 as appropri-

ate. 

 

Use  the  "V"  socket  on  the  right  for  voltage 
measurements. 

 

Set  the  switch  to 

mode  and  select  the 

desired measuring range. lf the voltage to be 
measured  is  unknown  beforehand,  set  the 
measurement  range  dial  to  the  highest 
range  and  work  down.  The  100  mV  voltage 
range  is  associated  with  a  current  range  of 
0.1 mA. 

 

V

A

10A

 

Fig. 2 Voltage measurement 

 

5.3  Measuring  current  and  voltage  without 

reconnecting 

Isolated  current  and  voltage  sockets  allow 
measurements of both current and voltage to be 
made  in  sequence  without  disconnecting  the 
measuring  leads.  That  means  that  in  both  AC 
and  DC  modes,  resistance,  conductance,  im-
pedance  and  admittance  can  all  be  calculated 
forming the appropriate quotients. 

 

V

A

10A

 

Fig. 3 Simultaneous current and voltage measure-
ment 

 

5.4   Resistance and conductance 

According  to  the  defining  equation  for  a  linear 
resistance, 

R

 = 

V

/

I,

 or conductance 

I

/

V

, the 

Escola 10 can be connected in circuit as in Fig. 
3  so  that  current  and  voltage  can  be  measured 
one  after  the  other,  allowing  for  resistance 
measurements  to  be  made  over  a  range  from 
m

 up to several M

Therefore conductances in a range from under 1 
µS  up  to  30  S  can  be  measured  by  taking  the 
reciprocal of the resistance. 

One very key advantage of the Escola 10 is that 
the instrument, when connected as in Fig. 3, can 
measure  both  current  and  voltage  ranges  with-
out having to disconnect it. 

 

Содержание Escola 10

Страница 1: ...rements on experiments and laboratory apparatus It is not approved for measurements on low voltage distribution equipment such as sockets fuses etc CAT I Signal layer telecommunications elec tronic equipment with only minor excess voltage transients CAT II Domestic appliances mains sockets portable instruments etc CAT III Supply via buried cabling built in swit ches automatic circuit breakers sock...

Страница 2: ... not be used to make measurement on circuits which exhibit co rona discharge high voltage Particular care is to be taken when making measurements on high frequency circuits where dangerous voltages may arise due to superimposition of components Make very sure that the voltage value be tween the measured contact and earth or between the ground socket and the meas urement socket does not exceed 1000...

Страница 3: ...ides for measurements with the needle centred on the dial for DC quantities Resistance R and con ductance G or impedance Z and admittance Y can be obtained by division All measurement ranges are selected by means of the measurement range dial Two linearised mirror scales graded in a 1 3 ratio guarantee easy read ability of the measured values The equipment is turned on by means of the operating mo...

Страница 4: ...ore making any voltage measurements set the mode switch to or as appropri ate Use the V socket on the right for voltage measurements Set the switch to V mode and select the desired measuring range lf the voltage to be measured is unknown beforehand set the measurement range dial to the highest range and work down The 100 mV voltage range is associated with a current range of 0 1 mA V A 10A Fig 2 V...

Страница 5: ...se it to move to the left The scales are la belled accordingly smaller auxiliary numbering 5 7 Calibrating zero point Turn on the multimeter set the maximum voltage range of 600 V and connect the ground socket to the voltage socket using a short lead Set up the multimeter in a horizontal position and correct the zero point of the dial with the help of the adjustment knob 5 8 Switching off Always s...

Страница 6: ...ve Dirt or moisture in the measurement sockets can affect readings Shake out any dirt that may be in the meas urement sockets Soak a new swab with isopropyl alcohol and work around the inside of each measure ment socket 7 Disposal The packaging should be disposed of at local recycling points Should you need to dispose of the equip ment itself never throw it away in normal domestic waste Local regu...

Отзывы: