Measuring device GMD (GMDK) can be used only for determination of CO
2
content in the range of
measured values determined by technical conditions. Never connect the measuring device to the
withdrawal spots where measured parameters are over measuring capacity of the device.
Check device before each measurement. Do not use visibly damaged device and contact the
qualified service personnel who provides service for delivered device.
Warning!
It is forbidden to use the device for pressure higher than 600 kPa and for temperature
higher than +30 degrees Celsius. It could cause device destruction and staff injury.
4.0
T
ECHNICAL DATA
range of CO
2
measurement ....................... 2.0 – 9.99 g/l or in mass percentage 0.2 – 0.99 m%
range of temperature
measurement ..... 0 – +30 °C
range of pressure
measurement ............. 0 – 600 kPa
accuracy of CO
2
content measurement. ± 0.1 g/l or in mass percentage ± 0.01 m%
accuracy of temperature measurement ± 0.1 °C
accuracy of pressure measurement ....... ± 0.5 %
dimensions ....................................................... 280x180x300 mm
weight (of empty device) ............................ about 1.5kg
protection .......................................................... IP 54 (with threaded cover on connector)
5.0
O
PERATING INSTRUCTIONS
5.1
M
EASUREMENT PROCEDURE
You can see the instruction video on:
https://youtu.be/XRgAYRne3HM
5.1.1
T
AKING OF SAMPLES
Measuring device for measurement of CO2 content in beer, soft drinks and mineral water in
cylindro-conical fermenters, tanks and KEG casks completed with withdrawal head for kegs
(Optional accessories).
Check device before each measurement. Do not use visibly damaged device
.
Notice!
Sludge the sampling point before connecting the device. Sludge the sedimented yeast to
fill inside the chamber only beer without sedimented yeast!