background image

4

Understanding Your Device

The Lumin operates using UV-C light. The drawer is made 

of polished aluminum which reflects the UV light to improve 

coverage. The Lumin operates using a form of high energy light 

that is focused narrowly on the wavelength that is most effective 

in reducing bacteria* and viruses.* 

5

Using Your Lumin

1. Plug the Lumin power cord in a wall outlet.
2. Place the article face up on the acrylic rack.
3. If the article is visibly soiled please first wipe thoroughly with a 

wipe.

4. Close the Lumin chamber drawer.
5. Press round power button, the Cycle-On orange light will turn 

on indicating that the 5 minute cleaning cycle has started.

6. After 5 minutes the Lumin will emit a beep and the UV light will 

no longer be visible. A green light will indicate completion of 

the cleaning cycle.

No waiting period is necessary

and your article is now ready to use.

Caring for Your Lumin Device

The Lumin bulb is designed to exceed the life of your device 

so changing the bulb should not be necessary. If, however, 

it does become necessary, the bulb is removed by simply 

grabbing the base and pulling firmly. The bulb is easily replaced 

by inserting the new bulb and pushing it into the socket. If you 

are in need of a replacement bulb please contact 3B Medical 

at 863-226-6285 for a replacement bulb. Alternatively you may 

also log on to www.3BLumin.com and order a bulb or email 

[email protected]

The Lumin drawer is lined with polished aluminum. Do not 

use baking soda or other alkali-based cleaners on aluminum 

as it can cause discoloration. Aluminum scratches fairly easily, 

so avoid abrasive cleaners and scrubbers when cleaning and 

polishing. If using a metal or glass cleaner, test the cleaner on 

a small portion of the drawer. Polish with a soft lint free cloth. 

Should the casing of your 3B Lumin become soiled you may 

wipe down with any moist cloth. Do not immerse your Lumin 

into water or any cleaning solvents. Place your Lumin on a level 

counter or tabletop. Do not drop your Lumin or the bulb and or 

glass drawer may shatter.

SKU:  

LM3000

Weight:  

5.5 lbs

Dimensions:  

12.25 x 8.5 x 7.75”

Voltage: 

AC 110 V   0.2A 60Hz

 

AC 220 V   0.15A 50Hz

Fuse: 

250 V, 2A, Interrupting

 

Capability: 50A

TUV Lamp Useful Life:  9000 hr
Lamp Wattage: 

9 W

UV-C Radiation: 

2.3 W

Environmental Conditions:

      Storage Temperature: 0˚ to 120˚ F / -18˚ to 49˚ C

      Operating Temperature: 20˚ to 100˚ F / 7˚ to 38˚ C

      Humidity: Up to 93%, non-condensing

      Atmospheric Pressure: 800-1060hPa

      Altitude: 2000m

SPECIFICATIONS

Lab Tested on Hard Non-Porous Surfaces:
*  Bacteria: E. coli 99.99%, S. enterica 99.99%, S. aureus 99.99%. K. pneumoniae 99.99%
Lab Tested on Porous Soft Surfaces:
**  Viruses: Influenza A (H1N1) >99.9% and Human Coronavirus 229E >99.9%,   

 

Human Adenovirus 2 (ATCC VR-846) >99.9%