SPECIFICATIONS

Range:   –58 to 2372 °F

 

–50 to 1300 °C

Resolution:  0.1° or 1° (selectable)

THERMOCOUPLE

The sensor supplied is a stainless-steel Type-K 

probe.  Maximum temperature for this sensor 

is 446 °F (230°C).  Accessory thermocouples 

are available for measuring temperatures 

throughout the entire range of the unit (See 

“Accessories”).  

Only Type-K thermocouples 

may be used with this unit.

Longer leads and extensions may be used.  

The lead effect for 164 feet (50 meters) is 

generally less than 0.2 °C with a limit of error 

of 2.2 °C.  Type-K wire and connectors must 

be used.

OPERATION

Insert the thermocouple plug into the socket 

located on the front of the instrument.  The 

plug and socket are keyed with a small and 

large pin.  Make certain the plug is inserted 

correctly, do not insert the plug incorrectly and 

force it.
Press the POWER button to turn the unit on.  
Pressing the °F button will display the tem-

perature in Fahrenheit (°F will appear on the 

display).   Pressing the °C button will display 

the temperature in Celsius (°C will appear on 

the display).
Press the 0.1° or 1.0° button to select the 

desired temperature resolution.  (Note: 

0.1° resolution can only be used when the 

temperature being measured is between -50° 

and 199.9°)
Place the thermocouple in contact with 

the material to be measured and read the 

temperature on the display.

Press the POWER button to turn the unit off.  

To conserve battery life, always turn the unit 

off when not in use.

HOLD

Press the HOLD button to “freeze” the display 

at the current reading (“H” will appear at the 

top left corner of the display).  Press the 

HOLD button again to return to the current 

reading.

OFFSET ADJUSTMENT (Fine tuning)

By adjusting the offset control, found on the 

front of the unit, the measurement accuracy 

may be optimized for a particular thermocou-

ple at a particular temperature.

While offset adjustments are possible 

they will VOID with the factory calibration

This unit has been calibrated Traceable

®

 to 

standards provided by the National Institute

of Standards and Technology (NIST).  

Traceable

®

 recalibration must be made by 

the factory (See the “Warranty, Service or 

Recalibration” section).

BENCH STAND/WALL MOUNT

The unit is supplied with a bench stand that 

is on the back of the case.  Locate the stand 

on the back of the unit.  With your thumb 

and forefinger, lift the stand out.  To close the 

stand, simply snap it shut.
The unit may be wall mounted by using a 

screw.  Set the screw into the wall at the 

desired location.  Do not set the screw flush 

to the wall, the head of the screw will need 

to slip into the receptacle on the back of the 

unit.  Hang the unit by sliding the receptacle 

located on the back of the unit over the head 

of the screw.