Digital Weighing Indicator 

 

  User Manual 

  V 1.10 

 

 

 
 

 

 
 
 

 

 

Shanghai Yaohua Weighing System Co.,

 Ltd

Summary of Contents for XK3190-A12+SS

Page 1: ...Digital Weighing Indicator User Manual V 1 10 Shanghai Yaohua Weighing System Co Ltd ...

Page 2: ...ATOR CONNECTION 4 CHAPTER 3 CALIBRATION 5 3 1 GENERAL CALIBRATION 5 3 2 FAST CALIBRATION 6 CHAPTER 4 OPERATION 7 4 1 POWER ON AND AUTO ZERO SETTING 7 4 2 MANUAL ZERO SETTING 7 4 3 TARE FUNCTION 7 4 4 ACCUMULATING FUNCTION 7 4 5 COUNTING FUNCTION 7 4 6 USER S FUNCTION SETTING 7 CHAPTER 5 MAINTENANCE 9 CHAPTER 6 ERROR INDICATION 10 DEAR CUSTOMERS PLEASE READ THE USER GUIDE CAREFULLY BEFORE USING THE...

Page 3: ...ng temperature 25 55 12 relative humidity 85 RH Features 1 High precision A D conversion readability 1 30000 2 call and display inner code to replace weight observing and analysis tolerance 3 special digital filtering technology to strengthen anti vibration ability 4 able to set up digital filter intensity range and stable time 5 with counting function unit weight data save protection in case of p...

Page 4: ...efinition of each feet refers to picture 2 3 2 Indicator adopts 4 wire connection as default Indicator must be reliably connected to load cell and the shielded cable of load cell must be reliably connected to underground To protect the indicator and load cell we cannot plug or withdraw the connector when the indicator is power on Both the load cell and indicator are static sensitive devices You mu...

Page 5: ...ht net weight or tare The format of G W ww000 000kg or ww000 000lb Make sure that scoreboard and output lead wires are connected correctly If there is something wrong with connection damage will happen to output port of instrument and input port of scoreboard sometimes the damage is so big to influence the instrument and scoreboard Only specially provided connecting cable is allowed to be used Mak...

Page 6: ...3 Calibration 3 1 GENERAL CALIBRATION Connect load cell properly then turn on the indicator keep pressing key while it s initializing it will enter the calibration mode and display d X STEP OPERATION DISPLAY NOTES 1 Press TARE to change the division and press to confirm d X Division setting For example Press when displayed d 5 then the division is set to be 5 and the indicator enters decimal point...

Page 7: ...o the weighing mode Note if no pressing the calibration switch at back of indicator all the parameters won t be saved 3 2 FAST CALIBRATION Connect load cell properly then turn on the indicator keep pressing key while its initialization it will enter into the calibration mode and display d X 3 2 1 Fast calibration of zero point At any time before it showing nOLOAD press FUNC Indicator will keep the...

Page 8: ...le press Tare key indicator will deduct the displayed weight value as tare weight Then indicator displays net weight as 0 and Tare sign annunciator is on 4 4 COUNTING FUNCTION In weighing mode press Func to enter the counting state it will display count and press it will display C00000 then press Tare to move the digit corresponding with the small triangle press ZERO key the digit where the triang...

Page 9: ...ate setting x 1 9600 x 2 4800 x 3 2400 x 4 1200 4 P4 x RS232 Net Gross weight output option x 1 Net weight output x 2 Gross weight output x 3 Tare weight output 5 P5 x RS232 output mode option x 1 No transmission RS232 stop x 2 Continuous transmission x 3 Continuous transmission when stable x 4 Command mode Z zero T tare R transmit weight data once time x 5 scoreboard display Current loop output x...

Page 10: ...x 1 low x 2 middle x 3 high Chapter 5 Maintenance 5 1 To ensure the clarity and service life of the indicator it must be kept away from direct sunlight during using and the ground where the indicator stands must be smooth 5 2 It is improper to use this indicator in a dustful or vibrant or damp environment 5 3 The load cell and indicator need good connection System must have a good ground connectio...

Page 11: ...indicator and external equipment the power of both indicator and equipment should be cut off 5 7 Advice of the company our company is responsible for the indicator quality but not responsible for the problems of the system where the indicator locates Your attention is required when making purchase 5 8 Please use the indicator outward interfaces strictly as per the operating instruction manual Do n...

Page 12: ...7282812 Fax 021 67282810 Email sales yaohua com cn louise yaohua com cn Please make sure no load is on the scale when turning on ERR 4 The input sample quantity is zero when sampling in counting mode Please input the right sample quantity ERR 5 When full capacity calibrating the weight input is zero Please input the weight in accordance with the load on scale ERR 6 The unit weight is less than 0 2...