background image

INSTRUCTION MANUAL

ENGLISH

MM

700

10A

40M

Ω

• DATA & RANGE HOLD

• LOW IMPEDANCE

• AUDIBLE CONTINUITY

• MIN / MAX / RELATIVE

• TEMPERATURE

• DIODE TEST

• CAPACITANCE &

FREQUENCY

Auto-Ranging 

Digital Multimeter

True RMS

Measurement 

Technology

Summary of Contents for MM700

Page 1: ...10A 40MΩ DATA RANGE HOLD LOW IMPEDANCE AUDIBLE CONTINUITY MIN MAX RELATIVE TEMPERATURE DIODE TEST CAPACITANCE FREQUENCY Auto Ranging Digital Multimeter Digital Multimeter Digital Multimeter Digital Multimeter True RMS Measurement Technology ...

Page 2: ... to CSA STD C22 2 61010 1 61010 2 030 61010 2 033 IEC EN 61010 1 61010 2 030 61010 2 033 61326 1 Pollution degree 2 Accuracy of reading of least significant digits Drop Protection 6 6 ft 2m Ingress Protection IP42 except test lead jacks see WARNINGS Safety Rating CAT IV 600V CAT III 1000V Class 2 Double insulation CAT III Measurement category III is applicable to test and measuring circuits connec...

Page 3: ...0V DC CURRENT AUTO RANGING AC Current μA and mA 400 0μA 0 1μA 1 5 3 digits 4000μA 1μA 40 00mA 10μA 400 0mA 100μA 10 00A 0 01A 2 0 5 digits DC Current μA and mA 400 0μA 0 1μA 1 0 3 digits 4000μA 1μA 40 00mA 10μA 400 0mA 100μA 10 00A 0 01A 1 5 3 digits Overload Protection 500mA 1000V and 10A 1000V Fuses Frequency Range 50 to 400Hz Maximum Input µA mA setting 400mA AC RMS DC 10A setting 10A AC RMS DC...

Page 4: ... or 600V AC RMS FREQUENCY AUTO RANGING 9 999Hz 0 001Hz 1 0 5 digits 99 99Hz 0 01Hz 999 9Hz 0 1Hz 9 999kHz 1Hz 99 99kHz 10Hz 500 0kHz 100Hz Sensitivity 8V RMS Maximum Input 600V DC or 600V AC RMS DUTY CYCLE 1 to 99 9 0 1 1 2 2 digits Pulse width 0 1 100ms Frequency width 5Hz to 10kHz Sensitivity 8V RMS Maximum Input 600V DC or 600V AC RMS TEMPERATURE 0 to 1500 F 0 1 F 1 F 2 0 9 F 18 to 816 C 0 1 C ...

Page 5: ...g a known voltage or current Never use the meter on a circuit with voltages that exceed the category based rating of this meter Do not use the meter during electrical storms or in wet weather Do not use the meter or test leads if they appear to be damaged Use only with CAT IV rated test leads Ensure meter leads are fully seated and keep fingers away from the metal probe contacts when making measur...

Page 6: ...can lead to risk of electrical shock Follow all warnings and instructions detailed in this manual ENGLISH SYMBOLS ON LCD H Data Hold Audible Continuity Diode Auto Ranging AC Alternating Current DC Direct Current Low Battery Auto Power Off MAX Maximum Value MIN Minimum Value F Degrees Fahrenheit C Degrees Celsius M Mega value x 106 k kilo value x 103 m mili value x 10 3 μ micro value x 10 6 n nano ...

Page 7: ...6 NOTE There are no user serviceable parts inside meter 1 4000 count LCD display 7 RANGE button 2 Function selector switch 8 REL Relative button 3 10A jack 9 MAX MIN button 4 COM jack 10 HOLD button 5 VΩ jack 11 SEL Select button 6 mA µA jack ...

Page 8: ...between F and C between Hz and Duty cycle and between Continuity Resistance and Diode test The default function for each application is printed on the meter in white the secondary functions are printed on the meter in orange HOLD DATA HOLD BACKLIGHT BUTTON Press HOLD 10 to hold the measurement on the display Press again to release the display to return to live measuring Press and hold the HOLD but...

Page 9: ... is pressed the meter keeps track of the minimum and maximum value of the measurement as the meter continues to take samples 1 When measuring press MAX MIN 9 and the meter displays the maximum value If a new maximum occurs the display updates with that new value Pressing again reveals the minimum value If a new minimum occurs the display updates with the new value 2 Press MAX MIN 9 and hold for mo...

Page 10: ...letely TESTING IN CAT III CAT IV MEASUREMENT LOCATIONS Ensure the test lead shield is pressed firmly in place Failure to use the CAT III CAT IV shield increases arc flash risk TESTING IN CAT II MEASUREMENT LOCATIONS CAT III CAT IV shields may be removed for CAT II locations This will allow testing on recessed conductors such as standard wall outlets Take care not to lose the shields 5 32 4 mm 7 18...

Page 11: ...OTE If appears on the LCD the test leads are being applied to the circuit in reverse Swap the position of the leads to correct this NOTE When in a voltage setting and the test leads are open readings of order mV may appear on the display This is noise and is normal By touching the test leads together to close the circuit the meter will measure zero volts AC DC LO Z VOLTAGE LESS THAN 600V Voltage m...

Page 12: ... DC currents 400mA and 10A Insert RED test lead into 10A jack 3 and BLACK test lead into COM jack 4 and rotate function selector switch 2 to the 10A AC DC setting For mA AC DC currents 400mA Insert RED test lead into mA µA jack 6 and BLACK test lead into COM jack 4 and rotate function selector switch 2 to the mA AC DC setting For µA DC currents 400µA Insert RED test lead into μA mA jack 6 and BLAC...

Page 13: ... the SEL button 11 until the icon appears 2 Remove power from circuit 3 Test for continuity by connecting conductor or circuit with test leads If resistance is measured less than 50Ω an audible signal will sound and display will show a resistance value indicating continuity If circuit is open display will show OL OPERATING INSTRUCTIONS DO NOT attempt to measure continuity on a live circuit RESISTA...

Page 14: ...witch 2 to the Continuity Resistance Diode Capacitance setting NOTE The meter defaults to Continuity testing in this mode To enter Diode testing mode press the SEL button 11 twice The Diode icon will appear on the display 2 Touch test leads to diode A reading of 200 700mV on display indicates forward bias OL indicates reverse bias An open device will show OL in both polarities A shorted device wil...

Page 15: ... Measure capacitance by connecting test leads across the capacitor The meter will auto range to display the measurement in the most appropriate range FREQUENCY DUTY CYCLE 1 Insert RED test lead into VΩ jack 5 and BLACK test lead into COM jack 4 and rotate function selector switch 2 to the Frequency Duty Cycle setting NOTE The meter defaults to Frequency testing in this mode To enter Duty Cycle tes...

Page 16: ...display will stabilize The meter will auto range to display the measurement in the most appropriate range Remove thermocouple before switching meter to other measurement functions The thermocouple included with the original purchase is suitable for temperatures below 356 F 180 C only To measure higher temperatures a K type thermocouple with the appropriate measurement range should be used BAR GRAP...

Page 17: ...ied to the 10A jack 3 To access fuses 1 Remove screw from battery fuse door 2 Replace blown fuse s with μA mA jack 6 500mA 1000V fast blow Klein Cat No 69035 10A jack 3 10A 1000V fast blow Klein Cat No 69034 3 Replace battery fuse door and fasten securely with screw To avoid risk of electric shock disconnect leads from any voltage source before removing battery fuse door To avoid risk of electric ...

Page 18: ...nditions exceeding the limits mentioned in the General Specifications section allow the meter to return to normal operating conditions before using DISPOSAL RECYCLE Do not place equipment and its accessories in the trash Items must be properly disposed of in accordance with local regulations Please see www epa gov or www erecycle org for additional information CUSTOMER SERVICE 99 Washington Street...