Check button (9) on the top of the device if it is in secured position (10b). Afterwards connect the 
device by fixing of inlet hose (3) to the sampling point. The inlet hose (3) must be secured not to 
fall out with the pressure. 

If  there  is  a  big  difference  in  diameters  of  the  hose  and  sampling  valve,  use  a  reduction  in  the 
supplied hose and the other one in another hose satisfactory for safe sampling. We recommend to 
fix the hoses on the reduction with clamps. 

Place  the  outlet  hose  (5)  to  the  drain,  the  inlet  valve  (1)  is  shut  during  this  handling.  Afterwards 
open fully the valve of the sampling point  and let beer or saturated beverage go into the inlet hose 
(3) to inlet valve (1). Open fully the inlet valve (1) on the device. 

5.1.2

 

CO

2

 

C

ONTENT MEASUREMENT

 

Open  both valves (1), (2)  to let the sample fill in the transparent chamber 

(6) 

SLOWLY, to avoid 

foam creation. 

Let the beer flow through the device for at least 20s, 30s or more if you are rinsing the sample from 
a previous measurement or if it’s the first measurement and the GMD was stored in a warmer place 
before.  

Fill the transparent chamber (6) completely with the sample. Check visually if there is no foam or 
bubbles inside the chamber (6)!  

The device is on the ground during taking of sample. 

Now first shut the outlet (2) valve, then the inlet (1) valve. Press the power button (12) to turn on. 
Switch on the microcontroller by pressing the On key and when the device is ready, press the Enter 
key. The GMD starts measuring continuously the pressure and waits for depressurization. You have 
to open the outlet valve (2) to remove the overpressure from the tank. Open it slowly and close it 
immediately after hearing the sound emitted by the buzzer and the display shows "close v.", which 
indicates  that  the  pressure  is  under  threshold  level.  Then  push  out  the  button  (9)  of  the  secured 
position (10b) and pull  it out into the upper position (11) (look pic.2). Hold it for a while in this 
position  to  equalize  the  pressure.    Then  press  down  the  button  (9)  to  the  very  down    unsecured  
position  (10  a).  After  this  handling  dissolved  CO

2

  is  fluttered  and  the  pressure  increases  in  the 

chamber (6). Repeat this action one more time.  

You should not pump more than three times to get good results and a good repeatability. 

If the device was stored in a warm place before using it, the first measurement won’t be relevant 
because of big temperature variation.  

Then lock the button (9) in the down secured position (10 b).       

Then follow the procedure described  in 

5.2.3

 

CO2

 

C

ONTENT MEASUREMENT

 

 

 

Содержание GMD

Страница 1: ...67 580 01 Havlíčkův Brod Czech Republic tel 420 569 433 620 fax 420 569 422 144 1 cube 1 cube com http www 1 cube com GMD and GMDK MEASURING DEVICE OF CO2 CONTENT IN FERMENTING TANKS AND KEG CASKS USER S GUIDE ...

Страница 2: ...2 ...

Страница 3: ...s 8 5 1 Measurement procedure 8 5 1 1 Taking of samples 8 5 1 2 CO2 Content measurement 9 5 1 3 Discharging of the sample 10 5 1 4 Cleaning of the device 10 5 2 Control of the electronics 11 5 2 1 Switching ON and OFF the device 11 5 2 2 Charging of the device 11 5 2 3 CO2 Content measurement 12 5 2 4 Reading recorded values of CO2 from memory 13 5 2 5 Deleting the content of the memory 13 6 0 CO2...

Страница 4: ...l accessories withdrawal head for kegs 1unit Note accessories is not part of the supply only on customer s demand GMD is a measuring device for measurement of CO2 content in beer soft drinks and mineral water in tanks and KEG casks Note Adapted only for industrial application Whenever in the future GMD unit can be upgraded to GMDK unit if required With the comfortable version GMDK the CO2 measurem...

Страница 5: ...5 2 0 INSTALLATION PUTTING THE DEVICE INTO OPERATION AND ITS MAINTENANCE Pic 1 GMD set front view rear view 2 8 9 4 1 3 3 1 4 12 13 6 5 2 Pic 2 Button position 10 11 ...

Страница 6: ...2 Power button 13 Connector cover 2 1 TAKING OF SAMPLES The device is ready for measurement at the moment of delivery Preceding the own measurement connect the device by fixing of inlet hose 3 to the sampling point The measured sample must flow through stainless tube 4 to the bottom of the vessel The inlet hose 3 must be secured at the sampling point not to fall out during taking of sample under p...

Страница 7: ...be filled with water during cleaning During rinsing pull out and consequently press down several times the button 9 to rinse both pump and capillary It is important to remove any traces of beer to prevent the piston getting stuck After rinsing Once the cleaning is over switch off the water cock Disconnect the inlet hose 3 from the water cock If you plan to use the device next day do not discharge ...

Страница 8: ...ge of temperature measurement 0 30 C range of pressure measurement 0 600 kPa accuracy of CO2 content measurement 0 1 g l or in mass percentage 0 01 m accuracy of temperature measurement 0 1 C accuracy of pressure measurement 0 5 dimensions 280x180x300 mm weight of empty device about 1 5kg protection IP 54 with threaded cover on connector 5 0 OPERATING INSTRUCTIONS 5 1 MEASUREMENT PROCEDURE You can...

Страница 9: ...s no foam or bubbles inside the chamber 6 The device is on the ground during taking of sample Now first shut the outlet 2 valve then the inlet 1 valve Press the power button 12 to turn on Switch on the microcontroller by pressing the On key and when the device is ready press the Enter key The GMD starts measuring continuously the pressure and waits for depressurization You have to open the outlet ...

Страница 10: ...ter Both ball valves 1 2 have to be open during rinsing Warning Pressure of water must not exceed value 250kPa Whole internal space of the vessel 6 must be filled with water during cleaning During rinsing pull out and consequently press down several times the button 9 to rinse both pump and capillary It is important to remove any traces of beer to prevent the piston of the pump getting stuck After...

Страница 11: ...y for a second It is possible to work with the device for another 30 minutes If the device is not recharged during this time then the text Low voltage appears on the display and the device is shut down If you want to monitor how the battery is charged while working with the device press the rotation key the text battery appears on the first line of display and the voltage of the battery on the sec...

Страница 12: ...ing briefly the outlet valve 2 When the pressure is low enough you will hear the buzzer of the GMD and the display shows close v which indicates that you have to close immediately the valve 2 Now you can start releasing the CO2 from the beer with the pump on the top of the GMD look pic 2 Push out the button 9 of the secured position 10b and pull it out into the upper position 11 Hold it for a whil...

Страница 13: ... 5 2 5 DELETING THE CONTENT OF THE MEMORY The measured values can be deleted from memory at any time The memory capacity will be completely recovered 450 values and newly measured values will be saved in sequence 1 Press the key ON to switch on the device The text 1 CUBE with the program version appears on the display for a moment Then the following text appears ENT meas view 2 Press the key minus...

Страница 14: ...n the second line At the same time these values are displayed in DataViewer program window in column in turn from the oldest value to the newest ones and also plotted on the graph These values can be saved into file with the Save file button Note the measured values are displayed in g l in DataViewer program 7 0 SERVICE For service contact the manufacturer 1 CUBE Hamry 3567 580 01 Havlíčkův Brod C...

Отзывы: