RBRsolo and RBRduet operation and maintenance 

This section of the manual covers the basic operation of the RBRsolo and RBRduet data 
logger: changing the battery, inspecting and replacing O-rings, inserting desiccant, etc.  
Please see the Ruskin RBRsolo and RBRduet manual for instructions relating to logger 
programming and data retrieval.   

 

Opening and closing the logger 

 

To open an RBRsolo/duet  logger, unscrew the sensor end counter-clockwise while holding 
on to the yellow housing.  Once fully unscrewed, the housing can be carefully slid away from 
the sensor end cap, revealing the sensor carriage. 

The sensor carriage contains the battery compartment; desiccant holder and USB port (Figs. 1 
and 2). 

 

 

Fig. 1 RBRsolo/duet sensor carriage and housing 

 

Fig. 2 RBRsolo/duet sensor carriage, reverse side 

Summary of Contents for RBRsolo

Page 1: ...ccant etc Please see the Ruskin RBRsolo and RBRduet manual for instructions relating to logger programming and data retrieval Opening and closing the logger To open an RBRsolo duet logger unscrew the...

Page 2: ...eries you must use 3 6V lithium thionyl chloride cells The instrument will also not operate with rechargeable batteries RBR recommends using Tadiran brand batteries only part TL 5903 or TL 4903 in you...

Page 3: ...2014 Fig 5 Battery orientation label Communicating with the logger To establish communication between the RBRsolo duet and a computer open the instrument and remove the housing There is a micro USB c...

Page 4: ...and programmed the logger the instrument is ready to be deployed There are three precautions you should take to avoid damaging the logger 1 Heed the maximum pressure rating of 1700m This is indicated...

Page 5: ...logger is care of the O rings Any kind of water leak can damage the circuitry beyond repair and cause complete data loss Every logger s seal depends upon its O rings not the end cap tightness Proper...

Page 6: ...cleaning never use any material or tool which could scratch the O ring or any of its mating surfaces If any dirt is present in the O ring groove then the O ring should be removed as described below t...

Page 7: ...f oceanographic equipment replace all O rings before every deployment as a matter of routine The cost of an O ring is negligible compared with the cost of the instrument and its deployment Routine rep...