background image

Carbon dioxide

Synonyms: Carbonic acid gas, Carbonic anhydride, Carbon (IV) oxide; Dry ice (in
solid state).

Chemical formula

CO

2

Molar weight

44

Relative gas density (to air)

1.52

Conversion

1 ppm = 1.80 mg/m

3

Boiling point

Sublimes

Flammability

Nonflammable

Odour

Odourless

Hazards

Dusts  of  various  metals  (Mg,  Zr,  Ti,  Al,
Cr, Mn) are ignitable and explosive when
suspended in carbon dioxide.
In  concentrations  up  to  1%  (10 000
ppm),  it  will  make  some  people  feel
drowsy  and  give  the  lungs  a  stuffy
feeling.  Concentrations  of  7%  to  10%
(70 000  to  100 000  ppm)  may  cause
suffocation,  even  in  the  presence  of
sufficient 

oxygen, 

manifesting 

as

dizziness, headache, visual and hearing
dysfunction, and unconsciousness within
a few minutes to an hour.

Exposure 

limits

(Directive 2006/15/EC)

TWA

9000 mg/m

3

 /5000 ppm

STEL

Not specified

Conversion of ppm to mg/m

3

 is calculated for 25°C and 1 atm.

Indoor air quality in nonresidential buildings

CO2 level

Description

< 450 ppm

High quality (fresh air)

450-600 ppm

Medium quality

600-1000 ppm

Moderate  quality

> 1000 ppm

Low quality

Installation guidelines

(See Installation and connections section for general information.)
Since CO

2

 is heavier than air, it is recommended to locate the sensor not higher, than

the potential leakage. For air quality control place the sensor in the breathing zone.

Calibration

E2608-CO2  series  devices  have  been  calibrated  by  Manufacturer  with  standard
gas  mixtures  before  delivery.  Provided  that  the  sensor  is  used  under  moderate
conditions, field recalibration is recommended every  5 years. Please contact your
dealer for more information.

Maintenance

Do not perform any maintenance operation with the power on.
Clean the device with soft damp cloth. Do not use any abrasive cleaning agents. Do
not immerse the device into water or any cleaning media.

Delivery set

–Detector-transmitter E2608 (wall mount or duct mount version)
– Mounting accessories:

4 cross-shaped mounting lug with screws and 4 screws with plastic dowels

for wall mount version

rubber flange for duct mount version
fixing clamp for remote probe versions

E2608-CO2-10K_UM_EN 

         Rev 11.07.2019

Specifications

Sensor type

optical (NDIR)

Sampling method

diffusion

Detection range

0...10 000 ppm

Resolution / digital unit

1 ppm

Accuracy

± (50 ppm + 3% of reading)

in the range 300...5000 ppm

Response time T90

ca. 2 minutes

Sensor lifetime

ca. 10 years

Calibration interval

up to 5 years

Signal update

every 1 second

Power supply options

11...30 VDC, 24 VAC

or 90...265 VAC (with mains power unit)

Power consumption

< 2 VA

Analog outputs

2 × 4-20 mA / 0-10 V, user settable

Outputs assignment

OUT1 2 gas; OUT2 2 gas

Output scale width

> 10 × resolution

Recommended: 20-100% of the detection range

Digital interface

RS485, Modbus RTU protocol

no galvanic isolation

Load resistance

R

L

 < (Us - 2 V) / 22 mA for 4-20 mA

R

L  

> 250 kOhm for 0-10 V mode

Relay otputs

2 × SPST, max 5 A, 30 VDC / 250 VAC

Alarm setpoints

Determined by user within 5-95% of the detection range

Cable connections

screwless spring loaded terminals

Enclosure

light grey ABS, wall or duct mount, protection class IP65

Dimensions

H85 × W82 × D55 mm

Operating conditions

0...+50 °C, 0...85 %RH; atmospheric pressure ±10%*

non ATEXrated areas, non-aggressive atmosphere

CE marking

according to 2014/30/EU and EN61326-1 requirements

Other options

*Ambient pressure
compensation option

in the range 0,5...2 bar of absolute pressure

Warranty

This  product  is  warranted  to  be  free  from defects  in  material  and  workmanship  for  a
period  of  one  year  from  the  date  of  original  sale.  During  this  warranty  period
Manufacturer  will,  at  its  option,  either  repair  or  replace  product  that  proves  to  be
defective. This warranty is void if the product has been operated in conditions outside
ranges  specified  by  Manufacturer  or  damaged  by  customer  error  or  negligence  or  if
there has been an unauthorised modification.

    Evikon MCI OÜ

    Teaduspargi 7, Tartu

[email protected]

    Tel. +372 733 6310          50411 Estonia

www.evikon.eu

PluraSens

®

Carbon Dioxide Detector-Transmitter

E2608-CO2-10K

User Manual

Summary of Contents for E2608-CO2-10K

Page 1: ...ies 4 cross shaped mounting lug with screws and 4 screws with plastic dowels for wall mount version rubber flange for duct mount version fixing clamp for remote probe versions E2608 CO2 10K_UM_EN Rev 11 07 2019 Specifications Sensor type optical NDIR Sampling method diffusion Detection range 0 10 000 ppm Resolution digital unit 1 ppm Accuracy 50 ppm 3 of reading in the range 300 5000 ppm Response ...

Page 2: ...nections via RS485 interface should not exceed 1200 m Note The outputs are not galvanically isolated from 24 V power supply and share common 0V Allowed load resistance limits are stated in Specifications table To power the instrument from an external 24 VDC source connect terminals 0V and U to the source If the integrated mains power supply module is used connect terminals L and N to the mains Not...

Page 3: ...dec Default 0x0001 1 40002 R Hardware version 0x0002 2 40003 R Software version 0x0003 3 40004 R Product serial number 1 65535 0x0004 4 40005 RW Slave ID net address 1 247 1 0x0005 5 40006 RW Baudrate 1200 2400 4800 9600 19200 38400 57600 9600 0x0006 6 40007 RW Response delay ms 1 255 10 0x0007 7 40008 RW Stop bits parity bit 1 no parity bit 1 stop bit default after factory reset 2 no parity bit 2...

Page 4: ...t in MHR 40205 2 0x00CB 203 40204 RW Forced value for analog output 1 0 1000 0 0 100 0 of output scale 0 0x00CC 204 40205 RW Forced value for analog output 2 0 1000 0 0 100 0 of output scale 0 0x00D3 211 40212 RW Parameter tied to relay RE1 0 none 1 temperature 2 gas concentration 9 control by Modbus control state set in MHR 40214 2 0x00D4 212 40213 RW Parameter tied to relay RE2 0 none 1 temperat...

Page 5: ... 1 1 Hz 50 on 50 off LED signal off on if relay1 turned on bit 12 0 1 2 Hz 50 on 50 off LED signal off on if relay2 turned on user defined 0x0100 256 40257 R Raw temperature data C 100 signed integer 4000 8500 40 00 85 00 C 0x0101 257 40258 R Raw gas sensor data ADC data 0 4095 0x0102 258 40259 R Measured temperature C 100 signed integer 4000 12500 40 00 125 00 C 0x0103 259 40260 R Gas concentrati...

Reviews: