manualshive.com logo in svg
background image

MODEL

6535: 

TraceableGO™

 

Datalogging Thermometer

6537:

 TraceableGO™ Datalogging Hygrometer

Figure 1:

 6535 (left), 6537 (right)

Dimension

 L x H x D: 3.5 x 2 x 0.79” (89x51x20mm)

TEMPERATURE

Operating Range:

 –20 to 70°C

Resolution: 

0.1°C

Accuracy:

6535: ±0.4°C

6537: ±0.4°C between –10 and 70°C, otherwise 

±0.5°C

RELATIVE HUMIDITY (6537 ONLY):

Range:

 0% to 95%RH, non-condensing

Resolution:

 0.1% RH

Accuracy:

 ±3% RH between 5 to 75%, otherwise 

±5% RH

BATTERY

• 2 AAA Alkaline batteries (3.0V)

Battery Level Indication:

Battery level

LCD symbol

≥ 80% (2.78V)

≥ 60% (2.56V), < 80%

≥ 40% (2.34V), < 60%

≥ 20% (2.12V), < 40%

≥ 10% (2.01V), < 20%

< 10% Flashing

Note: 

Battery level is updated every 5 minutes.

Note: 

Once battery level drops to below 10%, device 

may not work properly. Replace batteries immediately.

Note:

 When replacing batteries, after removing the 

old batteries, wait 10 seconds before inserting new 

batteries. Otherwise, Bluetooth may not work properly.

MEASUREMENT READING UPDATE FREQUENCY

Temperature and humidity: 5 seconds;

Note:

 If a reading is out of operating range, the 

corresponding location on LCD will display ‘--.-‘, and 

such out of range reading will not trigger alarm.

DATA LOGGING FREQUENCY:

5 min by default, user-adjustable between 1 minute 

and 12 hours with 1 minute step.

DATA STORAGE CAPACITY

Alarm:

 most recent 90 Alarm events

Data:

 64K (65536) data points, 7.5 months on 

5-minute logging interval

Note:

 If maximum number of alarm events is reached, 

device will overwrite the oldest alarm events.

_____________________________________________________________________________________ 

Page 

3

 of 

10

 

Figure 2 

Device not initialized

Figure 3 

Device’s S/N not programmed

10.Device Operating Modes:

IDLE Mode

: First time battery inserted, and device has not been configured;

STANDBY Mode

: Device has been configured, but not started;

RUN Mode

: Device starts to measure and log data.

STOP Mode

: Device stops from RUN Mode. In STOP Mode, device does not update

measurements or log data, and the last measurements are displayed.

TEST Mode

: For test and calibration in production.

DTM Mode

: For Bluetooth radio test.

11.View Current Reading

1) 6535: Current reading, minimum/maximum reading since last clearing, run

time/alarm time toggling, Memory mode are shown on LCD.

2) 6537: Current temperature/humidity  reading  toggling  every 5 seconds,

minimum/maximum reading since last clearing, run time/alarm time, Memory mode

are shown on LCD.

12.View Current MIN/MAX

6535: Current temperature Min/Max values are displayed on LCD 

6537: Current temperature/humidity Min/Max values are toggled display. 

Note: Every time device is configured, or resumes to run from STOP Mode, Min/Max 

values are reset. 

13.Viewing RUN TIME/ALARM TIME

Run Time/Alarm Time is toggled to display on LCD. If Run Time is displayed, LCD symbol 

 appears; if Alarm TIME is displayed, LCD symbol 

 appears. 

Note: Alarm Time is accumulated for both Low Alarm and Hi Alarm for each channel 

(temperature, humidity). 

14.View Memory Usage, Current Date/Time

, Device’s Serial Number

1) Short press and release

DEVICE OPERATING MODES

• 

IDLE Mode: First time battery inserted, and device 

has not been confi gured;

• 

STANDBY Mode: Device has been confi gured, but 

not started;

•  RUN Mode: Device starts to measure and log data.

•  STOP Mode: Device stops from RUN Mode. 

In STOP Mode, device does not update 

measurements or log data, and the last 

measurements are displayed.

VIEW CURRENT READING

1.  Temperature ONLY unit: Current reading, 

minimum/ maximum reading, run time/alarm time 

toggling, Memory mode are shown on LCD.

2.  Humidity and Temperature unit: Current 

temperature/humidity reading toggling every 5 

seconds, minimum/maximum reading since last 

clearing, run time/alarm time, Memory mode are 

shown on LCD.

VIEW CURRENT MIN/MAX

Temperature ONLY unit: Current temperature Min/Max 

values are displayed on LCD.

Humidity and Temperature unit: Current temperature/

humidity Min/Max values are toggled display.

Note:

 Every time device is confi gured, or resumes to 

run from STOP Mode, Min/Max values are reset.

VIEWING RUN TIME/ALARM TIME

Run Time/Alarm Time is toggled to display on LCD. 

If Run Time is displayed, LCD symbol 

RUN TIME

 

appears; if Alarm Time is displayed, LCD symbol 

ALARM TIME 

appears.

Note:

 Alarm Time is accumulated for both Low 

Alarm and Hi Alarm for each channel (temperature, 

humidity).

VIEW MEMORY USAGE, CURRENT DATE/TIME, 

DEVICE’S SERIAL NUMBER

1.  Press and release 

START/STOP

 button

2.  Memory usage in percentage is shown on LCD. 

The percentage indicates how much internal data 

memory storage has been used;

3.  Number of days, hours, minutes left till memory is 

full are also displayed on the second line;

_____________________________________________________________________________________ 

Page 

4

 of 

10

 

2) Memory usage in percentage is shown on LCD. The percentage indicates how much

internal data memory storage has been used;

3) Number of days, hours, minutes left till memory is full are also displayed on the

second line;

Figure 

3

4) Within 10 seconds, short press

 again, current date/time is displayed on device 

LCD. The following figure shows 9/14/2017, 17:30 

Figure 

4

5) Within 10 seconds, short press

 again, device S/N will be displayed on LCD. 

Figure 

5

6) To go back to normal working status, press

 again, or wait for 10 seconds and 

device will automatically go back to normal working status. 

Note: 

If Memory Mode is set to WRAP WHEN FULL: LCD symbol 

starts to 

flash on the display when the memory is full. Once memory is full, the oldest data 

points will be overwritten with new data points.

4.  Within 10 seconds, press 

START/STOP

 again, 

current date/time is displayed on device LCD. The 

following fi gure shows 9/14/2017, 17:30

_____________________________________________________________________________________ 

Page 

4

 of 

10

 

2) Memory usage in percentage is shown on LCD. The percentage indicates how much

internal data memory storage has been used;

3) Number of days, hours, minutes left till memory is full are also displayed on the

second line;

Figure 

3

4) Within 10 seconds, short press

 again, current date/time is displayed on device 

LCD. The following figure shows 9/14/2017, 17:30 

Figure 

4

5) Within 10 seconds, short press

 again, device S/N will be displayed on LCD. 

Figure 

5

6) To go back to normal working status, press

 again, or wait for 10 seconds and 

device will automatically go back to normal working status. 

Note: 

If Memory Mode is set to WRAP WHEN FULL: LCD symbol 

starts to 

flash on the display when the memory is full. Once memory is full, the oldest data 

points will be overwritten with new data points.

5.  Within 10 seconds, press 

START/STOP 

again, 

device S/N will be displayed on LCD.

_____________________________________________________________________________________ 

Page 

4

 of 

10

 

2) Memory usage in percentage is shown on LCD. The percentage indicates how much

internal data memory storage has been used;

3) Number of days, hours, minutes left till memory is full are also displayed on the

second line;

Figure 

3

4) Within 10 seconds, short press

 again, current date/time is displayed on device 

LCD. The following figure shows 9/14/2017, 17:30 

Figure 

4

5) Within 10 seconds, short press

 again, device S/N will be displayed on LCD. 

Figure 

5

6) To go back to normal working status, press

 again, or wait for 10 seconds and 

device will automatically go back to normal working status. 

Note: 

If Memory Mode is set to WRAP WHEN FULL: LCD symbol 

starts to 

flash on the display when the memory is full. Once memory is full, the oldest data 

points will be overwritten with new data points.

6.  To return to normal working status, press 

START/

STOP 

again, or wait for 10 seconds and device will 

automatically go back to normal working status.

Note:

 If Memory Mode is set to WRAP WHEN FULL: 

LCD symbol 

_____________________________________________________________________________________ 

Page 

4

 of 

10

 

2) Memory usage in percentage is shown on LCD. The percentage indicates how much

internal data memory storage has been used;

3) Number of days, hours, minutes left till memory is full are also displayed on the

second line;

Figure 

3

4) Within 10 seconds, short press

 again, current date/time is displayed on device 

LCD. The following figure shows 9/14/2017, 17:30 

Figure 

4

5) Within 10 seconds, short press

 again, device S/N will be displayed on LCD. 

Figure 

5

6) To go back to normal working status, press

 again, or wait for 10 seconds and 

device will automatically go back to normal working status. 

Note: 

If Memory Mode is set to WRAP WHEN FULL: LCD symbol 

starts to 

flash on the display when the memory is full. Once memory is full, the oldest data 

points will be overwritten with new data points.

 starts to fl ash on the display when 

the memory is full. Once memory is full, the oldest 

data points will be overwritten with new data points.

If Memory Mode is set to STOP WHEN FULL: LCD 

symbol 

MEM

 starts to fl ash on the display when the 

memory is 95% full. Once memory is full, the device 

will stop logging new data points.

ENABLE/DISABLE BLUETOOTH

1.  Quick press twice 

START/STOP 

to enable 

Bluetooth, and Bluetooth LCD symbol   appears;

2.  Device starts to advertise to be found, and 

to be connected. Device Name shown on 

TraceableGO™ App discovered list looks like 

CC6535-xxxx, or CC6537-xxxx, where xxxx is the 

last 4 digits of device serial number.

3.  If no connection has been made for 

ONE MINUTE

Bluetooth will be disabled to save battery life, and 

Bluetooth LCD symbol   disappears, or quick 

press twice again to disable Bluetooth. To enable 

again, go to Step (1).

HOW TO SETUP DEVICE 

Confi gure device before use

1.  Enable Bluetooth on device.

2.  Use TraceableGO™ App to connect to device.

3. 

Follow TraceableGO™ App instruction to confi gure 

device.

4. 

The following parameters are confi gured through 

TraceableGO™ App: START Mode, STOP Mode, 

ALARM Enable/Disable, Celsius/Fahrenheit, 

Memory Mode, Data Logging Interval, Alarm 

Setting.

5. 

Once device has been confi gured, device enters 

STANDBY Mode.

Note:

 While device is connected to our Traceable-

GO™ App, if no action has been taken for 5 minutes, 

device will terminate the connection from our 

TraceableGO™ App.

Note:

 If Memory Mode is set to WRAP WHEN FULL, 

_____________________________________________________________________________________ 

Page 

4

 of 

10

 

2) Memory usage in percentage is shown on LCD. The percentage indicates how much

internal data memory storage has been used;

3) Number of days, hours, minutes left till memory is full are also displayed on the

second line;

Figure 

3

4) Within 10 seconds, short press

 again, current date/time is displayed on device 

LCD. The following figure shows 9/14/2017, 17:30 

Figure 

4

5) Within 10 seconds, short press

 again, device S/N will be displayed on LCD. 

Figure 

5

6) To go back to normal working status, press

 again, or wait for 10 seconds and 

device will automatically go back to normal working status. 

Note: 

If Memory Mode is set to WRAP WHEN FULL: LCD symbol 

starts to 

flash on the display when the memory is full. Once memory is full, the oldest data 

points will be overwritten with new data points.

 appears on LCD; If Memory Mode is set to 

STOP WHEN FULL, 

MEM

 appears on LCD.

ALARM

1.  Once an alarm is triggered by any of temperature, 

humidity out of set alarm range, LCD symbol 

LOW 

ALM

 and/or 

HI ALM

 starts to fl ash. An Alarm Event 

will be logged.

2.  Pressing 

START/STOP 

once will clear the alarm, 

LCD symbol stops fl ashing. An Alarm Acknowledge 

Event will be logged.

3.  If temperature, or humidity falls back to normal 

range, an Alarm Event will be logged. If any of 

temperature, humidity measurement goes back to 

out of alarm range, alarm will be triggered again.

HOW TO START, STOP, AND RESUME DATA 

LOGGING

1.  Devices are factory initialized with default setting:

• START Mode: Push Start

• STOP Mode: Push Stop

• 

Memory Mode: Wrap when Memory is Full

• Temperature Unit: Celsius

• Data Logging Interval: 5 minutes

• Alarm Enable/Disable: Enabled

• Alarm Setting High Alarm:

▪ Temperature: 8°C

▪ Humidity: 75% RH (6537 only)

• Alarm Setting Alarm Low:

▪ Temperature: 2°C

▪ Humidity: 25% RH (6537 only)

• Date/Time is also set to current to central time.

Note: 

Devices are shipped with battery, powered 

on, and ready to use if using default setting. Without 

powered on, device will lose track of date/time.

Note:

 If a user intends to use device for different 

setting other than default, device must be re-

initialized.

_____________________________________________________________________________________ 

Page 

1

 of 

10

 

6

535

/

6

537

 

B

LUETOOTH

 

D

ATA

 

L

OGGER

 

I

NSTRUCTION

 

11/01/2017

1.

Mode Number:

6535: Bluetooth Data Logging Thermometer

6537: Bluetooth Data Logging Hygrometer

2.

Device Display:

LCD View Angle: 6 

O’clock

Figure 1: 

6535 (left), 6537 (right)

 

Dimension L x H x D: 

3.5” x 2” x 0.

79

” (

89mm x 51mm x 20mm) 

3. Temperature:

Operating Range: -20°C to 70°C

Resolution: 0.1°C

Accuracy:

6535: ±0.4°C 

6537: ±0.4°C between -10°C and 70°C, otherwise ±0.5°C 

4. Relative humidity (6537 Only):

Range: 0% to 95%RH, non-condensing

Resolution: 0.1% RH

Accuracy: ±3% RH between 5 to 75%, otherwise ±5% RH

Recommended Humidity Operating Range: 20% to 80% (non-condensing) over -10°C to 60°C.

5. Battery

2 AAA Alkaline batteries (3.0V)

Battery Level Indication:

Battery Level 

LCD Symbol 

 80% (2.78V) 

Отзывы: