background image

MadgeTech, Inc.

PO Box 50 

l

 

Warner, NH 03278

Phone 603.456.2011

 

l

 

Fax 603.456.2012

www.madgetech.com

 

l

 

[email protected]

Red 485
Dark Blue Pantone 289
Light Blue 58% of Pantone 289

REV F 2010.04.15

To view the full MadgeTech product line,

visit our website at 

www.madgetech.com.

QuadTemp

4-Channel Thermocouple Based Temperature Recorder

Red 485
Dark Blue Pantone 289
Light Blue 58% of Pantone 289

Product Information Card

QuadTemp and OctTemp

OctTemp

8-Channel Thermocouple Based Temperature Recorder

Product Quick Reference Card

Part Number

QuadTemp

OctTemp

 Internal Channel

Temperature Sensor

 Semiconductor

 Internal Channel

Temperature Resolution

0.05°C

Internal Channel Accuracy

+0.5°C

Remote Channel Temperature

Sensor Range, Resolution & Accuracy

*See Table for Details

Cold Jct. Compensation

Automatic

Channels

1 internal & 4 remote

1 internal & 8 remote

Memory

500,000/channel

500,000/channel

Sample Rate

4 seconds up to 24 hours

LED Indicator

None

Required Interface Package

 IFC110 or IFC200

Baud Rate

115,200

Typical Battery Life

18 months typical

Operating Environment

–20 to +60°C, 0 to 95%RH  (Non–Condensing)

Material

Anodized aluminum

Dimensions

7.24” x 2.7” x 1.02” 

(183mm x 68mm x 26mm)

7.24” x 2.7” x 1.26” 

(183mm x 68mm x 32mm)

Approvals

CE

Specifications subject to change.

See MadgeTech’s terms and conditions at www.madgetech.com

Device Maintenance

Battery Replacement

Materials:

3/32” HEX Driver (Allen Key)

Replacement Battery (U9VL-J)

- Remove the cover from the device by unscrewing the four 

screws.

- Remove the battery from its compartment and unsnap it 

from the connector.

- Snap the new battery into the terminals and verify it is secure.

- Replace the cover taking care not to pinch the wires. Screw 

the enclosure back together securely.

Note: Be sure not to over tighten the screws or strip the threads.

Battery Warning

WARNING: FIRE, EXPLOSION, AND SEVERE BURN HAZARD.  DO NOT 

SHORT CIRCUIT, CHARGE, FORCE OVER DISCHARGE, DISASSEMBLE, 

CRUSH,  PENETRATE  OR  INCINERATE.  BATTERY  MAY  LEAK  OR 

EXPLODE IF HEATED ABOVE 60°C (140°F).

Recalibration

 

The  QuadTemp  or  OctTemp  standard  calibration  is  one  point 

at  25°C  for  the  internal  temperature  sensor  and  0mVs  for  the 

thermocouple channels.

Pricing:

Recalibration traceable to NIST 

$60.00

Recalibration 

$40.00

 

Additional:

As Found Data 

$15.00 per parameter/channel

Verification Point 

$15.00 per point

Channel (1st) 

$30.00 at 25°C

 

$45.00 at custom point

Additional Channels 

$3.00 at 25°C

 

$4.50 at custom point

To  send  the  devices  back,  visit  www.madgetech.com,  select  

Services

 then 

RMA Process

.

* Remote Channel Range, Resolution & Accuracy 

Thermocouple

Range (°C)

Resolution

Accuracy

J

-210 to +760

0.1°C

+0.5°C

K

-260 to +1370

0.1°C

+0.5°C

T

-260 to +400

0.1°C

+0.5°C

E

-260 to +980

0.1°C

+0.5°C

R

-50 to +1760

0.5°C

+2.0°C

S

-50 to +1760

0.5°C

+2.0°C

B

+60 to +1820

0.5°C

+2.0°C

N

-260 to +1300

0.1°C

+0.5°C

Reviews: