DIN-Rail DC Strain Gage  Conditioner

TECHNICAL MANUAL

Copyright © 2016 LORD Corporation • All Rights Reserved

Part No. 234200B   DCN 10087

237 Commerce Drive • Amherst, NY 14228 • USA

Tel: 716.250.1900 • Fax: 716.250.1909

Web: stellartech.com • Email: [email protected]

Due to the nature of technology, changes are inevitable. For latest technical specifications, see our website.

ISO 9001/AS9100

1 of 9

Summary of Contents for AP5103 Series

Page 1: ...All Rights Reserved Part No 234200B DCN 10087 237 Commerce Drive Amherst NY 14228 USA Tel 716 250 1900 Fax 716 250 1909 Web stellartech com Email info stellartech com Due to the nature of technology...

Page 2: ...tion technique which is outlined in section three The supplied calibration resistor is 59 kilohms 1 which is located internally to the AP5103 case Series AP5103 DC Strain Gage Module Access switch set...

Page 3: ...ent damage Common Mode Rejection Ratio 60 dB 1 2 excitation Input Impedance Differential and Common Mode 10 000 M Offset user adjustable vs Temperature 30 ppm C vs Time 10 ppm month Gain Accuracy Limi...

Page 4: ...e recommended transducer cabling would be eight wire individually shielded twisted pair wired as indicated Fig 1 Sense lines must be connected at the transducer as recommended or at the AP5103 screw t...

Page 5: ...e transducer calibration data mentions some other resistance value In Section 4 of this manual the techniques described above are demonstrated If however the installed value of calibration resistor is...

Page 6: ...to obtain your nominal full scale analog output Use Fine Span control to adjust the signal to the precise value required or for an analog value that correspondes with the appropriate Shunt Calibration...

Page 7: ...mote Calibration Check The instrument can be placed in the calibration mode positive or negative by shorting pin Power Common and SHP for positive or SHN for negative on the lower I O connector Figure...

Page 8: ...is determined by the direction of the loading or off zero reading C The inability to balance correctly where the instrument output reads totally off scale and the BALANCE controls have no authority c...

Page 9: ...ance will be made for any expenses incurred for correcting any defective workmanship and or material without written consent by Stellar Technology This warranty is in lieu of all other warranties expr...