background image

T3PS40381P, T3PS60251P,

T3PS062001P

Programmable DC Power Supplies

Programming Manual

Summary of Contents for T3PS062001P

Page 1: ...T3PS40381P T3PS60251P T3PS062001P Programmable DC Power Supplies Programming Manual ...

Page 2: ...roy company The information in this manual was correct at the time of printing However Teledyne LeCroy continues to improve products and reserves the rights to change specification equipment and maintenance procedures at any time without notice Unless otherwise specified all the materials and processes are compliant with RoHS Directive 2011 65 EU in its entirety inclusive of any further amendments...

Page 3: ...verview 10 Appearance 13 Configuration Settings 20 REMOTE CONTROL 28 Interface Configuration 29 Command Syntax 48 Command List 51 Status Register Overview 118 Error List 131 APPENDIX 140 T3PS Factory Default Settings 140 Error Messages Messages 143 LED ASCII Table Character Set 144 INDEX 145 ...

Page 4: ...instrument in the best possible condition Safety Symbols These safety symbols may appear in this manual or on the instrument WARNING Warning Identifies conditions or practices that could result in injury or loss of life CAUTION Caution Identifies conditions or practices that could result in damage to the T3PS or to other properties DANGER High Voltage Attention Refer to the Manual Protective Condu...

Page 5: ... Measurement categories EN61010 1 2010 and EN61010 2 030 specifies the measurement categories and their requirements as follows The T3PS falls under category II Measurement category IV is for measurement performed at the source of low voltage installation Measurement category III is for measurement performed in the building installation Measurement category II is for measurement performed on the c...

Page 6: ...lative Humidity 20 85 no condensation Altitude 2000m Temperature 0 C to 50 C Pollution Degree EN61010 1 2010 and EN61010 2 030 specifies the pollution degrees and their requirements as follows The T3PS falls under degree 2 Pollution refers to addition of foreign matter solid liquid or gaseous ionized gases that may produce a reduction of dielectric strength or surface resistivity Pollution degree ...

Page 7: ...midity 90 no condensation Disposal Do not dispose this instrument as unsorted municipal waste Please use a separate collection facility or contact the supplier from which this instrument was purchased Please make sure discarded electrical waste is properly recycled to reduce environmental impact ...

Page 8: ... be connected to the terminal which is marked with the letter N or coloured Blue or Black The wire which is coloured Brown must be connected to the terminal marked with the letter L or P or coloured Brown or Red If in doubt consult the instructions provided with the equipment or contact the supplier This cable appliance should be protected by a suitably rated and approved HBC mains fuse refer to t...

Page 9: ...n After going through the overview please read the theory of operation to become familiar with the operating modes protection modes and other safety considerations T3PS Series Overview 10 Series lineup 10 Main Features 11 Accessories 12 Appearance 13 Front Panel 13 Display and Operation Panel 16 Rear Panel 18 Configuration Settings 20 ...

Page 10: ...els covering a number of different current voltage and power capacities Model name Voltage Rating1 Current Rating2 Power T3PS062001P 6V 200A 1200W T3PS40381P 40V 38A 1520W T3PS60251P 60V 25A 1500W 1 Minimum voltage guaranteed to 0 2 of rating voltage 2Minimum current guaranteed to 0 4 of rating current ...

Page 11: ... master slave operation with active current sharing Remote sensing to compensate for voltage drop in load leads 19 rack mounted ATE applications OVP OCP and OHP protection Preset memory function Adjustable voltage and current slew rates Bleeder circuit ON OFF setting CV CC priority start function Prevents overshoot with output ON Interface Built in RS 232 485 LAN and USB interface Analog output pr...

Page 12: ...re included Standard Accessories Part number Description Qty Output terminal cover 1 Analog connector plug kit 1 Output terminal M8 bolt set 6V 60V model 1 Input terminal cover 1 1U Handle ROHS 2 1U BRACKET LEFT RoHS 1 1U BRACKET RIGHT RoHS 1 Download Name Description T3PS_cdc inf T3PS USB driver Other Name Description Certificate of traceable calibration ...

Page 13: ...t USB A port for data transfer loading test scripts etc 3 Air Inlet Air inlet for cooling the inside of the T3PS series 4 Voltage Knob Voltage Used to set the voltage value or select a parameter number in the Function settings Display Area The display area shows setting values output values and parameter settings The function LEDs below show the current status and mode of the power supply See page...

Page 14: ...tion functions that have been activated 8 Function Button Function M1 Used to configure the various functions M1 Button Shift Used to recall the M1 setup Shift and hold Used to save the current setup to M1 9 Test Button TEST M2 Used to run customized scripts for testing M2 Button Shift Used to recall the M2 setup Shift and hold Used to save the current setup to M2 10 Set Button SET M3 Used to set ...

Page 15: ...11 Shift Button Shift Used to enable the functions that are written in blue characters below certain buttons 12 Output Button Output Used to turn the output on or off 13 Output ON LED Lights in green when the output is on ...

Page 16: ...ays the current or the value of a Function parameter 16 CV LED Lights in green during constant voltage mode 17 CC LED Lights in green during constant current mode 18 VSR LED Lights up when CV Slew Rate Priority is enabled 19 LAN LED Lights up when the LAN interface is connected 20 RMT LED Lights in green during remote control 21 ERR LED Lights in red when an error has occurred 22 DLY LED The Outpu...

Page 17: ...LED Lights in green when the memory value are being recalled or saved 26 M2 LED Lights in green when the memory value are being recalled or saved 27 M3 LED Lights in green when the memory value are being recalled or saved Note Only the ERR and ALM LED s are red All the others are green ...

Page 18: ...mp connector 2 DC Output Output terminals for 6V to 60V models 3 USB USB port for controlling the T3PS remotely 4 LAN LAN RS 485 232 Ethernet port for controlling the T3PS remotely 5 Remote IN LAN RS 485 232 Two different types of cables can be used for RS232 or RS485 based remote control RS232 support via user supplied cable with DB9 RJ45 connectors wired according to the table on page 124 of the...

Page 19: ... supplies with the Remote IN port to form a communication bus Serial link cable with RJ45 shielded connectors wired according to the table on page 129 of the T3PS user manual 7 Analog Control ANALOG PROGRAMMING External analog control connector 8 Remote Sense S LS S LS NC Compensation of load wire drop 9 Ground Screw Connectors for grounding the output two positions shown in red L N AC INPUT 100 2...

Page 20: ...g F 89 Show Version can only be viewed not edited Configuration settings F 90 F 98 cannot be edited in the Normal Function settings Use the Power On Configuration settings See page 21 for details Steps 1 Press the Function key The function key will light up Function 2 The display will show F 01 on the left and the configuration setting for F 01 on the right VOLTAGE CURRENT VSR LAN RMT ERR DLY ALM ...

Page 21: ... Function Setting Power On Configuration Settings Background The Power On Configuration settings can only be changed during power up to prevent the configuration settings being inadvertently changed Ensure the load is not connected Ensure the power supply is off Steps 1 Hold the Function key whilst turning the power on Lock Local PROT Function Test Set Output Unlock ALM_CLR M 1 M 2 M 3 Shift Long ...

Page 22: ...ting Voltage Range F 90 F 98 4 Use the Current knob to set the parameter for the chosen F setting Current Press the Voltage knob to save the configuration setting ConF will be displayed when it is configuring Voltage VOLTAGE CURRENT VSR LAN RMT ERR DLY ALM ISR M 1 M 2 M 3 RUN C V C C V A Exit Cycle the power to save and exit the configuration settings ...

Page 23: ...ec T3PS40381P 0 001 0 6V msec T3PS60251P Falling voltage slew rate F 05 0 001 0 06V msec T3PS062001P 0 001 0 4V msec T3PS40381P 0 001 0 6V msec T3PS60251P Rising current slew rate F 06 0 001 2A msec T3PS062001P 0 001 0 38A msec T3PS40381P 0 001 0 25A msec T3PS60251P Falling current slew rate F 07 0 001 2A msec T3PS062001P 0 001 0 38A msec T3PS40381P 0 001 0 25A msec T3PS60251P Internal resistance ...

Page 24: ...20 0 None 1 Mass Storage Show rear panel USB status F 21 0 None 1 Linking to PC Setup rear USB Speed F 22 0 Disable USB 1 Full Speed 2 Auto Detect Speed SCPI Emulation F 26 0 Teledyne LeCroy 1 TDK GEN 2 Agilent 5700 3 Kikusui PWX 4 AMREL SPS LAN Settings Show MAC Address 1 F 30 0x00 0xFF Show MAC Address 2 F 31 0x00 0xFF Show MAC Address 3 F 32 0x00 0xFF Show MAC Address 4 F 33 0x00 0xFF Show MAC ...

Page 25: ...UART Data Bits F 72 0 7 bits 1 8 bits UART Parity F 73 0 None 1 Odd 2 Even UART Stop Bit F 74 0 1 bit 1 2 bits UART TCP F 75 0 SCPI 1 Emulation mode UART Address For multi unit remote control F 76 00 30 UART Multi Drop control F 77 0 Disable 1 Master 2 Slave 3 Display information UART Multi Drop status F 78 Displayed parameter AA S AA 00 30 Address S 0 1 Off line On line status System Settings Fac...

Page 26: ...when Power ON F 92 0 Safe Mode Always OFF 1 Force Mode Always ON 2 Auto Mode Status before last time power OFF Master Slave Configuration F 93 0 Independent 1 Master with 1 slave in parallel 2 Master with 2 slaves in parallel 3 Master with 3 slaves in parallel 4 Slave parallel External Output Logic F 94 0 High ON 1 Low ON Monitor Voltage Select F 96 0 5V 1 10V Control Range F 97 0 5V 5kΩ 1 10V 10k...

Page 27: ...ory number F106 1 3 M1 M3 Trigger Output Pulse Width F120 0 60ms 0 trigger output is set to the active level not pulse width Trigger Output Level F121 0 LOW 1 HIGH If F120 0 Trigger Source F122 0 None 1 Switching the output on or off 2 Changing a setting 3 Recalling a memory Special Function Settings Calibration F 00 0000 9999 Note Power On Configuration settings can only be set during power up Th...

Page 28: ...op mode 37 Multiple units Function Check 40 Configure Ethernet Connection 41 Sockets Server Configuration 41 Socket Server Function Check 43 Command Syntax 48 Command List 51 Status Register Overview 118 Introduction to the Status Registers 118 The Status Registers 119 Questionable Status Register Group 120 Operation Status Register Group 124 Standard Event Status Register Group 127 Status Byte Re...

Page 29: ... communications device class Steps 1 Connect the USB cable to the rear panel USB B port 2 Change the Rear panel USB F 22 setting to 2 Auto Detect Speed or 1 USB Full Speed Note If you are not using the rear panel USB device port set F 22 to 0 Disable USB 3 The RMT indicator will turn on when a remote connection has been established RMT indicator A V VOLTAGE CURRENT VSR LAN RMT ERR DLY ALM ISR M 1 ...

Page 30: ...System Hardware tab Run this query command via the terminal application after the instrument has been configured for USB remote control page 29 idn This should return the Manufacturer Model number Serial number and Firmware version in the following format Teledyne T3PS40381P TW123456 T0 01 12345678 Manufacturer Teledyne Model number T3PS40381P Serial number TW123456 Firmware version T0 01 12345678...

Page 31: ...wn below RS232 cable with DB9 connector DB 9 Connector Remote IN Port Remarks Pin No Name Pin No Name Housing Shield Housing Shield 2 RX 7 TX Twisted pair 3 TX 8 RX 5 SG 1 SG 5 1 9 6 1 8 RS485 cable with DB9 connector DB 9 Connector Remote IN Port Remarks Pin No Name Pin No Name Housing Shield Housing Shield 9 TXD 6 RXD Twisted pair 8 TXD 3 RXD 1 SG 1 SG 5 RXD 5 TXD Twisted pair 4 RXD 4 TXD ...

Page 32: ... Interface 0 Disable UART 1 RS232 or 2 RS485 F 71 0 7 Set the baud rate 0 1200 1 2400 2 4800 3 9600 4 19200 5 38400 6 57600 7 115200 F 72 0 or 1 Data bits 0 7 or 1 8 F 73 0 3 Parity 0 none 1 odd 2 even F 74 0 or 1 Stop bits 0 1 1 2 F 75 0 or 1 TCP 0 SCPI 1 TDK emulation mode F 76 00 30 UART address for multi unit remote connection F 77 0 3 Multi Drop control 0 Disable 1 Master 2 Slave 3 Display In...

Page 33: ...OM port No see the Device Manager in the PC For WinXP Control panel System Hardware tab Run this query command via the terminal application after the instrument has been configured for either RS232 or RS485 remote control page 31 idn This should return the Manufacturer Model number Serial number and Firmware version in the following format Teledyne T3PS40381P TW123456 T0 01 12345678 Manufacturer T...

Page 34: ...ters have to be configured in this mode The second mode allows the user to enter the SCPI commands developed for the instrument Multi Drop mode In this mode only the Multi Drop parameters have to be specified For both modes each unit is assigned a unique address and can then be individually controlled from the host PC Legacy Multi Drop mode Operation 1 Check the F 89 System version and build date ...

Page 35: ...ettings F 70 1 or 2 Configure the master unit as you normally would for RS232 or RS485 remote control see page 31 F 71 0 7 Set the baud rate set all units the same See page 31 F 72 1 Set to 8 data bits F 73 0 Parity to none F 74 0 1 Stop bit F 75 1 Set the UART TCP to TDK emulation mode F 76 00 30 Set the address of the master unit It must be a unique address identifier 7 Press the Function key to...

Page 36: ...ultiple units can now be operated at the same time Only legacy commands can be used in this mode See the programming manual or see the function check below for usage details Slave serial link cable with RJ 45 shielded connectors RS 485 slave serial link pin assignment 8 Pin Connector IN 8 Pin Connector OUT Pin No Name Pin No Name Housing Shield Housing Shield 1 SG 1 SG 6 TXD 6 TXD 3 TXD 3 TXD 5 RX...

Page 37: ...ct the OUT port on the first unit to the IN port of the second unit using the master serial link cable 5 Connect all the remaining units between the OUT port and the IN port with the slave serial link cable until all the desired units have been daisy chained together RS 485 232 OUT Unit 1 RS 485 232 OUT IN Unit 2 RS 485 232 OUT IN Unit N End terminal connector To PC Master serial link cable gray p...

Page 38: ...eck the slaves addresses by using the F 77 parameter on the master unit F 77 3 Display on each slave units the configured address This can show if identical addresses have been assigned individually to each slave units 13 Set the Multi Drop setting parameter F 77 to Master F 77 1 Set the Multi Drop setting to master 14 You can display the status of each slave unit by using the F 78 parameter F 78 ...

Page 39: ...o Name Pin No Name Housing Shield Housing Shield 1 SG 1 SG 6 TXD 6 TXD 3 TXD 3 TXD 5 RXD 5 RXD 4 RXD 4 RXD Master serial link cable with RJ 45 shielded connectors RS 485 master serial link pin assignment 8 Pin Connector IN 8 Pin Connector OUT Pin No Name Pin No Name Housing Shield Housing Shield 1 SG 1 SG 6 TXD 5 RXD 3 TXD 4 RXD 5 RXD 6 TXD 4 RXD 3 TXD 1 8 ...

Page 40: ...ion Check Functionality check Invoke a terminal application such as Realterm To check the COM port No see the Device Manager in the PC For WinXP Control panel System Hardware tab Below shows examples using the Legacy Multi Drop mode and the Multi Drop mode ...

Page 41: ...onfigured Ethernet configuration Parameters For details on how to configure the Ethernet settings please see the configuration chapter on page 18 MAC Address display only LAN Enable Disable DHCP Enable Disable IP Address Subnet Mask Gateway DNS Address Sockets Server Enable Disable Sockets Server Configuration Configuration This configuration example will configure the T3PS socket server The follo...

Page 42: ... 172 IP Address part 1 of 4 F 40 16 IP Address part 2 of 4 F 41 5 IP Address part 3 of 4 F 42 133 IP Address part 4 of 4 F 43 255 Subnet Mask part 1 of 4 F 44 255 Subnet Mask part 2 of 4 F 45 128 Subnet Mask part 3 of 4 F 46 0 Subnet Mask part 4 of 4 F 47 172 Gateway part 1 of 4 F 48 16 Gateway part 2 of 4 F 49 21 Gateway part 3 of 4 F 50 101 Gateway part 4 of 4 F 57 1 Enable Sockets ...

Page 43: ...or the VISA Run time Engine page or downloads at the following URL http www ni com visa Requirements Operating System Windows XP 7 8 Functionality check 1 Start the NI Measurement and Automation Explorer MAX program Using Windows press Start All Programs National Instruments Measurement Automation 2 From the Configuration panel access My System Devices and Interfaces Network Devices 3 Press Add Ne...

Page 44: ...t Manual Entry of Raw Socket from the popup window 4 5 Enter the IP address and the port number of the T3PS The port number is fixed at 2268 6 Click the Validate button 7 A popup will appear if a connection is successfully established 8 Click Next ...

Page 45: ... Alias name of the T3PS connection In this example the Alias is T3PS_DC1 10 Click finish 9 10 T3PS_DC1 11 The IP address of the T3PS will now appear under Network Devices in the configuration panel Select this icon now 12 Click Open VISA Test Panel ...

Page 46: ... O Settings 15 Make sure the Enable Termination Character check box is checked and the terminal character is n Value xA 16 Click Apply Changes 14 13 15 16 17 Click the Input Output icon 18 Enter IDN in the Select or Enter Command dialog box if it is not already 19 Click the Query button ...

Page 47: ...NTROL 47 20 The IDN query will return the Manufacturer model name serial number and firmware version in the dialog box Teledyne T3PS40381P TW123456 T0 02 20131205 18 17 19 20 Teledyne T3PS40381P T0 02 20131205 n ...

Page 48: ...h node in the command tree Each keyword node of a SCPI command is separated by a colon For example the diagram below shows an SCPI sub structure and a command example DC MEASure SCALar POWer VOLTage CURRent DC DC MEASure SCALar CURRent DC Command types There are a number of different instrument commands and queries A command sends instructions or data to the unit and a query receives data or statu...

Page 49: ... command must begin at the last node of the first command A semi colon and colon are used to combine two commands from different nodes Example meas volt dc meas curr dc Command Forms Commands and queries have two different forms long and short The command syntax is written with the short form of the command in capitals and the remainder long form in lower case The commands can be written in capita...

Page 50: ...ow Both DISPlay MENU NAME and DISPlay MENU are both valid forms Command Format 1 5 5 2 1 2 3 4 5 APPLY 1 Command header 2 Space 3 Parameter 1 4 Comma no space before after comma 5 Parameter 2 Parameters Type Description Example Boolean Boolean logic 0 1 NR1 integers 0 1 2 3 NR2 decimal numbers 0 1 3 14 8 5 NR3 floating point 4 5e 1 8 25e 1 NRf any of NR1 2 3 1 1 5 4 5e 1 block data Definitive leng...

Page 51: ...ITiate IMMediate TRANsient 59 Instrument Commands INSTrument SCAN 60 INSTrument SELect 60 INSTrument STATe 60 INSTrument DISPlay 61 Measure Commands MEASure SCALar ALL DC 62 MEASure SCALar CURRent DC 62 MEASure SCALar VOLTage DC 62 MEASure SCALar POWer DC 63 Memory MEMory TRIGgered 64 Output Commands OUTPut DELay ON 65 OUTPut DELay OFF 65 OUTPut MODE 66 OUTPut STATe IMMediate 66 OUTPut STATe TRIGg...

Page 52: ...ionable INSTrument ISUMmary n CONDition 72 STATus QUEStionable INSTrument ISUMmary n ENABle 73 STATus PRESet 73 Source Commands SOURce CURRent LEVel IMMediate AMPLitude 75 SOURce CURRent LEVel TRIGgered AMPLitude 76 SOURce CURRent LIMit AUTO 76 SOURce CURRent PROTection DELay 77 SOURce CURRent PROTection LEVel 77 SOURce CURRent PROTection STATe 78 SOURce CURRent PROTection TRIPped 78 SOURce CURRen...

Page 53: ...r INPut SOURce 95 SYSTem CONFigure TRIGger INPut WIDTh 96 SYSTem CONFigure TRIGger OUTPut SOURce 96 SYSTem CONFigure TRIGger OUTPut WIDTh 97 SYSTem CONFigure TRIGger OUTPut LEVel 97 SYSTem COMMunicate ENABle 98 SYSTem COMMunicate GPIB SELF ADDRess 99 SYSTem COMMunicate LAN IPADdress 99 SYSTem COMMunicate LAN GATeway 100 SYSTem COMMunicate LAN SMASk 100 SYSTem COMMunicate LAN MAC 101 SYSTem COMMuni...

Page 54: ... MODE 108 SYSTem ERRor ENABle 109 SYSTem LANGuage EMULation 109 SYSTem LANGuage SELect 109 SYSTem PRESet 110 SYSTem VERSion 110 SYSTem REBoot 110 Trigger Commands TRIGger OUTPut SOURce 111 TRIGger OUTPut IMMediate 111 TRIGger TRANsient SOURce 112 TRIGger TRANsient IMMediate 112 Common Commands CLS 113 ESE 113 ESR 114 IDN 114 OPC 114 RCL 115 RST 115 SAV 115 SRE 116 STB 116 TRG 116 TST 116 WAI 117 ...

Page 55: ...ommand sets the voltage and current at the same time Syntax Query Syntax APPLy NRf V MINimum MAXimum NRf A MINimu m MAXimum APPLy Parameter Return parameter NRf V Voltage setting MINimum Minimum voltage level MAXimum Maximum voltage level NRf A Current setting MINimum Minimum voltage level MAXimum Maximum voltage level Example APPL MIN MIN Sets the current and voltage to the minimum settings ...

Page 56: ...n menu Syntax Query Sytax DISPlay MENU NAME NR1 DISPlay MENU NAME Parameter Return parameter NR1 Description 0 Measure voltage current 1 2 Not Used 3 Set Menu 4 OVP OCP Menu 5 99 Not Used 100 199 F 00 99 Menu 200 229 F100 F129 Menu Example DISP MENU NAME 0 Sets the display to the Voltage Current display screen DISPlay WINDow TEXT CLEar Set Description Clears the text on the main screen from the DI...

Page 57: ...EXT DATA string DISPlay WINDow TEXT DATA Parameter Return parameter string ASCII character 20H to 7EH can be used to in the string parameter The string must be enclosed in quotes STRING Example DISP WIND TEXT DATA STRING Writes STRING to the display Query Example DISP WIND TEXT DATA STRING Returns the text data string on the screen DISPlay BLINk Set Query Description Turns blink on or off for the ...

Page 58: ...Syntax Query Syntax INITiate CONTinuous TRANsient bool OFF ON INITiate CONTinuous TRANsient Parameter OFF 0 OFF ON 1 ON Return parameter 0 OFF 1 ON Example INIT TRAN 1 Turns on the continuous trigger INITiate IMMediate NAME Set Description The INITiate command starts the TRANsient or OUTPut trigger Syntax INITiate IMMediate NAME TRANsient OUTPut Parameter TRANSient Starts the TRANsient trigger OUT...

Page 59: ...t Description This command controls the enabling of output triggers When a trigger is enabled a trigger causes the specified action to occur If the trigger system is not enabled all triggers are ignored Syntax INITiate IMMediate TRANsient Example INIT ...

Page 60: ...Specifies the address of the unit to which communication will be established when using the Multi Drop mode Syntax Query Syntax INSTrument SELect NR1 INSTrument SELect Parameter NR1 The address of the unit to be selected 0 30 Return parameter NR1 The currently selected address Example INST SEL 30 The currently selected address is 30 INSTrument STATe Query Description Displays the status on line of...

Page 61: ...et to 1 when the corresponding unit is on line Second value This value represents the master address Example INST STAT 33 0 33 0b100001 The units at address 0 and address 5 are on line 0 Master device s address is 0 INSTrument DISPlay Set Description Displays information configured address for all slave units when using the Multi Drop mode Syntax INSTrument DISPlay Example INST DISP ...

Page 62: ...ASure SCALar ALL DC Return parameter 0 0000 0 0000 voltage current Returns the voltage V and current A respectively MEASure SCALar CURRent DC Query Description Takes a measurement and returns the average output current Syntax MEASure SCALar CURRent DC Return parameter 0 0000 Returns the current in amps MEASure SCALar VOLTage DC Query Description Takes a measurement and returns the average output v...

Page 63: ...REMOTE CONTROL 63 MEASure SCALar POWer DC Query Description Takes a measurement and returns the average output power Syntax MEASure SCALar POWer DC Return 0 0000 Returns the power measured in watts ...

Page 64: ...rigger input is received and the trigger input is configured to load a memory setting Related Commands SYSTem CONFigure TRIGger INPut SOURce SYSTem CONFigure TRIGger OUTPut SOURce Syntax Return Syntax MEMory TRIGgered NR1 MINimum MAXimum MEMory TRIGgered MINimum MAXimum Parameter NR1 0 M1 2 M3 MINimum MAXimum Return Parameter NR1 Returns the memory setting ...

Page 65: ...ntax OUTPut DELay ON NR2 MINimum MAXimum OUTPut DELay ON Parameter NR2 0 00 99 99 seconds where 0 no delay Return parameter 0 00 Returns the delay on time in seconds until the output is turned on OUTPut DELay OFF Set Query Description Sets the Delay Time in seconds for turning the output off The delay is set to 0 00 by default Syntax Return Syntax OUTPut DELay OFF NR2 MINimum MAXimum OUTPut DELay ...

Page 66: ...ameter NR1 Returns the output mode OUTPut STATe IMMediate Set Query Description Turns the output on or off Syntax Query Syntax OUTPut STATe IMMediate bool OFF ON OUTPut STATe IMMediate Parameter OFF 0 Turns the output off ON 1 Turns the output on Return parameter bool Returns output status of the instrument OUTPut STATe TRIGgered Set Query Description Turns the output on or off when a software tri...

Page 67: ...ge over current and over temperature OVP OCP OTP protection circuits It also clears the shutdown and sense protection circuit The AC failure protection cannot be cleared Syntax OUTPut PROTection CLEar OUTPut PROTection TRIPped Query Description Queries the unit to see if a protection circuit has been tripped Syntax OUTPut PROTection TRIPped Return boolean 0 No protection error 1 A protection error...

Page 68: ...UNt Set Query Description Sets or queries the level of smoothing for the average setting Syntax Return Syntax SENSe AVERage COUNt NR1 LOW MIDDle HIGH SENSe AVERage COUNt Parameter LOW 0 Low setting MIDDle 1 Middle setting HIGH 2 High setting Return Parameter NR1 Returns the average setting ...

Page 69: ...Tus QUEStionable PTRansition 71 STATus QUEStionable NTRansition 72 STATus QUEStionable INSTrument ISUMmary n EVENt 72 STATus QUEStionable INSTrument ISUMmary n CONDition 72 STATus QUEStionable INSTrument ISUMmary n ENABle 73 STATus PRESet 73 STATus OPERation EVENt Query Description Queries the Operation Status Event register and clears the contents of the register Syntax STATus OPERation EVENt Ret...

Page 70: ...ATus OPERation PTRansition Set Query Description Sets or queries the bit sum of the positive transition filter of the Operation Status register Syntax STATus OPERation PTRansition NR1 STATus OPERation PTRansition Parameter NR1 0 32767 Return parameter NR1 0 32767 STATus OPERation NTRansition Set Query Description Sets or queries the bit sum of the negative transition filter of the Operation Status...

Page 71: ...egister Query Syntax STATus QUEStionable CONDition Return parameter NR1 0 32767 STATus QUEStionable ENABle Set Query Description Sets or queries the bit sum of the Questionable Status Enable register Syntax Query Syntax STATus QUEStionable ENABle NR1 STATus QUEStionable ENABle Parameter NR1 0 32767 Return parameter NR1 0 32767 STATus QUEStionable PTRansition Set Query Description Sets or queries t...

Page 72: ...t Query Description Queries the bit sum of the Questionable Instrument Summary Status Event register This query will also clear the contents of the register Multi Drop mode Query Syntax STATus QUEStionable INSTrument ISUMmary n EVENt Parameter n 1 2 or 3 Return parameter NR1 0 32767 STATus QUEStionable INSTrument ISUMmary n CONDition Query Description Queries the status bit sum of the Questionable...

Page 73: ...rn parameter NR1 0 32767 STATus PRESet Set Description This command resets the ENABle register the PTRansistion filter and NTRansistion filter on the Operation Status and Questionable Status Registers The registers filters will be reset to a default value Default Register Filter Values Setting QUEStionable Status Enable 0x0000 QUEStionable Status Positive Transition 0x7FFF QUEStionable Status Nega...

Page 74: ...able Status Enable registers and the Operation Status Enable registers are both reset to 0 The Questionable Status and Operation Status Positive Transition filters are all set high 0x7FFF and the Negative Transition filters are all set low 0x0000 I e only positive transitions will be recognized for the Questionable Status and Operation Status registers Syntax STATus PRESet ...

Page 75: ...SOURce VOLTage LEVel TRIGgered AMPLitude 81 SOURce VOLTage LIMit AUTO 82 SOURce VOLTage LIMit LOW 82 SOURce VOLTage PROTection LEVel 83 SOURce VOLTage PROTection TRIPped 83 SOURce VOLTage SLEWrate RISing 84 SOURce VOLTage SLEWrate FALLing 84 SOURce CURRent LEVel IMMediate AMPLitude Set Query Description Sets or queries the current level in amps For externally set current levels from the analog con...

Page 76: ...gered AMPLitude Parameter NR2 0 105 of the rated current output in amps MIN Minimum current level MAX Maximum current level Return Parameter NR2 Returns the current level Example SOUR CURR LEV TRIG AMPL 38 000 Returns the maximum possible current level in amps SOURce CURRent LIMit AUTO Set Query Description Enables or disables the limit on the current setting Syntax SOURce CURRent LIMit AUTO bool ...

Page 77: ...d delay time MIN The minimum allowed delay time Return parameter NR2 Returns the delay time in seconds Example SOUR CURR PROT DEL MAX Sets the current protection delay to the maximum SOURce CURRent PROTection LEVel Set Query Description Sets or queries the OCP over current protection level in amps Syntax SOURce CURRent PROTection LEVel NR2 A MINimum MAXimum Query Syntax SOURce CURRent PROTection L...

Page 78: ...bool OFF ON Query Syntax SOURce CURRent PROTection STATe Parameter OFF 0 Turns the OCP off ON 1 Turns the OCP on Return parameter bool Returns the over current protection state in bool format Example SOUR CURR PROT STAT OFF Turns OCP off SOURce CURRent PROTection TRIPped Set Query Description Returns the state of the current protection circuits Query Syntax SOURce CURRent PROTection TRIPped Return...

Page 79: ...vided by 100msec Return parameter NR2 Returns the step current in amps Example SOUR CURR SLEW RIS 0 950 Sets the rising current slew rate to 0 950 A ms SOURce CURRent SLEWrate FALLing Set Query Description Sets or queries the falling current slew rate This is only applicable for CC slew rate priority mode Syntax SOURce CURRent SLEWrate FALLing NR2 A MINimum MAXimum Query Syntax SOURce CURRent SLEW...

Page 80: ...ce MODE Return parameter string Returns the output state as a string CC CV OFF Example SOUR MODE CC The power supply is currently in CC mode SOURce RESistance LEVel IMMediate AMPLitude Set Query Description Sets or queries the internal resistance in ohms Syntax SOURce RESistance LEVel IMMediate AMPLitude NR2 OHM MINimum MAXimum Query Syntax SOURce RESistance LEVel IMMediate AMPLitude Parameter NR2...

Page 81: ...oltage level MAX Maximum voltage level Return parameter NR2 Returns the voltage level in volts Example SOUR VOLT LEV IMM AMPL 10 Sets the voltage level to 10 volts SOURce VOLTage LEVel TRIGgered AMPLitude Set Query Description Sets or queries the voltage level in volts when a software trigger has been generated Syntax SOURce VOLTage LEVel TRIGgered AMPLitude NR2 V MINimum MAXimum Query Syntax SOUR...

Page 82: ...e UVL setting is higher than the voltage setting the UVL setting will be set equal to the voltage setting Syntax SOURce VOLTage LIMit AUTO bool OFF ON Query Syntax SOURce VOLTage LIMit AUTO Parameter OFF 0 Disable the limit setting ON 1 Enable the limit setting Return parameter bool Returns the setting in bool format Example SOUR VOLT LIM AUTO 0 Disables the limit setting SOURce VOLTage LIMit LOW ...

Page 83: ...um Depends on the unit type if Vrated 0 1 5V then Minimum 5V else Minimum Vrated 0 1 Maximum Vrated 1 1 MIN Minimum OVP level MAX Maximum OVP level Example SOUR VOLT PROT LEV MAX Sets the OVP level to its maximum SOURce VOLTage PROTection TRIPped Set Query Description Sets or queries the overvoltage protection level Query Syntax SOURce VOLTage PROTection TRIPped Return parameter bool 0 Protection ...

Page 84: ...ivided by 100msec Return parameter NR2 Returns the slew rate in V msec Example SOUR VOLT SLEW RIS MAX Sets the rising voltage slew rate to its maximum SOURce VOLTage SLEWrate FALLing Set Query Description Sets or queries the falling voltage slew rate This is only applicable for CV slew rate priority mode Syntax SOURce VOLTage SLEWrate FALLing NR2 V MINimum MAXimum Query Syntax SOURce VOLTage SLEWr...

Page 85: ...REMOTE CONTROL 85 Example SOUR VOLT SLEW FALL MIN Sets the falling voltage slew rate to its minimum ...

Page 86: ...96 SYSTem CONFigure TRIGger OUTPut WIDTh 97 SYSTem CONFigure TRIGger OUTPut LEVel 97 SYSTem COMMunicate ENABle 98 SYSTem COMMunicate GPIB SELF ADDRess 99 SYSTem COMMunicate LAN IPADdress 99 SYSTem COMMunicate LAN GATeway 100 SYSTem COMMunicate LAN SMASk 100 SYSTem COMMunicate LAN MAC 101 SYSTem COMMunicate LAN DHCP 101 SYSTem COMMunicate LAN DNS 102 SYSTem COMMunicate RLSTate 102 SYSTem COMMunicat...

Page 87: ...s Syntax Query Syntax SYSTem BEEPer IMMediate NR1 MINimum MAXimum SYSTem BEEPer IMMediate MINimum MAXimum Parameter NR1 0 3600 seconds MINimum Sets the beeper time to the minimum 0 seconds MAXimum Sets the beeper time to the maximum 3600 seconds Return parameter NR1 Returns the remaining beeper duration time in seconds or returns the maximum or minimum beeper time in seconds for the MINimum MAXimu...

Page 88: ...s SYSTem CONFigure BLEeder STATe Set Query Description Sets or queries the status of the bleeder resistor Syntax Query Syntax SYSTem CONFigure BLEeder STATe NR1 OFF ON AUTO SYSTem CONFigure BLEeder STATe Parameter OFF 0 Turns the bleeder resistor off ON 1 Turns the bleeder resistor on AUTO 2 Turn the AUTO mode on Return parameter NR1 Returns bleeder resistor status SYSTem CONFigure CURRent CONTrol...

Page 89: ... 0kΩ Io max 4 VISolation External voltage control isolated The resistance value depends on the SYSTem CONFigure CONTrol RANGe command If the range is high then the resistance is 10kΩ else it is 5kΩ Return Parameter NR1 Returns the current control configuration Example SYST CONF CURR CONT VOLT SYST REB Sets the current control by external voltage and reboot the unit to active the setting SYSTem CON...

Page 90: ...rol 2 RRISing External resistance control 10kΩ or 5kΩ Io max 0kΩ Io min 3 RFALling External resistance control 10kΩ or 5kΩ Io min 0kΩ Io max 4 VISolation External voltage control isolated The resistance value depends on the SYSTem CONFigure CONTrol RANGe command If the range is high then the resistance is 10kΩ else it is 5kΩ Return Parameter NR1 Returns the voltage control configuration Example SY...

Page 91: ...rns on in the same state the unit was in prior to the previous shut down The output is set to off default FORCe ON 1 The T3PS turns on in the same state the unit was in prior to the previous shut down The output is set to on AUTO 2 The T3PS turns on in the same state the unit was in prior to the previous shut down but with the same output on off setting Return parameter 0 The power on output setti...

Page 92: ... fixed AUTO The output is turned on automatically when the cause of the alarm is fixed SYSTem CONFigure MSLave Set Query Description Sets or queries the unit operation mode Note This is the equivalent to the F 93 Master Slave Configuration power on configuration settings Syntax Query Syntax SYSTem CONFigure MSLave NR1 SYSTem CONFigure MSLave Parameter Return parameter NR1 Description 0 Master Loca...

Page 93: ...en cycled Syntax Return Syntax SYSTem CONFigure OUTPut EXTernal MODE NR1 LOW HIGH SYSTem CONFigure OUTPut EXTernal MODE Parameter HIGH 0 Active high LOW 1 Active low Return Parameter NR1 Returns the logic setting Example SYST CONF OUTP EXT MODE 0 SYST REB Set active high for the logic for the external analog control contact and reboot the unit to active the setting SYSTem CONFigure OUTPut EXTernal...

Page 94: ...al output control and reboot the unit to active the setting SYSTem CONFigure MONitor RANGe Set Query Description This command is used to select the monitor voltage range This is the equivalent to the F 96 Monitor Voltage Select power on configuration settings Note The setting will only be valid after the power has been cycled Syntax Return Syntax SYSTem CONFigure MONitor RANGe NR1 LOW HIGH SYSTem ...

Page 95: ...5V 5kΩ HIGH 1 10V 10kΩ Return Parameter NR1 Returns the range setting Example SYST CONF CONT RANG LOW SYST REB Set external analog control voltage or resistance range to 0 5V 5kΩ and reboot the unit to active the setting SYSTem CONFigure TRIGger INPut SOURce Set Query Description Sets or queries what action will be performed on receiving a trigger Syntax Return Syntax SYSTem CONFigure TRIGger INPu...

Page 96: ...RIGger INPut WIDTh NR2 MINimum MAXimum SYSTem CONFigure TRIGger INPut WIDTh MINimum MAXimum Parameter NR2 0 60ms MINimum Minimum width 0 MAXimum 60ms Return Parameter NR2 Returns the trigger input width SYSTem CONFigure TRIGger OUTPut SOURce Set Query Description Sets or queries the output trigger source Syntax Return Syntax SYSTem CONFigure TRIGger OUTPut SOURce NR1 NONE OUTPut SETTing MEMory SYS...

Page 97: ...ONFigure TRIGger OUTPut WIDTh NR2 MINimum MAXimum SYSTem CONFigure TRIGger OUTPut WIDTh MINimum MAXimum Parameter NR2 0 60ms MINimum Minimum width 0 MAXimum 60ms Return Parameter NR2 Returns the trigger output width SYSTem CONFigure TRIGger OUTPut LEVel Set Query Description Sets the polarity of the output trigger level when the output trigger pulse width is set to 0 Syntax Return Syntax SYSTem CO...

Page 98: ...l SYSTem COMMunicate ENABle GPIB USB LAN SOCKets WEB SERial Parameter 1 OFF 0 Disables the selected interface ON 1 Enables the selected interface Parameter 2 GPIB Select GPIB USB Select USB LAN Select LAN SOCKets Select Sockets WEB Select the web server SERial Selected Serial UART Return Parameter bool Returns the status of the selected mode Example SYST COMM ENAB 1 USB SYST REB Turns the USB inte...

Page 99: ... Sets the GPIB address to 15 and reboot the unit to active the setting SYSTem COMMunicate LAN IPADdress Set Query Description Sets or queries LAN IP address Note The setting will only be valid after the power has been cycled Syntax Query Syntax SYSTem COMMunicate LAN IPADdress string SYSTem COMMunicate LAN IPADdress Parameter Return string LAN IP address in string format address Applicable ASCII c...

Page 100: ...xample SYST COMM LAN GATe 172 16 0 254 SYST REB Sets the LAN gateway to 172 16 0 254 and reboot the unit to active the setting SYSTem COMMunicate LAN SMASk Set Query Description Sets or queries the LAN subnet mask Note The setting will only be valid after the power has been cycled Syntax Query Syntax SYSTem COMMunicate LAN SMASk string SYSTem COMMunicate LAN SMASk Parameter Return string Subnet ma...

Page 101: ...MM LAN MAC 02 80 AD 20 31 B1 Returns the MAC address SYSTem COMMunicate LAN DHCP Set Query Description Turns DHCP on off Queries the DHCP status Note The setting will only be valid after the power has been cycled Syntax Query Syntax SYSTem COMMunicate LAN DHCP bool OFF ON SYSTem COMMunicate LAN DHCP Parameter OFF 0 DHCP off ON 1 DHCP on Return parameter bool Returns the DHCP status Example SYST CO...

Page 102: ...ets the DNS to 172 16 1 252 reboot the unit to active the setting SYSTem COMMunicate RLSTate Set Query Description Enables or disables local remote state of the instrument Syntax Query Syntax SYSTem COMMunicate RLSTate LOCal REMote RWLock SYSTem COMMunicate RLSTate Parameter Return parameter LOCal All keys are valid This instrument is controlled by the front panel controls REMote All keys are inva...

Page 103: ...r queries the communication protocol for the serial port Syntax Query Syntax SYSTem COMMunicate SERial LANGuage SELect SCPI LEGACY SYSTem COMMunicate SERial LANGuage SELect Parameter Return parameter SCPI Sets the communication protocol to SCPI LEGACY Sets the communication protocol to legacy mode Emulate TDK Genesys Example SYST COMM SER LANG SCPI Indicates that the communication protocol is set ...

Page 104: ...AN BAUD 2400 Returns the baud rate settings SYST COMM SER TRAN BAUD 9600 SYST REB Set the UART baud rate to 9600bps and reboot the unit to active the setting SYSTem COMMunicate SERial RECeive TRANsmit BITS Set Query Description Sets or queries the UART number of data bits Note The setting will only be valid after the power has been cycled Syntax Query Syntax SYSTem COMMunicate SERial RECeive TRANs...

Page 105: ...iption Sets or queries the parity of the UART connection Note The setting will only be valid after the power has been cycled Syntax Query Syntax SYSTem COMMunicate SERial RECeive TRANsmit PARity NR1 SYSTem COMMunicate SERial RECeive TRANsmit PARity Parameter Return parameter 0 None 1 Odd 2 Even Example SYST COMM SER TRAN PARity 1 Indicates that odd parity is used for the UART connection SYST COMM ...

Page 106: ...s NR1 SYSTem COMMunicate SERial RECeive TRANsmit SBITs Parameter Return parameter 0 1 stop bit 1 2 stop bits Example SYST COMM SER TRAN SBITs 1 Indicates that one stop bit is used for the UART connection SYST COMM SER TRAN SBITs 1 SYST REB Sets the UART stop bits to 2 stop bits and reboot the unit to active the setting SYSTem COMMunicate SERial MULTidrop CONTrol Query Description Queries the Multi...

Page 107: ...s the speed of the rear panel USB B port Note The setting will only be valid after the power has been cycled Syntax SYSTem COMMunicate USB REAR MODE NR1 DISable AUTO FULL Query Syntax SYSTem COMMunicate USB REAR MODE Parameter 0 DISable Disable 1 AUTO Auto detect speed 2 FULL Full speed Return parameter NR1 0 Disable 1 Auto detect speed 2 Full speed Example SYST COMM USB REAR MODE 2 SYST REB Sets ...

Page 108: ... the error queue Query Syntax SYSTem ERRor Return parameter string Returns an error code followed by an error message as a single string Example SYSTem ERRor 100 Command error SYSTem KLOCk Set Query Description Enables or disables the front panel key lock Syntax Query Syntax SYSTem KLOCk bool OFF ON SYSTem KLOCk Parameter OFF 0 Panel keys unlocked ON 1 Panel keys locked Return parameter bool Retur...

Page 109: ...ERRor ENABle SYSTem LANGuage EMULation Set Query Description Sets or queries the command language Syntax Query Syntax SYSTem LANGuage EMULation NONE N5700 GENSYS PWX SYSTem LANGuage EMULation Parameter Return parameter NONE Emulation is not used This is the default setting N5700 N5700 N8700 emulation is used GENSYS GENESYS emulation is used PWX PAG emulation is used SYSTem LANGuage SELect Set Quer...

Page 110: ...nguage SYSTem PRESet Set Description Loads the preset default settings Syntax SYSTem PRESet SYSTem VERSion Query Description Returns the version of the T3PS SCPI version Query Syntax SYSTem VERSion Return string Returns the SCPI version as a string Query Example SYST VERS 1999 9 SYSTem REBoot Set Description Reboots the T3PS system Syntax SYSTem REBoot ...

Page 111: ...Mediate EXTernal Query Syntax TRIGger OUTPut SOURce Parameter Return parameter BUS Output trigger is generated by the bus IMMediate Output trigger is immediately generated EXTernal The output trigger is generated when an external signal triggers it Example TRIGger OUTPut SOURce EXT Sets the output trigger source to EXT TRIGger OUTPut IMMediate Set Description Generates an immediate trigger for the...

Page 112: ...e Parameter Return parameter BUS Transient trigger is generated by the bus IMMediate Transient trigger is immediately generated EXTernal The transient trigger is generated when an external signal triggers it Example TRIG SOUR EXT Sets the transient trigger source to EXT TRIGger TRANsient IMMediate Set Description Generates an immediate trigger for the transient trigger system Syntax TRIGger TRANsi...

Page 113: ...LS Set Description The CLS command clears all the event registers including the status byte event status and error queue Syntax CLS ESE Set Query Description Sets or queries the Standard Event Status Enable register Syntax Query Syntax ESE NR1 ESE Parameter NR1 0 255 Return parameter NR1 Returns the bit sum of the Standard Event Status Enable register ...

Page 114: ...l number and firmware version of the T3PS Query Syntax IDN Return parameter string Returns the instrument identification as a string in the following format Teledyne T3PS40381P TW123456 T0 01 12345678 Manufacturer Teledyne Model number T3PS40381P Serial number TW123456 Firmware version T0 01 12345678 OPC Set Query Description The OPC command sets the OPC bit bit0 of the Standard Event Status Regis...

Page 115: ... the M1 memory contents MAX Recalls the M3 memory contents RST Set Description Performs a device reset Configures the unit to a known configuration preset default settings This known configuration is independent of the usage history Syntax RST SAV Set Description Saves the settings into memory slot M1 M2 or M3 Syntax SAV NR1 MIN MAX Return parameter NR1 0 1 2 as memory M1 M2 M3 MIN Saves the M1 me...

Page 116: ...he Service Request Enable register STB Query Description Queries the bit sum of the Status Byte register with MSS Master summary Status replacing the RQS bit bit 6 Query Syntax STB Return parameter NR1 Returns the bit sum of the Status Byte register with the MSS bit bit 6 TRG Set Description The TRG command is able to generate a get Group Execute Trigger If the T3PS cannot accept a trigger at the ...

Page 117: ...rameter 0 Returns 0 if there are no errors NR1 Returns an error code NR1 if there is an error WAI Set Description Prevents any other commands or queries from being executed until all outstanding commands have completed Syntax WAI ...

Page 118: ...us registers are used to determine the status of the power supply The status registers maintain the status of the protection conditions operation conditions and instrument errors The T3PS Series have a number of register groups Questionable Status Register Group Standard Event Status Register Group Operation Status Register Group Status Byte Register Service Request Enable Register Service Request...

Page 119: ...ng for Trigger 5 Not Used 6 Not Used 7 CV Constant Voltage Mode 8 CP Constant Power Mode 9 CC Constant Current Mode 10 OND Output ON Delay 11 OFD Output OFF Delay 12 Not Used 13 PR Program Running 14 Not Used 15 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 OPC Operation Complete RQC Request Control QUE Query Error DDE Device Depe...

Page 120: ...5 6 7 8 9 10 11 12 13 14 15 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 Questionable Status Register Logical OR To Status Byte Register STATus QUEStionable CONDition STATus QUEStionable EVENt STATus QUEStionable NTRansition NRf STATus QUEStionable NTRansition STATus QUEStionable PTRansition NRf STATus QUEStionable PTRansition STATus QUEStionable ENABle NRf STATus QUEStionable ENABle Bit Summary Event Bi...

Page 121: ...6384 Condition Register The Questionable Status Condition Register indicates the status of the power supply If a bit is set in the Condition register it indicates that the event is true Reading the condition register does not change the state of the condition register PTR NTR Filters The PTR NTR Positive Negative transition register determines the type of transition conditions that will set the co...

Page 122: ... the type of transition conditions will set the corresponding bits in the Event Register If the Event Register is read it will be cleared to 0 Enable Register The Enable register determines which Events in the Event Register will be used to set the QUES bit in the Status Byte Register ...

Page 123: ... SRQ20 7 7 7 SRQ21 8 8 8 SRQ22 9 9 9 SRQ23 10 10 10 SRQ24 11 11 11 SRQ25 12 12 12 SRQ26 13 13 13 SRQ27 14 14 14 0 15 15 15 STATus QUEStionable ISUMmary2 CONDition STATus QUEStionable ISUMmary2 ENABle NRf STATus QUEStionable ISUMmary2 ENABle CONDITION EVENT ENABLE ISUM2 0 0 0 SRQ0 1 1 1 SRQ1 2 2 2 SRQ2 3 3 3 SRQ3 4 4 4 SRQ4 5 5 5 SRQ5 6 6 6 SRQ6 7 7 7 SRQ7 8 8 8 SRQ8 9 9 9 SRQ9 10 10 10 SRQ10 11 11...

Page 124: ...2 3 4 5 6 7 8 9 10 11 12 13 14 15 Operation Status Register Logical OR To Status Byte Register STATus OPERation CONDition STATus OPERation EVENt STATus OPERation NTRansition NRf STATus OPERation NTRansition STATus OPERation PTRansition NRf STATus OPERation PTRansition STATus OPERation ENABle NRf STATus OPERation ENABle Bit Summary Event Bit Bit Weight CAL Calibration mode Indicates if the T3PS is ...

Page 125: ...ng Indicates if a Test is running 14 16384 Condition Register The Operation Status Condition Register indicates the operating status of the power supply If a bit is set in the Condition register it indicates that the event is true Reading the condition register does not change the state of the condition register PTR NTR Filters The PTR NTR Positive Negative transition register determines the type ...

Page 126: ...ister will dictate the type of transition conditions will set the corresponding bits in the Event Register If the Event Register is read it will be cleared to 0 Enable Register The Enable register determines which registered Events in the Event Register will be used to set the OPER bit in the Status Byte Register ...

Page 127: ... Status Register To Status Byte Register ESR Logical OR ESE NRf ESE Bit Summary Event Bit Bit Weight OPC Operation complete The OCP bit is set when all selected pending operations are complete This bit is set in response to the OPC command 0 1 RQC Request control 1 2 QUE Query Error The Query Error bit is set in response to an error reading the Output Queue This can be caused by trying to read the...

Page 128: ... bit is set when a syntax error has occurred The CME bit can also be set when a GET command is received within a program message 5 32 URQ User Request 6 64 PON Power On Indicates the power is turned on 7 128 Event Register Any bits set in the event register indicate that an error has occurred Reading the Event register will reset the register to 0 Enable Register The Enable register determines whi...

Page 129: ...rror Que Status Byte Register 0 1 2 3 4 5 6 7 SRE NRf SRE Service Request Enable Register Logical OR STB From Operation Status Register From Questionable Status Register From Standard Event Status Register Bit Summary Event Bit Bit Weight ERR Error Event Queue If data is present in the Error queue the ERR bit will be set 2 4 QUES Questionable Status Register The summary bit for the Questionable St...

Page 130: ... the summary bit for the Operation Status Register Group 7 128 Status Byte Register Any bits set in the Status byte register acts as a summary register for all the three other status registers and indicates if there is a service request an error in the Error Queue or data in the Output Queue Reading the Status Byte register will reset the register to 0 Service Request Enable Register The Service R...

Page 131: ...ntax error has been detected by the parser That is a controller to device message was received which is in violation of the IEEE 488 2 standard Possible violations include a data element which violates the device listening formats or whose type is unacceptable to the device An unrecognized header was received Unrecognized headers include incorrect device specific headers and incorrect or unimpleme...

Page 132: ...ram message unit MEAS VOLT DC MEASCURR DC 104 Data type error The parser recognized a data element different than one allowed for example numeric or string data was expected but block data was encountered 108 Parameter not allowed More parameters were received than expected for the header for example the KLOCk command only accepts one parameter so receiving SYSTem KLOCk 1 0 is not allowed 109 Miss...

Page 133: ... inconsistency with the number of instruments in the selected group 120 Numeric data error This error as well as errors 121 through 129 are generated when parsing a data element which appears to be numeric including the non decimal numeric types This particular error message should be used if the device cannot detect a more specific error 121 Invalid character in number An invalid character for th...

Page 134: ...ement was encountered but was not allowed by the device at this point in parsing 160 Block data error This error as well as errors 161 through 169 are generated when parsing a block data element This particular error message should be used if the device cannot detect a more specific error 161 Invalid block data A block data element was expected but was invalid for some reason see IEEE 488 2 7 7 6 ...

Page 135: ...th the device s capabilities A valid program message could not be properly executed due to some device condition Execution errors shall be reported by the device after rounding and expression evaluation operations have taken place Rounding a numeric data element for example shall not be reported as an execution error Events that generate execution errors shall not generate Command Errors device sp...

Page 136: ...ations for example the device was not ready to respond Note a DT0 device always ignores GET and treats TRG as a Command Error 213 Init ignored Indicates that a request for a measurement initiation was ignored as another measurement was already in progress 220 Parameter error Indicates that a program data element related error occurred This error message should be used when the device cannot detect...

Page 137: ...ror bit bit 3 in the event status register IEEE 488 2 section 11 5 1 to be set The meaning of positive error codes is device dependent and may be enumerated or bit mapped the error message string for positive error codes is not defined by SCPI and available to the device designer Note that the string is not optional if the designer does not wish to implement a string for a particular error the nul...

Page 138: ...trument has detected a problem with the message exchange protocol described in IEEE 488 2 chapter 6 The occurrence of any error in this class shall cause the query error bit bit 2 in the event status register IEEE 488 2 section 11 5 1 to be set These errors correspond to message exchange protocol errors described in IEEE 488 2 section 6 5 One of the following is true An attempt is being made to re...

Page 139: ...rror Code Description 400 Query error This is the generic query error for devices that cannot detect more specific errors This code indicates only that a Query Error as defined in IEEE 488 2 11 5 1 1 7 and 6 3 has occurred ...

Page 140: ...t Setting Output ON delay time F 01 0 00s Output OFF delay time F 02 0 00s V I mode slew rate select F 03 0 CV high speed priority Internal resistance setting F 08 0 000Ω Bleeder circuit control F 09 1 ON Buzzer ON OFF control F 10 1 ON OCP Delay Time F 12 0 1sec Current Setting Limit F 13 0 OFF Voltage Setting Limit F 14 0 OFF Display Memory parameter when recalling F 15 0 OFF Auto parallel Contr...

Page 141: ...ble UART Baudrate F 71 7 115200 UART Data Bits F 72 1 8 bits UART Parity F 73 0 None UART Stop Bit F 74 0 1 bit UART Transmission Control Protocol F 75 0 SCPI UART Address F 76 30 UART Multi Drop control F 77 0 Disable Power On Configuration setting Setting Default Setting CV Control F 90 0 Power On Configuration CC Control F 91 0 Control by local Output Status when Power ON F 92 0 Safe Mode Alway...

Page 142: ...F100 0 trigger controlled by trigger level Trigger Input Action F102 0 None Output State When Receiving Trigger F103 0 OFF Apply Voltage Setting on Trigger F104 0 0V Apply Current Setting on Trigger F105 0 0A Recall memory number F106 1 M1 Trigger Output Pulse Width F120 0ms Trigger Output Level F121 0 LOW Trigger Source F122 0 None ...

Page 143: ... Over current protection FAN FAIL Fan failure SHUT DOWN Force shutdown Err 001 USB mass storage is not present Err 002 No such file in USB mass storage Err 003 Empty memory location Err 004 File access error Err 007 Slave occurs Off line Multi drop mode Normal Messages Description MSG 001 External control of output Output off F 94 0 High on MSG 002 External control of output Output off F 94 1 Low ...

Page 144: ...T3PS Series Programming Manual 144 LED ASCII Table Character Set Use the following table to read the LCD display messages 0 1 2 3 4 5 6 7 8 9 A B C D E F G H I J K L M N O P Q R S T U V W X Y Z ...

Page 145: ...conversion 144 List of features 11 Messages 143 Model differences 10 Optional accessories 12 Package contents 12 Power on off safety instruction 5 Rear panel diagram 18 Remote control 28 Command list 51 Command syntax 48 Error list 131 Ethernet configuration 41 interface configuration 29 local bus configuration 34 37 multi unit configuration 34 37 multi unit function check 40 sockets configuration...

Page 146: ...to market Designers engineers and educators rely on Teledyne Test Tools solutions to meet their most challenging needs for testing education and electronics validation Location and Facilities Headquartered in Chestnut Ridge New York Teledyne Test Tools and Teledyne LeCroy has sales service and development subsidiaries in the US and throughout Europe and Asia Teledyne Test Tools and Teledyne LeCroy...

Reviews: