Yellow Light
• Low O
2
• Warning status
• Cartridge replacement
• Cannula missing
A yellow indicator light alerts you to a warning status If no cannula is
detected a solid yellow light will show Low O
2
purity between 70%-
85% the yellow indicator light will blink while inhaling, the Aer X unit
IS usable during this time, but will need to be checked by a service
professional Cartridge Replacement needed-the yellow light will blink
and a message will show “change cartridge” in the machine log. Light
will blink twice and beep if no breath is detected
Red Light
• Error status unit requires
servicing
• Low battery (shut off
impending)
• Unit is overheating
Solid red indicator light indicates the Aer X unit cannot complete a
cycle and may not be delivering oxygen, this requires urgent servicing
Very low O
2
purity >70% will cause the Aer X to blink a red indicator
light 4 times while simultaneously beeping. Critically low battery power
will cause the red-light indicator to blink and beep 7 times prior to the
Aer X shutting down
27