background image

INSTRUCTION MANUAL

ENGLISH

1000V

10A

60M

Ω

•  HIGH VISIBILITY 

DISPLAY

•  DATA HOLD

•  LOW IMPEDANCE

•  AUDIBLE CONTINUITY

•  MIN / MAX / RELATIVE

•  TEMPERATURE

•  DIODE TEST

•  CAPACITANCE & 

FREQUENCY

Auto-Ranging 

Digital Multimeter

True RMS

Measurement 

Technology

MM720

2m

IP42

CAT IV

600V

CAT III

1000V

5000573

www.calcert.com

[email protected]

1.800.544.2843

0

5

10

15

20

25

30

Summary of Contents for 092644693762

Page 1: ...LOW IMPEDANCE AUDIBLE CONTINUITY MIN MAX RELATIVE TEMPERATURE DIODE TEST CAPACITANCE FREQUENCY Auto Ranging Digital Multimeter True RMS Measurement Technology MM720 2m IP42 CAT IV 600V CAT III 1000V 5...

Page 2: ...3 85 7 47 7 mm Weight 12 2 oz 345 g Calibration Accurate for one year Standards Conforms to UL STD 61010 1 61010 2 030 61010 2 033 Certified to CSA STD C22 2 61010 1 61010 2 030 61010 2 033 EC EN 6101...

Page 3: ...A 0 1 A 1 0 5 digits 6000 A 1 A 60 00mA 10 A 600 0mA 100 A 6 000A 1mA 2 0 3 digits 10 00A 0 01A 2 0 5 digits DC Current A and mA 600 0 A 0 1 A 1 0 3 digits 6000 A 1 A 60 00mA 10 A 600 0mA 100 A 6 000...

Page 4: ...its 99 99Hz 0 01Hz 999 9Hz 0 1Hz 9 999kHz 1Hz 99 99kHz 10Hz 500 0kHz 100Hz Voltage Range 2V to 220V RMS Maximum Input 600V DC or 600V AC RMS DUTY CYCLE 1 0 to 99 9 0 1 1 2 2 digits Pulse Width 0 1 100...

Page 5: ...uit with voltages that exceed the category based rating of this meter Do not use the meter during electrical storms or in wet weather Do not use the meter or test leads if they appear to be damaged Us...

Page 6: ...ectrical Shock Improper use of this meter can lead to risk of electrical shock Follow all warnings and instructions detailed in this manual ENGLISH SYMBOLS ON LCD H Data Hold Audible Continuity Diode...

Page 7: ...er 1 6000 count Reverse Contrast Display 7 RANGE button 2 Function selector switch 8 REL Relative button 3 10A jack 9 MAX MIN button 4 COM jack 10 HOLD Backlight button 5 V jack 11 SEL Select button 6...

Page 8: ...meter in white the secondary functions are printed on the meter in orange HOLD DATA HOLD BACKLIGHT BUTTON Press HOLD 10 to hold the measurement on the display Press again to release the display to ret...

Page 9: ...ON When the MAX MIN button 9 is pressed the meter keeps track of the minimum and maximum value of the measurement as the meter continues to take samples 1 When measuring press MAX MIN 9 and the meter...

Page 10: ...ENT LOCATIONS Ensure the test lead shield is pressed firmly in place Failure to use the CAT III CAT IV shield increases arc flash risk TESTING IN CAT II MEASUREMENT LOCATIONS CAT III CAT IV shields ma...

Page 11: ...TE If appears on the LCD the test leads are being applied to the circuit in reverse Swap the position of the leads to correct this NOTE When in a voltage setting and the test leads are open readings o...

Page 12: ...rt RED test lead into 10A jack 3 and BLACK test lead into COM jack 4 and rotate function selector switch 2 to the 10A AC DC setting For mA AC DC currents 600mA Insert RED test lead into mA A jack 6 an...

Page 13: ...utton 11 until the icon appears 2 Remove power from circuit 3 Test for continuity by connecting conductor or circuit with test leads If resistance is measured less than 50 an audible signal will sound...

Page 14: ...otate function selector switch 2 to the Continuity Resistance Diode Capacitance setting NOTE The meter defaults to Continuity testing in this mode To enter Diode testing mode press the SEL button 11 t...

Page 15: ...est leads across the capacitor The meter will auto range to display the measurement in the most appropriate range FREQUENCY DUTY CYCLE 1 Insert RED test lead into V jack 5 and BLACK test lead into COM...

Page 16: ...e make contact between the thermocouple tip and the object being measured When thermocouple tip and object are in thermal equilibrium the measurement on the display will stabilize The meter will auto...

Page 17: ...move screw from battery fuse door 2 Replace blown fuse s with A mA jack 6 800mA 1000V fast blow Klein Cat No 69390 10A jack 3 10A 1000V fast blow Klein Cat No 69034 3 Replace battery fuse door and fas...

Page 18: ...ts mentioned in the General Specifications section allow the meter to return to normal operating conditions before using FCC IC COMPLIANCE See this product s page at for FCC compliance information Can...

Reviews: