6.  Position the nasal cannula on your face and breathe through your nose.

If no breath is detected for 3 minutes, your Aer X will power off.

The Auto X will sense the onset of inhalation 

and deliver a burst of oxygen at a precise time 

when you inhale. The Auto X will sense each 

breath and continue to deliver oxygen in this 

manner. As your breathing rate changes, the 

Auto X will sense these changes and deliver 

oxygen only as you need it. At times, if you 

inhale very quickly between breaths the Auto 

X may ignore one of the breaths, giving the 

appearance of a missed breath. This may be 

normal as the Auto X senses and monitors the 

changes in your breathing pattern. The Auto X 

will normally sense the next breath and deliver 

oxygen accordingly.
Make certain that the nasal cannula is properly 

aligned on your face and you are breathing 

through your nose.

18

Содержание Aer X

Страница 1: ...myAerX com USER MANUAL ...

Страница 2: ...erating Instructions 16 General Instructions 16 Auto X 19 Traveling by Air 21 Battery Operating Instructions 24 Battery Care and Maintenance 25 4 Audible and Visible Signals 26 5 Troubleshooting 28 6 Cleaning Care and Maintenance 30 7 Symbols Used on Concentrator and Accessories 33 8 System Specifications 35 9 Limited Warranty 43 Table of Contents ...

Страница 3: ......

Страница 4: ... restricts this device to sale by or on the order of a physician May also be applicable in other countries CAUTION Use of a cannula other than a 3B approved may restrict oxygen delivery and or attachment to cannula adapter WARNING Availability of an alternative source of oxygen is recommended in case of power outage or mechanical failure Consult your equipment provider for type of back up system r...

Страница 5: ...adaptor away from children and pets CAUTION The Aer X is not a Toy so don t let children or pets touch or turn on the device CONTRAINDICATIONS 5 WARNING This device is NOT INTENDED to be life sustaining or life supporting CAUTION Under certain circumstances the use of non prescribed oxygen therapy can be hazardous This device should be used only when prescribed by a physician CAUTION Additional mo...

Страница 6: ...6 WARNING If you begin to feel ill or are experiencing discomfort while using this device consult your physician immediately CAUTION The Aer X is designed to provide a flow of high purity oxygen An advisory yellow light will inform you if oxygen concentration drops below 85 If the yellow light persists contact your equipment provider See the section Indicator Lights for additional information CAUT...

Страница 7: ...2 MAIN COMPONENTS AND FUNCTIONS Magnetic Cannula Port for Safety Full Color LCD Screen Indicator Lights Offer Feedback on Your Device Battery Door Easy Press Buttons to Control Your Aer X Device 7 ...

Страница 8: ...ress to turn OFF MENU SELECT UP Use the Up and Down arrow to select your O2 pulse setting Settings are available in 1 5 lpm increments Please do not change settings unless you have been advised by a medical professional DOWN DISPLAY BACKLIGHT 8 USER CONTROLS ...

Страница 9: ...n Please See Troubleshooting Chart on page 28 for more details Audible Signals An audible signal beep indicates either a change in operating status or a condition that may need response alarm Your 3B Aer X will sound a long beep upon start up and shut down When you press menu buttons a short audible beep will sound Turning the volume off or low may interfere with alerts and alarms please use cauti...

Страница 10: ...INPUT OUTPUT CONNECTIONS Inlet Air Filter Cartridge The user changeable inlet air filter should be checked and cleaned weekly The Inlet air filter should be changed every 6 months or as needed 10 ...

Страница 11: ... cannula adapter connects to the top port on the 3B Aer X The port outputs oxygenated air The port is magnetized and will detach from the device for your safety DC POWER IN Connection for external power from the Power Supply 11 ...

Страница 12: ...nal power source When fully charged a single battery will provide approximately 4 hours of operation at a setting of 2 Recharging time is up to 3 hours for a fully depleted battery See information in the Battery Care and Maintenance section 12 WARNING Check Battery case before use don t use cracked case ...

Страница 13: ...s from 100V to 240V 47 63HZ permitting use with most power sources throughout the world Battery Charger The 3B Medical Battery Charger is used to charge the 3B Aer X concentrator battery s from an AC power source This allows the user to charge batteries where AC power is available The Aer X Battery Charger is specifically designed for use with the Aer X Oxygen Concentrator Batteries The Battery Ch...

Страница 14: ...nd accessories The Carry Case is specifically designed to allow proper airflow around the Aer X and battery during operation Use caution when utilizing any other carrying case or pack with your Aer X device improper venting may cause a fire hazard Adsorbent Cartridges The Aer X has patient changeable sieve bed cartridges These cartridges siphon out the oxygen in the room air to provide you with co...

Страница 15: ...lear access Place the 3B Aer X in such a way that auditory alarms may be heard WARNING Avoid use of the Aer X in the presence of pollutants smoke or fumes Do not use the Aer X in the presence of flammable anesthetics cleaning agents or other chemical vapors CAUTION Do not obstruct air intake or exhaust when operating the device Blockage of air circulation or proximity to a heat source may lead to ...

Страница 16: ... Make sure the Power Supply cools down before handling CAUTION The Power Supply is not intended to be used near water sources Doing so will put you at risk of damage to your unit and electric shock CAUTION Do not disassemble the Power Supply This may lead to component failure and or safety risk CAUTION Do not place anything in the Power Supply port other than the supplied wall cord Avoid the use o...

Страница 17: ...and the Aer X will begin to deliver oxygen Set the Aer X Concentrator to the output setting prescribed by your physician or clinician Use the arrow keys to adjust the Aer X to the desired output setting The current setting can be viewed on the display CAUTION To facilitate oxygen flow ensure the nasal cannula is properly connected to the port the port is properly connected to the Aer X and that th...

Страница 18: ...ner As your breathing rate changes the Auto X will sense these changes and deliver oxygen only as you need it At times if you inhale very quickly between breaths the Auto X may ignore one of the breaths giving the appearance of a missed breath This may be normal as the Auto X senses and monitors the changes in your breathing pattern The Auto X will normally sense the next breath and deliver oxygen...

Страница 19: ...se your oxygen intake according to those needs 0 200 400 600 800 1000 1200 1400 10 15 20 25 30 O2 minute cc Min Delivered in pulse each inhale Respiratory Rate BPM Breaths per Minute Aer X Active X Oxygen delivered vs Respiratory Rate Setting 1 Setting 2 Setting 3 Setting 4 Setting 5 19 Aer X Auto X Oxygen delivered vs Respiratory Rate ...

Страница 20: ...7 Aer X Screen Options Start Up Screen Main Screen Settings Menu Options Aer X Device Info 20 ...

Страница 21: ...aximum flow rate corresponding to the pressure in the cabin under normal operating conditions A new statement will not be necessary each time you fly but the statement should be available during every flight Some airlines may equip their aircraft with onboard electrical power You may have an opportunity to request a seat with a power port which can be used to power your Aer X However availability ...

Страница 22: ...ghts on regional airlines you will need sufficient battery power for the duration of the flight plus a conservative estimate of unanticipated delays CAUTION Airlines may not be equipped to provide backup oxygen Arrive at the airport early Airport Security screening personnel may require extra time to inspect your Aer X While waiting to board your flight you may be able to conserve battery power by...

Страница 23: ...ine personnel to ensure compatibility During taxi takeoff and landing stow your Aer X under the seat in front of you Your Aer X will fit upright under most airline seats However if it doesn t fit you may turn it on its side It is not necessary to turn off your Aer X during taxi takeoff and landing if your physician s written statement requires you receive oxygen during these periods 23 ...

Страница 24: ...ttery life is low do one of the following Plug the Aer X into an AC power source using the Power Supply wall or car outlet Replace the battery with a charged battery after turning off the Aer X by pressing the ON OFF button If the battery is drained charge the battery Normal Battery Charging To ensure that your battery is properly charging make sure the battery charger is properly connected to the...

Страница 25: ...displays battery level remaining The on screen battery display reflects the battery power left in the battery The Battery Level Display will show a circle with a value of 10 100 At 10 battery life the RED INDICATOR light will blink on and off as a warning When the battery life reaches a critically low level the Aer X device will beep and the indicator light will blink red 7 times This is to alert ...

Страница 26: ...ings Indicator Lights Troubleshooting Green Light Device Operating Properly Blinks on start up Blinks during oxygen bolus delivery Green light will blink on start up and during Oxygen inhalation The Green light indicates your Aer X is running normally and no issues are detected 26 ...

Страница 27: ...ight will blink and a message will show change cartridge in the machine log Light will blink twice and beep if no breath is detected Red Light Error status unit requires servicing Low battery shut off impending Unit is overheating Solid red indicator light indicates the Aer X unit cannot complete a cycle and may not be delivering oxygen this requires urgent servicing Very low O2 purity 70 will cau...

Страница 28: ...Solution Device does not operate Unknown mechanical electrical failure Check charge battery Change battery Use power supply Contact service provider check machine log if screen turns on Battery low Seat battery correctly Battery not installed correctly 28 ...

Страница 29: ...kips a pulse dispense of oxygen Battery pack appears to have no power Did not sense inhale Battery completely discharged Adjust cannula Physically check battery gauge on bottom of battery Breathing faster than 30 times per minute Cannula disconnected 29 ...

Страница 30: ...up to 4 feet in length is recommended to ensure proper breath detection and oxygen delivery CAUTION Use of a cannula other than what is recommended by 3B Medical may restrict oxygen delivery and may cause impurities in the delivered oxygen Carry Case and Device Case Cleaning Turn off and disconnect power before cleaning replacing parts You may clean the outside case and the carry case using a clot...

Страница 31: ...lean the particle filter more often in dusty environments To purchase additional particle filters contact your equipment provider or 3B Medical WARNING Do not disassemble the 3B Aer X or any of the accessories or attempt any maintenance other than tasks described in the troubleshooting section disassembly creates a hazard of electrical shock and will void your warranty Do not remove tamper evident...

Страница 32: ... EQUIPMENT AND ACCESSORIES Follow your local governing ordinances for disposal and recycling of the 3B Aer X and accessories MAINTENANCE ITEMS LIST Aer X battery Replacement inlet air filter Adsorbent cartridge 32 ...

Страница 33: ... in significant injury CAUTION A caution indicates that a precaution or service procedure must be followed Disregarding a caution could lead to a minor injury or damage to equipment See User Manual for Instructions Risk of Electrical Shock U S Federal Regulation Restricts this Device to Sale by Order of Physician May also be applicable in other Countries AC Power DC Power only 33 ...

Страница 34: ...e Battery Keep Dry Indoor or Dry Location Only Do Not Get Wet Use No Oil or Grease Do Not Disassemble Do Not Dispose of In Unsorted Municipal Waste Type BF Applied Part Not Intended for Cardiac Application Class II Device Complies with Applicable EU Directives Including Medical Device Directive 34 ...

Страница 35: ...ttery L W H 7 22 X 8 38 X 2 64 Weight 4 25Ib 3 5Ib without battery Noise 40 dBA Warm Up Time 2 minutes Oxygen Concentration 87 to 94 on all settings Output Settings 5 settings 1 to 5 8 SYSTEM SPECIFICATIONS continued 35 ...

Страница 36: ...utput 15 VDC 5 35A max Rechargeable Battery Voltage 10 4 12 6 VDC Battery Duration Up to 4 hours at output setting of 2 Battery Charging Time Up to 3 hours Environmental Ranges Intended for Use Temperature 38 ºF to 100 ºF 2 ºC to 38 ºC Humidity 0 to 60 non condensing Altitude 0 to 10 000 ft 0 to 3 048 meters 36 ...

Страница 37: ...on condensing Altitude 0 to 10 000 ft 0 to 3048 meters Transportation Keep Dry Handle with Care SETTING DISPENSE VOLUME CC MAXIMUM BREATH RATE BPM 1 11 30 2 22 30 3 33 30 4 44 30 5 55 30 setting of 4 30 pm dispenses 48 The Aer X will wait for the next breath when maximum breath rate is exceeded Will slightly reduce the O2 output 37 ...

Страница 38: ...e BPM Breaths per Minute Aer X Pulse Volume Setting vs Respiratory Rate Setting 1 Setting 2 Setting 3 Setting 4 Setting 5 0 10 20 30 40 50 60 10 15 20 25 30 O2 Delivered per pulse cc Breath Delivered in pulse each inhale Respiratory Rate BPM Breaths per Minute Aer X Pulse Volume Setting vs Respiratory Rate Setting 1 Setting 2 Setting 3 Setting 4 Setting 5 ...

Страница 39: ...inst Electrical Shock Class II Degree of Protection Against Electrical Shock Type BF Not intended for cardiac application Degree of Protection Against Ingress of Water IP22 Degree of Safety for Application in Presence of Anesthetic Gases Not suitable for such application 39 ...

Страница 40: ...ere is no guarantee that interference will not occur in a particular installation If this equipment interferes with other devices which can be determined by turning the equipment off and on the user is encouraged to try to correct the interference by one or more of the following measures 1 Reorient or relocate the receiving device 2 Increase the distance between the equipment 3 Connect the equipme...

Страница 41: ...creased emissions and decreased immunity of the equipment or system 41 GENERAL Keep the Aer X inside the Carry bag at all times as WARNING it works like a barrier from any excessive heat don t wear the carry bag while connecting the Aer X to the power supply to charge the battery During the running of the Aer X especially outdoor or in up normal environment the patient doesn t need to touch hot pa...

Страница 42: ...ents including domestic establishments and those directly connected to the public low voltage voltage power supply network that supplies buildings used for domestic purposes Harmonic Emissions IEC 61000 3 2 Not Applicable Voltage Fluctuations Flicker Emissions Not Applicable Guidance and Manufacturer s Declaration Electromagnetic Emissions The Aer X Oxygen Concentrator is intended for use in the e...

Страница 43: ...thority to determine if misuse or neglect is resulting in a fraudulent replacement If a confirmed malfunction or failure occurs it is at 3B Medical s discretion to provide a new or refurbished unit as a replacement All refurbished units are professionally handled and are in like new condition Warranty begins from date of shipment The malfunctioning unit MUST be shipped within 30 days of receipt of...

Страница 44: ...member to always consult a physician prior to using any medical device the information provided herein is not to be construed as a medical diagnosis or medical advice No medical claims are implied or intended with the sale of this device Purchaser accepts responsibility for correct and proper use of device www 3Bproducts com ...

Отзывы: