The 

MAXIASPEED

 aspirator is a professional device suitable for outpatient use, specifically for the aspiration of secretions. It is also suit-

able for use in hospital wards, on tracheotomy patients, or for surgical applications. It is equipped with 4 anti-static wheels (1.7) with a brake 
device, a telescopic pulling handle (1.1), vacuum regulator (1.3), vacuum gauge (1.2) and two vessels of 2 L. or 4 L. (depending on version). 
The vessels feature a protection device against liquid entering the vacuum pump, which interrupts the aspiration flow. It is lubrication free, 
easy to handle, simple to use, reliable, strong and quiet. The 

MAXIASPEED

 aspirator also has a switch (1.4) for switching the aspiration 

flow from one vessel to the other, very easily and without having to touch any possibly contaminated parts. The 

MAXIASPEED

 aspirator 

has the following accessories: 2 Vessels of 2 L. or 4 L. with protection device, power supply cable, 2 sterilisable silicone connection tubes Ø 
8x14 - short, 2 sterilisable silicone connection tubes Ø 8x14 - long, 2 aspiration fittings, 2 antibacteria filters, pedal switch (for versions with 
the relevant socket only). 

N.B.: Only use genuine accessories supplied by 3A Health Care; the accessories supplied with the 

device by the manufacturer have been tested and proven compliant with the latest current safety standards. WARNING! 
Correct operation of the device is not guaranteed in the event of use of accessories other than those supplied.

INTENDED USE

Mains-operated medical /surgical aspirator intended for outpatient use.

Medical purposes: 

This product is intended for use for the aspiration of body fluids.

Intended users of the product:

•  Legally certified medical personnel (doctors, nurses and therapists).
•  Users must also be able to understand the operation of the medical device, and the contents of the instruction manual, in general terms.

Intended patients for the product: 

People who need to remove body fluids (saliva, blood, etc.).

Environment: 

This product is intended for outpatient use.

Expected duration: 

Duration may vary based on the operating environment. The lifetime of the device is 5 years and that of the col-

lection vessel and the silicone tubes 1 year or 30 sterilisation cycles. The antibacterial filter is disposable device and must therefore be 
replaced after each application. Frequent use of the product may shorten the duration.

Precautions for use: 

The warnings and cautions described in the instruction manual must be observed.

  IMPORTANT WARNINGS

This is a medical device and it must be used by authorised personnel. It must be operated as instructed in this 
use manual. It is important for the operator to read and understand the use and maintenance information of the 
device. Contact your retailer for any questions you may have. MICROBIAL CONTAMINATION: where pathologies 
are present with microbial infection and contamination risks, perform the cleaning and disinfecting operations 
scrupulously after every use on the patient.  

The manufacturer ensures that everything necessary is done to ensure that every device is of the highest quality and safety. Neverthe-
less, basic safety rules must always be followed, as for every electrical device. 
•  The device must only be used by specialised personnel.
•  This device must be exclusively used for the purpose for which it was designed; in particular as a surgical aspirator. All other uses 

are improper and dangerous, and the manufacturer cannot be held responsible for any consequences deriving from improper use.  

•  Never use adaptors for power supply voltages different to those shown on the plate data label on the back of the device.
•  Keep the cable far away from hot surfaces.
•  The device is not suitable for use in the presence of anaesthetic mixes inflammable with air, or oxygen, or nitrogen oxide.
•  Never handle the power supply plug with wet hands.

Never leave the device close to water; do not submerge it in liquids, dot not wet it. If this should occur, pull the plug out of the 
power supply socket immediately, before touching the device. Do not use the device if the plug or power supply cable are dam-
aged, or wet (send it to your retailer immediately).

•  The device casing is not protected against the entry of liquids.
•  Always take out the power supply plug immediately after use.
•  Maintenance and/or repair must only be performed by authorised personnel. Unauthorised repairs invalidate the guarantee.
•  Ensure that the connections and vessel closing is performed carefully to prevent aspiration leakages. 
•  Do not upset the vessels while they are connected to the device when running, because the liquid comes into direct contact with the hy-

drophobic antibacteria filter and immediately blocks aspiration; if this should happen, empty the vessel and replace the antibacteria filter. 

•  In the event of aspiration without vessel and/or antibacterial filter, or if you suspect that solid or liquid substances have entered 

the aspiration circuit, send the device to customer service.

Содержание MAXIASPEED

Страница 1: ...1 MAXIASPEED ...

Страница 2: ...LOTTEUR CAGE SCHWIMMER KORB FLOTADOR JAULA 1 5 1 6 1 2 1 3 1 10 1 9 1 8 Fig Abb 1a Fig Abb 1b Fig Abb 3a Fig Abb 3b Fig Abb 5 OK KO GABBIA IN ASSE CAGE IN AXIS CAGE DANS L AXE FLUCHTENDER KORB JAULA EN EJE GABBIA FUORI ASSE CAGE OUT OF AXIS CAGE HORS AXE NICHT FLUCHTENDER KORB JAULA FUERA DE EJE 1 4 2 1 6 ...

Страница 3: ...e prise appropriée 1 7 Roulettes antistatiques avec frein 1 8 Prise d alimentation avec fusible pouvant être inspecté 1 9 Compartiment porte câble et interrupteur à pédale 1 10 Prise pour interrupteur à pédale quand il existe 2 Filtre antibactérien 3 Tuyau court Ø 8x14 mm 4 Récipients 2000cc or 4000cc 4 1 Prise VACUUM 4 2 Prise PATIENT 5 Tuyau long Ø 8x14 mm 6 Raccord d aspiration 7 Outil pour poi...

Страница 4: ...s for use The warnings and cautions described in the instruction manual must be observed IMPORTANT WARNINGS This is a medical device and it must be used by authorised personnel It must be operated as instructed in this use manual It is important for the operator to read and understand the use and maintenance information of the device Contact your retailer for any questions you may have MICROBIAL C...

Страница 5: ...on vessel is complete with overflow valve lid and vessel in a clear mate rial polycarbonate The antibacterial filter also protects the suction circuit from possible contaminants sucked in during use 1 3 Connect the long tube to the PATIENT connector 4 2 of the selected plug 2 Operation of the telescopic handle for space reasons the aspirator is supplied with the telescopic handle lowered to raise ...

Страница 6: ... cage is positioned correctly inside its housing It must be in axis with the plug aspiration hole so that the floating valve can block it in the circumstance in which the aspirated liquid exceeds the maximum volume the vessel can receive Do not connect the cage as shown in figure 3b it must be connected as shown in figure 3a Check that the float is fitted in the correct position and that it is fre...

Страница 7: ... conditions Class of risk according to the 93 42 EEC directive Storage conditions Operating storage air pressure min 690 hPa max 1060 hPa Double pump version High flow high vacuum device 230V 50 60Hz 330VA approx 90 l min approx 50 dBA 1 m T2 5A 250V IIa Temperature min 10 C max 40 C Humidity min 10 max 95 Temperature min 25 C max 70 C Humidity min 10 max 95 approx 12 kg approx 13 kg approx 16 kg ...

Страница 8: ...13 ...

Страница 9: ...29 ...

Страница 10: ...30 ...

Страница 11: ...31 ...

Страница 12: ... in your area or returning it to the distributor when buying a new device of the same type to be used with the same functions This procedure of separated collection of electric and electronic devices is carried out forecasting a European environmental policy aiming at safeguarding protecting and improving environment quality as well as avoiding potential effects on human health due to the presence...

Отзывы: