background image

IRwin Quick Start Guide

E

ective October 2019

Southern Cross Inc.

   www.southerncrossinc.com

3175 Corners North Court 

Peachtree Corners, GA 30071

1.800.241.5057 

Summary of Contents for Southern Cross IRwin

Page 1: ...IRwin Quick Start Guide Effective October 2019 Southern Cross Inc www southerncrossinc com 3175 Corners North Court Peachtree Corners GA 30071 1 800 241 5057 ...

Page 2: ...o set the lower explosive limit to 5 vol 7 How to set unit to one point calibration 9 How to set the threshold alarm 14 Data transfer Bluetooth 16 Charging the battery 19 Troubleshooting error messages 22 How to set your leak detector in Universal Mode 5 ...

Page 3: ...How to Assemble your IRwin Methane Leak Detector Components of the IRwin 3 Bell Probe Probe Rod Hand Probe ...

Page 4: ...our IRwin Methane Leak Detector 1 Attach the probe rod to the bell probe 2 Attach the hand probe to the probe rod Connect the probe tube of the hand probe to the gas inlet on the leak detector Your unit is assembled 3 4 ...

Page 5: ...it For Operation in countries not following German DVGW directives 1 Start the device by pressing the ON button Press followed by 2 3 4 5 Answer the function test questions and confirm test results by pressing Use to navigate to General and press Use to navigate to Custom and press 5 ...

Page 6: ... for Universal Mode 6 Use to navigate to Universal Mode and press 7 Enter the Supervisor password code 1111 unless changed by supervisor 8 Use to navigate to Universal Mode and press Press to tick Universal Mode box 9 10 Press followed by and 6 ...

Page 7: ...ct 100 CH4 LEL Conc Use to navigate to Custom and press How to set the Lower Explosive Limit LEL at 5 0 vol on IRwin Methane Leak Detector 1 5 If prompted enter the password code 1111 unless changed by supervisor and repeat step 4 If not prompted go ahead to the next step 7 ...

Page 8: ...How to set the Lower Explosive Limit LEL at 5 0vol on IRwin Methane Leak Detector 6 Use to select 5 0 vol in the drop down menu press 7 Press followed by and 8 ...

Page 9: ...Universal Mode Use to navigate to General and press 1 Make sure your IRwin is set in Universal Mode see the How to set your IRwin for Universal Mode quick start guide 2 3 4 5 Press followed by Use to navigate to AbG Cal Setup and press Use to navigate to Custom and press 9 ...

Page 10: ...d press followed by 8 9 Use and to navigate to Calibration and press Use to navigate to Settings and press 10 If prompted enter the min Intermediate password code If not prompted go ahead to next step 6 If prompted enter the Supervisor Advanced password code If not prompted go ahead to next step 10 ...

Page 11: ...tor for 1 Point Calibration 2 5 Press to get to the drop down menu 12 Press on the first choice of the drop down menu 13 Navigate to far right drop down menu and press 14 Use to select Disabled and press 15 Press again to view drop down menu 11 ...

Page 12: ...o select Disabled and press 19 20 In Universal Low Range use to select the second choice in the drop down menu press Navigate to far right drop down menu and press Press again to view the drop down menu In Universal Low Range use to select the third choice in the drop down menu press 12 ...

Page 13: ... 2 5 21 22 23 Select Set by using and and press Navigate to far right drop down menu and press Use to navigate to drop down menus Set the actual value of the calibration gas you have nominally 2 2 or 2 5 by entering 2 and press 2 5 and press 24 Press followed by and 13 ...

Page 14: ...ppm on your IRwin Methane Leak Detector Use to navigate to Thresholds and press 1 2 3 4 Press followed by If prompted enter the Intermediate password code and repeat step 3 If not prompted go ahead to the next step To select CxHy press 14 ...

Page 15: ...5 Press to open the drop down menu Use to navigate to 100 and press 6 How to set the Threshold Alarm at 100ppm on your IRwin Methane Leak Detector Press followed by and 15 ...

Page 16: ... computers will prompt you to acknowledge that a connection password has been correctly transferred to the detector You must confirm this on both computer and detector Start the program IRwin App on your computer In the IRwin App on your computer Click Connection menu item Port and select your detector Data Transfer Bluetooth The serial number of the leak detector is displayed in the drop down men...

Page 17: ...sis as this can severely affect the GC analysis 5 3 4 Real time streaming from leak detector to computer You have installed IRwin App software on your computer You have activated the Bluetooth interface on your computer Switch on the leak detector Perform self test 17 4 5 6 7 8 9 10 11 12 1 2 Set the leak detector in the desired mode This must be one of the following operating modes Bar Hole Gas P...

Page 18: ...While streaming you can change mode as desired Streamed data cannot be saved Fig 10 Streamed measurement data 18 ...

Page 19: ...dels The only equipment allowed to be connected to the Charging socket of any of the IRwin SXnn models is the Ex Certified IRwin Charging Adapter Connect this adapter to the leak detector first and then connect the charger or car cable to the adapter NOTICE Fast charging reduces the lifetime of the battery Do not use fast charging regularly The ambient temperature during the charging process shoul...

Page 20: ...rging socket and charging plug to each other Connect the Charger for IRwin S models 100 240V 580 605 or Charger cable for car 591 361 to the Certified IRwin Charging Adapter 580 604 Plug the extension cable into the charger and connect to outlet The status LED is green when battery is charging normally The status LED goes out when charging is completed 2 3 1 Charging IRwin S from serial number 929...

Page 21: ...t button The detector is now charging fast and the status LED turns cyan Normal Fast charge 100 4 h from empty to full charge produces 9 5 h operation time 3 h from empty to full charge produces 9 h operation time Short charge 1 h charging gives 3 25 h of operation 1 h charging gives 4 h of operation ...

Page 22: ...alarm remains Contact INFICON service W4 Exhaust Blocked Exhaust blocked Check that nothing is blocking the exhaust If alarm remains Contact INFICON service W5 O2 sensor error System error Restart detector If problems remains Contact INFICON service W6 O2 sensor error System error Restart detector If problems remains Contact INFICON service W7 H2 S sensor error System error Restart detector If pro...

Page 23: ...thout any gas present Recalibrate TOX mode If still reappearing Contact INFICON service W21 CO2 STEL level reached STEL and limit shown Confirm Short Term Exposure alarm for CO2 Normal function If warning reapperas without any gas present Recalibrate TOX mode If still reappearing Contact INFICON service W22 CO2 LTEL level reached LTEL and limit shown Confirm Long Term Exposure alarm for CO2 Normal...

Page 24: ...Pump Valve test has red cross in function test result screen Exhaust blocked Restart detector If problems remains Contact INFICON service Pump dirty or broken Contact INFICON service E13 Pump Valve test has red cross in function test result screen Valve clogged Restart detector If problems remains Contact INFICON service System error Restart detector If problems remains Contact INFICON service E14...

Page 25: ...blems remains Contact INFICON service E21 ROM test error Power off Memory damaged or test error Restart detector If problems remains Contact INFICON service E22 CRC error power off Memory damaged or test error Restart detector If problems remains Contact INFICON service E23 Gas Sensors has red cross in function test result screen Sensor error or warm up drift Restart detector If problems remains C...

Reviews: