OPERATING INSTRUCTION

TENS+EMS DUO

Summary of Contents for TENS+EMS SUPER DUO

Page 1: ...1639 OPERATING INSTRUCTION TENS EMS DUO ...

Page 2: ...plies with the essential requirements of Directive 93 42 EEC on medical products Manufacturer The manufacturer wants to signal that it complies with the regulations of the Packaging Ordinance 3 Signs and symbols 4 Contents of the packaging 5 Getting to know your device 7 Important notes 12 Components of the device 12 Operating instructions step by step 13 Insertion changing of the batteries 14 Con...

Page 3: ...ations you will merely notice a slight tingling or vibrating sensation The electrical impulses that are sent into the tissue influence the transmission of stimulation into nerves nerve centres and muscle groups in the application area Electrostimulation usually only has an effect after regular applications With regard to muscles electrostimulation does not replace regular training However it is a ...

Page 4: ...rong muscle contractions could affect the healing process The device must never be used near to the heart The stimulation elec trodes must not be placed on any part of the front ribcage where the ribs and breastbone are located especially not on the two large pecto scientifically substantiated and approved as a form of medical treatment Any symptoms that could be relieved using TENS must be checke...

Page 5: ... that no metallic objects e g belt buckles or necklaces come into contact with the electrodes during stimulation If you are wearing jewellery or have piercings in the area to be treated e g a navel piercing rals this can increase the risk of ventricular fibrillation and induce cardiac arrest On the skeletal skull structure or around the mouth throat or larynx In the neck carotid artery area In the...

Page 6: ...ent these must be removed before using the device Failure to do so could result in spot burns Keep the device away from children Make sure not to confuse the electrode cables including contacts with your headphones or other devices and do not connect the electrodes to other devices Do not use the device whilst using other devices that transmit electrical impulses into your body Do not use in the v...

Page 7: ...r Apply let dry and attach the electrode as described in this manual In this way you decrease the risk of skin irritation and prolong the life of your electrodes 5 When you remove the electrodes always pull them in the direction of the hair growth 6 It may be helpful to apply skin lotion to the area of the electrode place ment when you are not wearing the electrodes Please refer to the overview of...

Page 8: ...itching the device off 1 The device turns off automatically after treatment time has elapsed 2 Press the OFF key for approximately three 3 seconds to switch off the device manually The screen goes blank and the device turns off 3 In an emergency you can remove the plug from the device and subse quently remove the pads NOTE Do not remove the device while it is still switched on to avoid unplea sant...

Page 9: ...ODE again and the program number flashes If you want to select the program of the selected modality use the ON increase or OFF decrease key 3 Press MODE again to save your setting The next time you switch on the device the time set by you will appear again NOTE a change in treatment programs during treatment leaves the chosen duration of treatment unchanged This can only be changed manually as de ...

Page 10: ...u should feel and advantages Recommendation P1 Training preparation This program warms the muscles gently before training it feels like a rhythmic massage Increase the intensity until you achieve a strong but soothing muscle movement Duration 10 minutes P2 Muscle strengthening This program uses a pulse frequency which let the muscle fibres pulsate It improves the oxygen free capacity of the muscle...

Page 11: ...xation dura tion 20 minutes P8 Building up of endurance This program uses a low frequency pulse sequence for the develop ment of oxygen dependent capacity and the supply of capillary vessels from which slow twitching muscle fibres profit It improves resistance to fatigue during prolonged exercise with moderate intensity The training includes a number of work phases which are interrupted by longer ...

Page 12: ...ent is interrupted in case the electrodes have not been correctly placed and or in case of an entry to change the mode The device is reset to the lowest level of intensity and 0 is indicated on the display BATTERY STATUS DISPLAY The indicator light for the battery comes on when the battery s energy is de pleted Troubleshooting Before use you should check the device and accessories to avoid damage ...

Page 13: ...sten the cloth with a mild soapy solution Do not use any chemical or abrasive cleaning agents Ensure that no water enters the device Reuse of the device Once it has been properly prepared the device can be used again Prepara tion includes replacement of the treatment electrodes as well as cleaning of the surface of the device using a cloth moistened with a mild soapy solution Storage Remove the ba...

Page 14: ...with the above can impair the performance of the device Disposal WARNING The empty completely flat batteries should be disposed of through specially designated collection boxes recycling points or electronics retailers You are legally required to dispose of the batteries For environmental reasons do not dispose of the device in the hou sehold waste at the end of its useful life Dispose of the unit...

Page 15: ...an send the device free of cost and what documentation is required Filing a warranty claim can only come into consideration if the buyer can submit a copy of the invoice proof of pur chase and the original product to EUROMEDICS or to a service cen tre nominated by EUROMEDICS upon demand The warranty period is 2 years from the date of purchase of a new un used product by the purchaser The warranty ...

Page 16: ...y Well Life Healthcare Ltd 6F No 168 Lide St Jhonghe District New Taipei City 23512 Taiwan EXCLUSIVELY IMPORTED AND DISTRIBUTED IN EUROPE BY EUROMEDICS GmbH Hauptstraße 169 D 59846 Sundern Germany info euromedics de www euromedics de prorelax com 1639 ...