background image

Operation Manual

ergometrics er900 PC

Art-Nr: 475.156 

!

 

Version:  12/01

Summary of Contents for ER900 PC

Page 1: ...Operation Manual ergometrics er900 PC Art Nr 475 156 Version 12 01...

Page 2: ...is only permitted after prior written consent by the manufacturer This Operation Manual is not registered for automatic update in case of eventual alterations Information regarding the latest version...

Page 3: ......

Page 4: ...Version 12 01 er900 PC Operation Manual Art Nr 475 156...

Page 5: ...Device Protecting Features 9 2 7 Measurement Accuracy Verification 9 2 7 1 Routine Control of the Blood Pressure Measurement Unit 9 2 7 2 Periodic Accuracy Verification for Blood Pressure Unit 9 2 7...

Page 6: ...6 1 General Preparations 21 6 2 Preparations for External Control via ECG or PC 22 6 3 Start of the Ergometry Session 22 6 4 Manual Operation of Ergometry Session 22 6 5 Operational Panel for Manually...

Page 7: ...be done at your ECG device This handbook is intended for operators of the unit It explains all the functions of the er gometer in detail and in such a manner that operation of the unit can be learned...

Page 8: ...Operation Manual er900 PC Operation Manual 4 Version 12 01 Art Nr 475 156 Ergometer Type er900 PC...

Page 9: ...ded via special statements by the operating com pany e g for supervision and obligatory registrations internal maintenance plans test ing intervals 1 1 Comments by the Manufacturer for the Operating C...

Page 10: ...g Electromagnetic compatibility 1 3 Warranty Regulation ergoline GmbH is only then responsible for the safety and reliability of this device when all revisions enhancements repairs and service to the...

Page 11: ...ergometer For such purpose special leaning ergometer or offered by ergoline c Application under outdoor conditions is not permitted d In case the ergometer is not being used according to above conditi...

Page 12: ...sonal 2 4 Normative Notices 2 4 1 Notice for EMC electromagnetic compatibility The EMC test was done with a interference level of 3V m The device should not be used in areas with a higher level The de...

Page 13: ...pabilities are cause for re certification 2 7 1 Routine Control of the Blood Pressure Measurement Unit The automatic blood pressure measurement unit in the console of the ergometer must be approved an...

Page 14: ...Operation Manual er900 PC Operation Manual 10 Version 12 01 Art Nr 475 156 Ergometer Type er900 PC 2 8 Identification plate Labelling Safety Notices label with serial number...

Page 15: ...surement head with the tin flaps into the handlebar pipe Attach grounding cable connection cable and tubing 3 Adjust black rear cover onto the measurement head and tighten with screw 4 Insert saddle p...

Page 16: ...ric size 200 850 saddle for children incl saddle pipe and sadle bellows 475 156 er900PC Operation Manual 475 190 er900 Service Unterlagen 3 4 Transport Please ensure that the mains cable is removed pr...

Page 17: ...ation the device Minimal fuse value 16A 4 1 Operating elements and Connections Putting into Operation 1 power connector fuse 2 main power switch 3 clamp handle for seat adjustment 4 clamp handle for h...

Page 18: ...156 Ergometer Type er900 PC 4 2 Measurement head 4 2 1 Keyboard Putting into Operation 1 LCD Display Enable blood pressure measurement Initiate or stop blood pressure measurement Increase load level...

Page 19: ...ect the unit s power plug to one of the white outlets on the utility cart The ergometer can then be switched on and off together with the suction device 4 3 2 Connector for Blood Pressure Cuff The cuf...

Page 20: ...3 1 Digital RS 232 interface connector for PC or ECG 2 Analog interface for analog control and analog data transfer to ECG 3 Fiber optic connector for ear finger clip or HR input from and ECG The int...

Page 21: ...gometer Type er900 PC 4 4 Power on Sequence Mains power switch The unit will subsequently execute a self test At the end of the self test the main menu will appear as below Putting into Operation P 00...

Page 22: ...PC without an ECG device The load controll can be done through 5 Watt steps by using the 5W and 5W keys The NIBP can be controlled through the two corresponding keys Use of the Blood Pressure Module...

Page 23: ...of the last blood pressure reading current heart rate in case of continuous HR signal t 00 min duration of the ergometry in min S 000 mmHg systolic value of the last blood pressure reading the mea sur...

Page 24: ...Operation Manual er900 PC Operation Manual 20 Version 12 01 Art Nr 475 156 Ergometer Type er900 PC Operation...

Page 25: ...at the test subject can comfortably sit upright To do this loosen the handlebar height clamp handle and adjust appropriately Next you can adjust the angle of the handlebar by loosening the handlebar a...

Page 26: ...ed Configure your ECG or PC in accordance with the appropriate operator s instructions for ergometry Start the ergometric session with the ECG or PC 6 3 Start of the Ergometry Session Start the ergome...

Page 27: ...d decreased in steps of Watt The pos sible load covers the range from 20 Watt to 995 Watts With the key Blutdruck I O the blood pressure measurement is enabled On or disabled Off The current status is...

Page 28: ...damage the unit s finish and lead to defects Recommended are disinfectants like Fugaten spray lysoform or Promanum N 7 2 Cleaning the Blood Pressure Cuff The blood pressure cuff may also be cleaned w...

Page 29: ...admissable patient weight 120 kg Seat adjustment continuously adjustable for body sizes 1 20m to 2 10m Handle bar adjustment continuously adjustable for body sizes 1 20m to 2 10m Display 1x LCD Displa...

Page 30: ...peration Manual 26 Version 12 01 Art Nr 475 156 Ergometer Type er900 PC Annex A Technical Data RPM independent range in accordance with DIN 13405 RPM independent range er900 The Speed correlated Capac...

Reviews: