CLEANING AND DISINFECTION OPERATIONS

•  Detach the long sterilisable silicone tubes Ø 8 x 14 - long (

5.

) from the vessel (

4.

). Disconnect the sterilisable silicone tubes Ø 8 

x 14 - short (

3.

) from the respective vessels and from the respective antibacterial filters and empty the vessel/s.

 The liquids col-

lected in the vessels must be disposed of in the biological waste container.

 

•  In order to clean the accessories, rinsing them under hot running water is sufficient. Before performing the vessel cleaning opera-

tions empty their content into the biological material vessels and clean them by sterilisation in an autoclave with a cycle at 121°C. 
The silicone aspiration tubes and aspiration fitting can be sterilised in an autoclave with a cycle at 121°C. The complete vessel 
should be replaced every 30 sterilisation cycles.During the sterilisation operations the operator must use latex gloves and suitable 
clothing in order not to come into contact with any possibly contaminating substances. 

•  The disposable antibacteria filter (

2.

) must be replaced after every application.

•  Never wash the device under water or by immersion; clean the outside of the device using a cloth moistened with detergent (non 

abrasive).

PERIODICAL CHECK FOR THE SAFETY OF THE DEVICE

MAXIASPEED

 does not require maintenance and/or lubrication. Nevertheless, several simple several simple checks must be performed 

before every use:
•  Check the integrity of the casing, power supply cable and the pedal switch, if supplied.
•  Block the aspiration connector with a finger and check that the vacuum level reaches -0.9 bar.
•  Check that there are no disturbing noises suspicious of malfunctioning.
•  Check that the cage is positioned correctly inside its housing. It must be in axis with the plug aspiration hole so that the floating valve 

can block it in the circumstance in which the aspirated liquid exceeds the maximum volume the vessel can receive.

 Do not connect 

the cage as shown in figure 3b; it must be connected as shown in figure 3a.

•  Check that the float is fitted in the correct position and that it is free to move inside the cage (dirt or encrustation may hinder its 

movement). Figure 4 shows the correct insertion.

•  Check that the plug seal is inserted with the edging as illustrated in figure 5.

PROBLEMS

POSSIBLE CAUSES

SOLUTIONS

Excessive noise  

Damaged pump or obstructions in the 
internal suction duct

Send to the customer care service

The unit turns on but it does not suck

- Damaged pump
- Vacuum regulator fully open.
Connection tubes disconnected
and/or badly connected, broken
connection tubes. Vessel not in a
vertical position, full, or defective
overflow valve. Possible blockage of
the hydraulic circuit inside the unit

- Send to the customer care service.
- Check the position of the vacuum 
regulator. Check the connections and the 
integrity of the tubes.
Position the vessel in a vertical position, 
check the overflow valve (blocked) and/or 
replace the silicon tubes

The vacuum value cannot be adjusted

Damage to the internal hydraulic circuit or 
obstruction of the connection tubes to the 
aspiration unit

Send to the customer care service

The protection fuse is activated any time 
the device is turned on

Pump damaged or in short circuit

Send to the customer care service

The vacuum gauge does not work

Liquid penetrating the pneumatic
circuit

Send to the customer care service

PROBLEMS, CAUSES and SOLUTIONS

Note: if you experience faults or malfunctioning problems different to those listed above, always and exclusively 
contact authorised assistance centres.

Содержание MAXIASPEED

Страница 1: ...1 MAXIASPEED ...

Страница 2: ...LOTTEUR CAGE SCHWIMMER KORB FLOTADOR JAULA 1 5 1 6 1 2 1 3 1 10 1 9 1 8 Fig Abb 1a Fig Abb 1b Fig Abb 3a Fig Abb 3b Fig Abb 5 OK KO GABBIA IN ASSE CAGE IN AXIS CAGE DANS L AXE FLUCHTENDER KORB JAULA EN EJE GABBIA FUORI ASSE CAGE OUT OF AXIS CAGE HORS AXE NICHT FLUCHTENDER KORB JAULA FUERA DE EJE 1 4 2 1 6 ...

Страница 3: ...e prise appropriée 1 7 Roulettes antistatiques avec frein 1 8 Prise d alimentation avec fusible pouvant être inspecté 1 9 Compartiment porte câble et interrupteur à pédale 1 10 Prise pour interrupteur à pédale quand il existe 2 Filtre antibactérien 3 Tuyau court Ø 8x14 mm 4 Récipients 2000cc or 4000cc 4 1 Prise VACUUM 4 2 Prise PATIENT 5 Tuyau long Ø 8x14 mm 6 Raccord d aspiration 7 Outil pour poi...

Страница 4: ...s for use The warnings and cautions described in the instruction manual must be observed IMPORTANT WARNINGS This is a medical device and it must be used by authorised personnel It must be operated as instructed in this use manual It is important for the operator to read and understand the use and maintenance information of the device Contact your retailer for any questions you may have MICROBIAL C...

Страница 5: ...on vessel is complete with overflow valve lid and vessel in a clear mate rial polycarbonate The antibacterial filter also protects the suction circuit from possible contaminants sucked in during use 1 3 Connect the long tube to the PATIENT connector 4 2 of the selected plug 2 Operation of the telescopic handle for space reasons the aspirator is supplied with the telescopic handle lowered to raise ...

Страница 6: ... cage is positioned correctly inside its housing It must be in axis with the plug aspiration hole so that the floating valve can block it in the circumstance in which the aspirated liquid exceeds the maximum volume the vessel can receive Do not connect the cage as shown in figure 3b it must be connected as shown in figure 3a Check that the float is fitted in the correct position and that it is fre...

Страница 7: ... conditions Class of risk according to the 93 42 EEC directive Storage conditions Operating storage air pressure min 690 hPa max 1060 hPa Double pump version High flow high vacuum device 230V 50 60Hz 330VA approx 90 l min approx 50 dBA 1 m T2 5A 250V IIa Temperature min 10 C max 40 C Humidity min 10 max 95 Temperature min 25 C max 70 C Humidity min 10 max 95 approx 12 kg approx 13 kg approx 16 kg ...

Страница 8: ...13 ...

Страница 9: ...29 ...

Страница 10: ...30 ...

Страница 11: ...31 ...

Страница 12: ... in your area or returning it to the distributor when buying a new device of the same type to be used with the same functions This procedure of separated collection of electric and electronic devices is carried out forecasting a European environmental policy aiming at safeguarding protecting and improving environment quality as well as avoiding potential effects on human health due to the presence...