TDAS G5 

Data Acquisition System 

User’s Manual

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 
 

 

 

 

 

 

October 2019 

 

Summary of Contents for TDAS G5

Page 1: ...11000 00010 MAN Rev 5 Diversified Technical Systems Inc All Rights Reserved TDAS G5 Data Acquisition System User s Manual October 2019...

Page 2: ...Considerations 9 Power Considerations 10 Power Consumption and Management 10 Power Requirements and Memory Retention 10 External Power Provisions 10 Communication Features 11 Communication Method 11 L...

Page 3: ...ail or on site visits The best way to contact a DTS support engineer is to submit a request through the DTS Help Center web portal support dtsweb com You must be registered support dtsweb com registra...

Page 4: ...e 216 position interface connector LED indicators for communication calibration power arm status and event status Back up battery input for safe operation even with loss of primary power Please see th...

Page 5: ...checks are performed under software control settings are calculated by the software and the applied shunt signal is automatically scaled to an appropriate percentage of the requested range See the TD...

Page 6: ...gh bandwidth EID Support Each TDAS G5 DAS measurement channel supports communication with silicon serial number devices manufactured by Dallas Semiconductor Maxim Integrated Products When you connect...

Page 7: ...treated with great care as any debris solvents or oil even from fingers can compromise the integrity of the connections signals Proper orientation can be determined by locating the board references J1...

Page 8: ...tempting any cleaning procedure When not in use or if shipping is required we suggest that you always place the unit in the padded carrying case originally provided with your system WARNING The gold p...

Page 9: ...as possible will minimize self heating It is unlikely that the units will be damaged by excessive heat because power is internally limited by self resetting thermal fuses the units will simply shut do...

Page 10: ...oring data After the data collection routine is complete the TDAS G5 DAS de energizes several circuits to minimize power consumption It takes a maximum of 125 seconds for 70 seconds of data at 10k sam...

Page 11: ...n status with the docking station and DAS CAL Provides calibration status PWR Provides status on power to the DAS ARM Provides arm start record and event status STAT Provides general status TDAS G5 DA...

Page 12: ...cle applications as needed In dummy Docking Station Configuration The iDummy docking station allows for easy installation and removal of the TDAS G5 DAS into virtually any crash dummy type Signals ent...

Page 13: ...for on board testing applications Power commu nication and event signals are easily accessible via the control panel A variety of larger sensor connectors such as LEMO 1B or equivalent can be easily...

Page 14: ...TDAS G5 DAS User s Manual October 2019 support dtsweb com 14 11000 00010 MAN Rev 5 Appendix A TDAS G5 DAS 216 position Interface Connector...

Page 15: ...AN Rev 5 Appendix B Mechanical Specifications Weight 200 grams TDAS G5 DAS TDAS G5 DAS standard high bandwidth 3 374 inches 85 7 mm 2 677 inches 68 mm 0 472 inches 12 mm 0 158 inches diam 4 0 mm diam...

Page 16: ...DAS Firmware Loader provided with the TDAS Control software establish communication with the DAS unit To do this change the fields to the appropriate hardware G5 serial number of the DAS unit and comm...

Page 17: ...Command fpget name read parameter fpset Set parameter set parameter into flash memory Command fpset name value set parameter Confirm or change the TDAS S N field Type the command on the Input Command...

Page 18: ...d or any text editor 4 Find the line titled Rack Inventory and change the IP address associated with the serial number for the TDAS G5 DAS to the new address Note Change only the IP address and not th...

Page 19: ...om the factory with the default IP address shown on the packing list for the equipment If the packing list is not available and you need information on the specifics of your equipment please submit a...

Page 20: ...d TDAS G5 DAS Interface Connector section Updated Appendix D DTS Support boilerplate and support link in footer Removed HQ address Rev 5 04 Dec 2013 EK Updated DTS Support Input Range Hardware Filters...