Quick Reference Cards   |   CG000543   |   Rev A

Visium CytAssist

for Instrument Operation

For use with:

Visium CytAssist & Accessory Kit, 12-Month Warranty, PN-1000441

 

Visium CytAssist Instrument Accessory Kit, PN-1000433

Visium CytAssist & Accessory Kit, 24-Month Warranty, PN-1000442

 

Visium CytAssist Instrument Accessory Kit, PN-1000433

Содержание Visium CytAssist

Страница 1: ...ist for Instrument Operation For use with Visium CytAssist Accessory Kit 12 Month Warranty PN 1000441 Visium CytAssist Instrument Accessory Kit PN 1000433 Visium CytAssist Accessory Kit 24 Month Warranty PN 1000442 Visium CytAssist Instrument Accessory Kit PN 1000433 ...

Страница 2: ... 10x Genomics products may be covered by one or more of the patents as indicated at www 10xgenomics com patents All products and services described herein are intended FOR RESEARCH USE ONLY and NOT FOR USE IN DIAGNOSTIC PROCEDURES The use of 10x Genomics products in practicing the methods set forth herein has not been validated by 10x Genomics and such non validated use is NOT COVERED BY 10X GENOM...

Страница 3: ...tain all packaging materials in the event that the instrument needs to be moved d Remove Shipping Bolt from the back of the instrument Retain for future use e Openthelid Carefullyremovetapefromthe fourTissueSlideClipsandtheVisiumSlideLock f Remove glass protective films from the alignment windows on both stages Close lid clicking sound indicates closure g Within Instrument Accessory Kit locate the...

Страница 4: ... a networked device login to your 10x Genomics account at register 10xgenomics com and register the instrument Enter the serial number instrument ID Read and accept the legal agreement d Update the firmware Internet connection After registration the instrument will automatically search for the latest firmware update Confirm data transmission settings for automatic upload download No Internet conne...

Страница 5: ...ue play button to start the test g A successful Readiness Test marks the end of setup Follow onscreen instruction in the event of a failed or incomplete test h Upon test completion follow onscreen instruction to unload slides i Store the Calibration Slide away from sun light and extreme temperature in the original packaging bag Retain for future use j Wipe down the Tissue Slide Stage and the Visiu...

Страница 6: ...Assist Slide serial number Custom run name temperature and time 37 C for 30 min is recommended for most applications Sample names the 6 5 mm capture areas are designated A1 and D1 The 11 mm capture areas are designated A and B c Open the lid d Follow onscreen instruction and load the blank slides and Calibration Slide onto the instrument See Load Slides section for details e Load reagents to each ...

Страница 7: ...m Visium CytAssist Training Slide Visium CytAssist Calibration Slide Load Tissue Slides onto Visium Slide Stage a Inspect the Visium Slide Stage for dust debris b Open the Visium Slide Lock by using one finger to pivot c Line up the slide with the label face up and oriented to the right d Fit the slide within the raised grooves on the left top and bottom e Secure the slide by pivoting the Visium S...

Страница 8: ...s the blue Run Start Button on the touchscreen Enter the serial number located on the Visium CytAssist Calibration Slide label b Locate two blank slides and the Visium CytAssist Calibration Slide in the accessory kit Clean with compressed air c Follow onscreen instruction to load the blank slides and calibration slide onto the instrument d Press the blue play button to start the test e Upon test c...