3. Introduction  

3B NET

log

™ is a multimedia data acquisition 

and evaluation system for current and voltage 
measurements and for measurements using 
sensors. The system can be used with or with-
out being connected to a computer. With the 
accompanying 3B NET

lab

™ software, meas-

urements can be conducted with arbitrarily 
chosen parameters as well as in prepared 
experiments. The prepared experiments take 
the user through an interactive experimental 
environment where the measurement parame-
ters have already been defined. When used-
within a network, the teacher and students can 
mutually monitor their results. 
The Ethernet port allows the 3B NET

log

™ unit 

to be directly connected to a local network, so 
that data can be sent to any computer in the 
network for evaluation. In order to use the unit 
with Ethernet an intranet in the same sub-
network must be in place. 
Without being connected to a computer, the 
3B NET

log

™ device can be used as a digital 

multimeter for current and voltage measure-
ments. Combined with different sensors, the 
3B NET

log

™ equipment can also be used as a 

handheld unit with automatic sensor recogni-
tion.  

 
 

4. Technical specifications  

4.1  Analogue inputs  

 

Voltage inputs (channels A and B):  
Measuring equipment:  2 Differential amplifi-
ers  
Measuring range: 

±200mV, ±2V, ±20V 

Surge voltage protection: 

Max. ±40V 

Connections: 

4-mm safety sockets  

 

Current input (channel A):  
Measuring range: 

±200mA, ±2A 

Excess current protection:  Max. ±2.5A 
Connections: 

4-mm safety sockets  

 

Sensor inputs (channels A and B):  
Sensor type: 

Analogue  

Sensor identification:  Automatic  
Sensor connections: 

8-pin mini DIN sockets  

Sampling: Continuous 

 

Sampling rate: 

50k samples/s  

Resolution: 

12 bit  

 

 

 

 

4.2  Analogue outputs (channels A and B)  

Reference point (earth): Common  
Measuring range: 

± 5V  

Connections: 4-mm 

safety 

sockets 

and 8-pin mini DIN 
sockets  

Sampling rate: 

10kSamples/s  

Resolution: 12bit 

 

 

4.3  Digital inputs  

Channels: 

4 (separated into 2 
TTL inputs, one of 
which is a high-speed 
time input, plus 2 in-
puts via optocoupler)  

Sampling rate: 

50k samples/s 
100k samples/s (high-
speed time input)  

Connection: 

8-pin mini DIN socket 

 

4.4 Digital outputs  

Channels: 6 

 

Signal: TTL 

 

Connection: 

8-pin mini DIN socket  

 

4.5 Additional data  

Computer connection:  USB  
Network port: 

Ethernet 

Internal data memory:  128k 
Display: 

64 x 122 matrix for 
measurements and 
units  

Power supply: 

4.5 V DC/300mA or 3 
batteries (AA, LR6 or 
AM3); owing to their 
longer battery life, use 
of alkaline batteries is 
recommended.  

Dimensions: 

21 x 8 x 4 cm

3

 approx. 

Weight: 

400g approx. (includ-
ing batteries) 

 
 
 
 
 
 
 
 
 
 
 
 
 
 

Содержание 3B Netlog 1000541

Страница 1: ...g device does not come into contact with water The 3B NETlog device should not be exposed to temperatures above 80 C The operating voltage input is protected against polarity reversal However it is no...

Страница 2: ...1 Analogue inputs Voltage inputs channels A and B Measuring equipment 2 Differential amplifi ers Measuring range 200mV 2V 20V Surge voltage protection Max 40V Connections 4 mm safety sockets Current...

Страница 3: ...lighting on and off briefly press the button Date Time button Switching on the battery and temperature displays Switching on the time display Switching on the date display Selection of operating mode...

Страница 4: ...a a computer never press the buttons on the 3B NETlogTM console 7 1 Driver installation Before installing the 3B NETlab software it is important to install the USB driver Insert the installation CD in...

Страница 5: ...r Click Yes to confirm that the file should be overwritten 7 2 Software installation Note on installing the software For this software please refer to the installa tion instructions available in the 3...

Страница 6: ...ETlog equipment is ready to conduct measurements 8 6 Setting the sampling rate Press the Rate button the options list for SAMPLE RATE will appear Select the desired sampling rate with the Rate or Stor...

Страница 7: ...button until the display shows the following Use the Date Time button to confirm this then plug in the Ethernet cable to connect the 3B NETlog unit to the local network Now you can modify the known fi...

Страница 8: ...ince the beginning of the read opera tion Time remaining shows the approximate time remaining for the read operation Options the following settings can be made here Decimal separator sets the decimal...

Страница 9: ...o search and install older firmware versions 12 Format of exported files Exported files have the following format the words shown here between the delimiters are simply placeholders for actual data En...

Страница 10: ...ype tested EN 61326 1 Tested for noise immunity and interference Conformity with EU guidelines is indicated by the CE marking on the device 16 Licence 3B Netlog and 3B Netlab are registered trademarks...

Страница 11: ...3B Scientific GmbH Rudorffweg 8 21031 Hamburg Germany www 3bscientific com Subject to technical modifications Copyright 2014 3B Scientific GmbH...

Страница 12: ......

Отзывы: